Skip to content

ARROW IPN035564 (REF CDC-42703-XP1A)Recalls & reported problems

Teleflex Incorporated · Multi-Lumen Central Venous Catheterization Kit · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20801902058588), received between Jun 2, 2021 and Aug 10, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Entrapment of Device” (1 report).

54 other recalls cover devices cleared under the same FDA submission (K071538); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Entrapment of Device1
  • Unraveled Material1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 10, 2021Entrapment of Device; Entrapment of Device

    COMPLAINT FOUND IN MAUDE DATABASE REPORTED AS: "A RIGHT IJ TRIPLE LUMEN CATHETER WAS INSERTED IN A PT WITH SEVERE SEPSIS. AT THE CONCLUSION OF THE PROCEDURE, IT WAS NOTED THE GUIDEWIRE WAS RETAINED WITHIN THE CATHETER AND PARTIALLY IN THE IVC. THE CATHETER WAS NOT FREE-FLOATING AND WAS SUCCESSFULLY RETRIEVED BY IR. UNFORTUNATELY, GUIDEWIRES HAVE A KNOWN DESIGN "FLOW" IN THAT THERE WAS NO HARD-STOP DESIGN MEASURES, OTHER THAN THE PROFESSIONAL USER, TO PREVENT A GUIDEWIRE FROM 'SLIPPING AWAY' INTO THE CATHETER ITSELF."

    Patient problems: Foreign Body In Patient

    MAUDE report 12302964 ↗
  • MalfunctionJun 2, 2021Unraveled Material; Unraveled Material

    THE CUSTOMER REPORTED "THE GUIDE WIRE FROM THE INCLUDED ARROW CVC KIT COMPLETELY UNWOUND DURING DEPLOYMENT LEADING TO LOOSE WIRE ATTEMPTING TO ENTER THE PATIENTS' BLOODSTREAM. ER DOCTOR WAS ABLE TO REMOVE AND RETRIEVE THE WIRE." NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11924696 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071538. One submission often covers many models, so check the product and lot details in each record.

  • Z-2376-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
  • Z-1914-2026 · Mar 11, 2026 · Open, Classified
    ARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4
  • Z-2123-2026 · Jan 7, 2026 · Open, Classified
    Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
  • Z-0002-2024 · Aug 10, 2023 · Open, Classified
    Arrow ErgoPack Complete, Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-42854- PCN1A; catheter, intravascular, therapeutic, long-term greater than 30 days
  • Z-0003-2024 · Aug 10, 2023 · Open, Classified
    Arrow Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
  • Z-0351-2023 · Oct 27, 2022 · Terminated
    Arrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducer
  • Z-0352-2023 · Oct 27, 2022 · Terminated
    Pressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-
  • Z-1604-2022 · May 25, 2022 · Open, Classified
    TELEFLEX LLC Arrow Central Venous Catheter Kit 16, 18 Gauge Multi Lumen Model Number:CDC-45703-XP1A (Model Numbers # IPN035619, IPN036437, IPN922677)
  • Z-1994-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Multi-Lumen CVC Kit; Product Codes: CDC-45703-P1A; CDC-45703-PB1A; CDC-42703-P1A; and CDC-15703-P1A
  • Z-1991-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Quad-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Codes: ASK-45854-PSFM a
  • Z-1998-2018 · Apr 11, 2018 · Terminated
    Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-427
  • Z-1990-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Two-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Code: ASK-45802-PHMW
  • Z-1992-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Two-Lumen CVC Kit; Product Code: CDC-42802-P1A
  • Z-1993-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable ARROWg+ard Blue PLUS(R) Two-Lumen Central Venous Catheterization Kit; Product Code: ASK-42802-PWHC1
  • Z-1988-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable ARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Codes: ASK-42703-PAM1, ASK-42703-PSHR1, ASK-45703-PNY, ASK-45703-PWHC1

…and 39 more.

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page