Skip to content

ARROW IPN033104Recalls & reported problems

Teleflex Incorporated · Cannon(R) II Plus Hemodialysis Catheter · Class 2 · Implant

All models of ARROW (1488)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20801902096610), received between Jan 26, 2026 and Apr 8, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K111117); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 10, 2026Class IOpen, Classified

    16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

    Z-2158-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Crack1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2026Material Separation

    IT WAS REPORTED THAT: "THE HUB CONNECTION SEPARATED FROM THE ASSEMBLY WHEN DR. W. WAS ATTEMPTING TO PLACE THE CATHETER. THE HUB ASSEMBLY FELL ON THE FLOOR, SO WE HAD TO OPEN ANOTHER CATHETER KIT. THE WAS NOT ON THE PATIENT YET. IT WAS AFTER OPENING THE KIT. THE PACKAGING WAS NOT DAMAGED. NO HARM CAME TO THE PATIENT. WE HAD TO OPEN ANOTHER CATHETER KIT. THE PATIENT IS STABLE AND HAS BEEN COMING TO DIALYSIS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24814136 ↗
  • MalfunctionJan 26, 2026Crack

    IT WAS REPORTED THAT: " THE PATIENT JUST STARTED DIALYSIS TREATMENT AND THE ARTERIAL HUB BROKE, WHEN CAP CAME OFF THE DAMAGE WAS SEEN. NO AIR OR PARTICULATE ENTERED THE PATIENT. THE ARTERIAL HUB WAS BEING CLEANED AFTER REMOVING THE CAP. THE PATIENT WAS NOT HOOKED UP TO THE DIALYSIS MACHINE. NO INJURIES OCCURRED. THE ISSUE WAS RESOLVED BY CHANGING OUT THE CATHETER FOR A NEW ONE. THE MEDICAL INTERVENTION WAS A CATHETER EXCHANGE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176229 ↗

FDA clearance and classification

Cleared through Special 510(k) K111117 on Jun 30, 2011. Submitted by Teleflexmedical, Inc. as “Cannon Ii Plus Chronic Hemodialysis Catheter, Edge Chronic Hemodialysis Catheter, Nextstep(R) Retrograde Chronic...”. The FDA review took 70 days. FDA record ↗

The FDA regulates product code MSD (catheter, hemodialysis, implanted) under 21 CFR 876.5540 as a Class II device. 21 CFR 876.5540 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Recalls under the same FDA submission

These recalls involve devices cleared under K111117. One submission often covers many models, so check the product and lot details in each record.

  • Z-2156-2026 · Apr 10, 2026 · Open, Classified
    Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I
  • Z-2159-2026 · Apr 10, 2026 · Open, Classified
    NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep A
  • Z-1007-2015 · Feb 4, 2013 · Terminated
    The Arrow Edge Hemodialysis Catheterization Set Long-Term Access and NextStep Antegrade Hemodialysis Catheterization Set, Long-Term Access, part numbers: AC-15192-SFX, AC-15232-SFX

Common questions about the ARROW IPN033104

Has the ARROW IPN033104 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2158-2026 (Class I), began on Apr 10, 2026: 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ARROW IPN033104?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 26, 2026 and Apr 8, 2026. In those reports, the most common specific device problems are “Crack” (1) and “Material Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ARROW IPN033104?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ARROW IPN033104 FDA approved?

It was cleared through Special 510(k) K111117 on Jun 30, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the ARROW IPN033104 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the ARROW IPN033104, and is it still sold?

The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN033104). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page