ARROW IPN035564 (REF CDC-42703-XP1A)Recalls & reported problems
Teleflex Incorporated · Multi-Lumen Central Venous Catheterization Kit · Class 2 · Implant
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20801902058588), received between Jun 2, 2021 and Aug 10, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Entrapment of Device” (1 report).
54 other recalls cover devices cleared under the same FDA submission (K071538); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Entrapment of Device1
- Unraveled Material1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 10, 2021Entrapment of Device; Entrapment of Device
COMPLAINT FOUND IN MAUDE DATABASE REPORTED AS: "A RIGHT IJ TRIPLE LUMEN CATHETER WAS INSERTED IN A PT WITH SEVERE SEPSIS. AT THE CONCLUSION OF THE PROCEDURE, IT WAS NOTED THE GUIDEWIRE WAS RETAINED WITHIN THE CATHETER AND PARTIALLY IN THE IVC. THE CATHETER WAS NOT FREE-FLOATING AND WAS SUCCESSFULLY RETRIEVED BY IR. UNFORTUNATELY, GUIDEWIRES HAVE A KNOWN DESIGN "FLOW" IN THAT THERE WAS NO HARD-STOP DESIGN MEASURES, OTHER THAN THE PROFESSIONAL USER, TO PREVENT A GUIDEWIRE FROM 'SLIPPING AWAY' INTO THE CATHETER ITSELF."
Patient problems: Foreign Body In Patient
MAUDE report 12302964 ↗MalfunctionJun 2, 2021Unraveled Material; Unraveled Material
THE CUSTOMER REPORTED "THE GUIDE WIRE FROM THE INCLUDED ARROW CVC KIT COMPLETELY UNWOUND DURING DEPLOYMENT LEADING TO LOOSE WIRE ATTEMPTING TO ENTER THE PATIENTS' BLOODSTREAM. ER DOCTOR WAS ABLE TO REMOVE AND RETRIEVE THE WIRE." NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11924696 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071538. One submission often covers many models, so check the product and lot details in each record.
- Z-2376-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
- Z-1914-2026 · Mar 11, 2026 · Open, ClassifiedARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4
- Z-2123-2026 · Jan 7, 2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
- Z-0002-2024 · Aug 10, 2023 · Open, ClassifiedArrow ErgoPack Complete, Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-42854- PCN1A; catheter, intravascular, therapeutic, long-term greater than 30 days
- Z-0003-2024 · Aug 10, 2023 · Open, ClassifiedArrow Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
- Z-0351-2023 · Oct 27, 2022 · TerminatedArrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducer
- Z-0352-2023 · Oct 27, 2022 · TerminatedPressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-
- Z-1604-2022 · May 25, 2022 · Open, ClassifiedTELEFLEX LLC Arrow Central Venous Catheter Kit 16, 18 Gauge Multi Lumen Model Number:CDC-45703-XP1A (Model Numbers # IPN035619, IPN036437, IPN922677)
- Z-1994-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PLUS Pressure Injectable Multi-Lumen CVC Kit; Product Codes: CDC-45703-P1A; CDC-45703-PB1A; CDC-42703-P1A; and CDC-15703-P1A
- Z-1991-2018 · Apr 11, 2018 · TerminatedPressure Injectable Quad-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Codes: ASK-45854-PSFM a
- Z-1998-2018 · Apr 11, 2018 · TerminatedMulti-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-427
- Z-1990-2018 · Apr 11, 2018 · TerminatedPressure Injectable Two-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Code: ASK-45802-PHMW
- Z-1992-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PLUS Pressure Injectable Two-Lumen CVC Kit; Product Code: CDC-42802-P1A
- Z-1993-2018 · Apr 11, 2018 · TerminatedPressure Injectable ARROWg+ard Blue PLUS(R) Two-Lumen Central Venous Catheterization Kit; Product Code: ASK-42802-PWHC1
- Z-1988-2018 · Apr 11, 2018 · TerminatedPressure Injectable ARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Codes: ASK-42703-PAM1, ASK-42703-PSHR1, ASK-45703-PNY, ASK-45703-PWHC1
…and 39 more.
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page