Simplexa COVID-19 Direct MOL4150Recalls & reported problems
Diasorin Molecular LLC · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20816101027020), received between Nov 13, 2024 and Nov 13, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “False Positive Result” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jan 30, 2023Class IIOpen, Classified
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Z-1209-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- False Positive Result1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2024False Positive Result; False Positive Result
DIASORIN MOLECULAR LLC RECEIVED A COMPLAINT ON (B)(6) 2024, ALLEGING FALSE POSITIVE RESULTS WITH THE SIMPLEXA COVID-19 DIRECT ASSAY WITH ONLY 1 OF 2 TARGETS DETECTED (NO S-GENE, BUT ORF1AB DETECTED). UPON REPEAT ON A COMPETITOR ASSAY (BD-MAX), THE RESULTS WERE NEGATIVE. THE CUSTOMER CONFIRMED NO ALLEGED HARM OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20679515 ↗
FDA clearance and classification
The FDA regulates product code QQX (respiratory specimen nucleic acid sars-cov-2 test) under 21 CFR 866.3981 as a Class II device. 21 CFR 866.3981 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: A device to detect and identify nucleic acid targets in respiratory specimens from SARS-CoV-2 that cause COVID-19 is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 nucleic acid targets in human clinical respiratory specimens from patients suspected of COVID-19 based on signs and symptoms of respiratory infections by their healthcare providers. The device is intended to aid in the diagnosis of COVID-19 in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Common questions about the Simplexa COVID-19 Direct MOL4150
Has the Simplexa COVID-19 Direct MOL4150 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1209-2023 (Class II), began on Jan 30, 2023: Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Simplexa COVID-19 Direct MOL4150?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 13, 2024. That report lists device problem “False Positive Result” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Simplexa COVID-19 Direct MOL4150?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Simplexa COVID-19 Direct MOL4150, and is it still sold?
The FDA's device database lists Diasorin Molecular LLC as the company responsible for it (brand Simplexa COVID-19 Direct, model MOL4150). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page