Skip to content

3100A HFOV 773997Recalls & reported problems

Carefusion Corporation · LONG PAT CKT, F&P850, BOX 4, 31A · Class 3

All models of 3100A HFOV (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20846446003373), received between Dec 18, 2018 and Feb 5, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Disconnection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Disconnection1

Patient problems

  • Low Oxygen Saturation1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 5, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER REPORTED A DEFECTIVE 3100A CIRCUIT, THE CUSTOMER EXPLAINED THAT WHEN THEY CONNECT THE CIRCUIT ON THE UNIT, THE HEATER ALARMED, THEY TRIED A NEW TEMP PORTS AND HUMIDIFIER BUT THE DEVICE STILL ALARMING. THE CUSTOMER REPLACED THE CIRCUIT AND SOLVED THE ISSUE. THE PROBLEM HAPPENED WHILE CONNECTED ON A PATIENT, HOWEVER, THE CUSTOMER ALSO CONFIRMED THAT THERE WAS NO HARM OR INJURY THAT WAS ASSOCIATED WITH THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9673498 ↗
  • InjuryDec 18, 2018Disconnection; Disconnection

    ADDITIONAL CUSTOMER INFORMATION RECEIVED ON 18DEC2018: THE CUSTOMER REPORTED THAT THE PATIENT WAS MANUALLY VENTILATED AND EXPERIENCED A LOW OXYGEN SATURATION DURING THIS EVENT.

    Patient problems: Low Oxygen Saturation; No Known Impact Or Consequence To Patient

    MAUDE report 8173434 ↗

Other Ventilator, High Frequency devices

All Ventilator, High Frequency devices →