Reference Patch CARTO 3 CREFP6Recalls & reported problems

Biosense Webster Inc · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 20846835000129), received between Oct 19, 2022 and Nov 6, 2025 — classified as 2 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 5 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Tear, Rip or Hole in Device Packaging7
  • Device Markings/Labelling Problem4
  • Expiration Date Error4
  • Delivered as Unsterile Product3
  • Manufacturing, Packaging or Shipping Problem2
  • Patient Device Interaction Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • Anaphylactic Shock1
  • Cardiac Arrest1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2025Expiration Date Error; Device Markings/Labelling Problem; Expiration Date Error; Device Markings/Labelling Problem

    IT WAS REPORTED THERE WERE SOME INCONSISTENCES BETWEEN THE REFERENCE PATCH CARTO 3 DEVICE AND THE PACKAGE LABEL. IT WAS REPORTED THAT WHEN THE PACKAGE WAS OPEN, IT WAS FOUND THAT THE LOT AND EXPIRATION DATE ON THE DEVICE WERE INCONSISTENT WITH THOSE ON THE PACKAGE AND THERE WERE MIXING OF EXPIRED MATERIALS. THE DEVICES WERE NOT USED ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23493849 ↗
  • MalfunctionNov 6, 2025Expiration Date Error; Device Markings/Labelling Problem; Expiration Date Error; Device Markings/Labelling Problem

    IT WAS REPORTED THERE WERE SOME INCONSISTENCE BETWEEN THE REFERENCE PATCH CARTO 3 DEVICE AND THE PACKAGE LABEL. IT WAS REPORTED THAT WHEN THE PACKAGE WAS OPEN, IT WAS FOUND THAT THE LOT AND EXPIRATION DATE ON THE DEVICE WERE INCONSISTENT WITH THOSE ON THE PACKAGE AND THERE WERE MIXING OF EXPIRED MATERIALS. THE DEVICES WERE NOT USED ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23493847 ↗
  • MalfunctionNov 6, 2025Expiration Date Error; Device Markings/Labelling Problem; Expiration Date Error; Device Markings/Labelling Problem

    IT WAS REPORTED THERE WERE SOME INCONSISTENCES BETWEEN THE REFERENCE PATCH CARTO 3 DEVICE AND THE PACKAGE LABEL. IT WAS REPORTED THAT WHEN THE PACKAGE WAS OPEN, IT WAS FOUND THAT THE LOT AND EXPIRATION DATE ON THE DEVICE WERE INCONSISTENT WITH THOSE ON THE PACKAGE AND THERE WERE MIXING OF EXPIRED MATERIALS. THE DEVICES WERE NOT USED ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23493833 ↗
  • MalfunctionNov 6, 2025Expiration Date Error; Device Markings/Labelling Problem; Expiration Date Error; Device Markings/Labelling Problem

    IT WAS REPORTED THERE WERE SOME INCONSISTENCES BETWEEN THE REFERENCE PATCH CARTO 3 DEVICE AND THE PACKAGE LABEL. IT WAS REPORTED THAT WHEN THE PACKAGE WAS OPEN, IT WAS FOUND THAT THE LOT AND EXPIRATION DATE ON THE DEVICE WERE INCONSISTENT WITH THOSE ON THE PACKAGE AND THERE WERE MIXING OF EXPIRED MATERIALS. THE DEVICES WERE NOT USED ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23493815 ↗
  • MalfunctionNov 3, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A REFERENCE PATCH CARTO 3. REFERENCE PATCHES ARRIVED WITH THE INTERNAL SEAL BREACHED. THEREFORE, FOR SAFETY, THE MEDICAL TEAM CHOSE NOT TO USE THE MATERIAL. NO RISK WAS CAUSED TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED. THEY DID NOT TAKE ANY PHOTOS AS THEY WERE IN A STERILE OPERATING ROOM. THE PRODUCT WAS DISPOSED OF ACCORDING TO LOCAL HOSPITAL PROTOCOL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23447481 ↗
  • MalfunctionJun 20, 2025Tear, Rip or Hole in Device Packaging; Manufacturing, Packaging or Shipping Problem; Tear, Rip or Hole in Device Packaging; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION PROCEDURE WITH A REFERENCE PATCH CARTO 3 AND WHEN THE MEDICAL TEAM OPENED THE OUTER BOX, THE INNER BAG WAS NOT SEALED WITH ADHESIVE AND WAS LEFT OPEN. THE REFERENCE PATCH WAS USED ON THE PATIENT BASED ON THE PHYSICIAN'S DECISION, AS IT IS NOT A STERILE PRODUCT. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294753 ↗
  • MalfunctionMar 28, 2025Tear, Rip or Hole in Device Packaging; Manufacturing, Packaging or Shipping Problem; Tear, Rip or Hole in Device Packaging; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A REFERENCE PATCH CARTO 3. WHEN THEY OPENED THE PACKAGING FOR THE PATCHES, EACH OF THE PATCHES WERE NOT SEALED. ADVISED THE CALLER TO NOT USE THE PATCH SET. NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED. THEY DID NOT TAKE ANY PHOTOS OF THE OPEN PACKAGING. THE PACKAGING WAS SENT BACK BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21721810 ↗
  • MalfunctionMar 27, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A REFERENCE PATCH CARTO 3. WHEN OPENED THE BOX, THE STERILE PACKAGING WAS ALREADY UNSEALED. NEW REFERENCE PATCH WAS OPENED AND USED. ISSUE OCCURRED PRE-OP. THE PROCEDURE WAS COMPLETED WITHOUT ANY PROBLEMS. NO PATIENT CONSEQUENCE. ADDITIONAL INFORMATION RECEIVED CONFIRMED THE DEVICE WAS NOT USED ON THE PATIENT. THE PACKAGING NOT AVAILABLE SINCE THE PACKAGING WAS DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21715687 ↗
  • MalfunctionMar 4, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES AND THERE WAS REPORTS OF "CONTAMINATED MATERIAL". NO FURTHER DETAILS WERE PROVIDED REGARDING TROUBLESHOOTING OF THE ISSUE OR COMPLETION OF THE PROCEDURE. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21519463 ↗
  • MalfunctionJan 16, 2025Display or Visual Feedback Problem; Delivered as Unsterile Product; Display or Visual Feedback Problem; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A REFERENCE PATCH CARTO 3 AND THE MEDICAL TEAM STATED THAT THERE WAS 'CONTAMINATION' OF THE MONITORING PATCH. NO FURTHER DETAILS WERE PROVIDED SPECIFYING THE NATURE OF THE CONTAMINATION. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21175310 ↗
  • MalfunctionDec 10, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE AND UPON OPENING, THE SEAL PART WAS TORN. THE DEVICE WAS NOT USED ON THE PATIENT. THEY REPLACED THE REFERENCE PATCH WITH ANOTHER ONE. NO ADVERSE PATIENT CONSEQUENCE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20898501 ↗
  • InjurySep 18, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION AND THE PATIENT EXPERIENCED ALLERGIC SKIN REACTION FROM THE CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES. THE ALLERGIC SKIN REACTION REQUIRED MEDICATION. PATIENT HAD AN ALLERGIC REACTION TO ALL OF THE PATCHES (ECG REFERENCE PATCHES, INDIFFERENT ELECTRODE, AND CARTO 3 SYSTEM REFERENCE PATCHES). THE RASH WAS DISCOVERED POST PROCEDURE. THE PATIENT WAS GIVEN HYDROCORTISONE AND BENADRYL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT WAS STABLE.

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 20251301 ↗
  • MalfunctionMay 15, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES. WHEN THE REFERENCE PATCH PACKAGING WAS OPENED, THEY NOTICED THAT THE PATCHES WERE NOT COMPLETELY SEALED. WHEN THE PATCH WAS REPLACED, THE ISSUE RESOLVED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19321078 ↗
  • MalfunctionMay 8, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE STERILIZATION POUCH OF THE CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES WERE NOT SEALED. IT WAS REPORTED THAT THE HOSPITAL STAFF OPENED A PACKAGE OF A CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES IN PREPARATION FOR A PROCEDURE AND FOUND THAT TWO BAGS INSIDE WERE NOT SEALED ON ONE OF THE SIDES. THE DEVICES WERE NOT USED ON ANY PATIENT. A REPLACEMENT OF THE SAME LOT WAS OPENED, AND NO PROBLEM WAS OBSERVED. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT'S CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19274580 ↗
  • MalfunctionMay 2, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES AND THE MEDICAL TEAM DETERMINED THAT THE PATCHES WERE CONTAMINATED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19235334 ↗
  • InjuryOct 19, 2022Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN IDIOPATHIC VENTRICULAR TACHYCARDIA (IDVT) ABLATION PROCEDURE WITH A CARTO® 3 SYSTEM EXTERNAL REFERENCE PATCHES AND THE PATIENT SUFFERED CARDIAC ARREST AND POSSIBLE ANAPHYLACTIC REACTION. WHILE BEGINNING TO INTUBATE THE PATIENT THE ANESTHESIA DOCTOR PUSHED VANCOMYCIN AND THE PATIENT CODED. THE BWI REPRESENTATIVE STATED THAT THERE WERE STILL VISIBLE P WAVES AND QRS ON THE ECG BUT NO CAROTID PULSE. CHEST COMPRESSIONS WERE PERFORMED AS WELL AS 7 CARDIOVERSIONS. A LIFT TEAM WAS BROUGHT IN ON STANDBY AND THE PHARMACY CAME TO PUSH MEDS. THE PATIENT STABILIZED AND THE PULSE RETURNED, THE CASE THEN PROCEEDED WITHOUT FURTHER INCIDENT. THE PATCHES WERE THE ONLY BIOSENSE WEBSTER PRODUCT INVOLVED AT THE TIME OF THE INCIDENT. THE DOCTOR INQUIRED IF THE CARTO PATCHES COULD CAUSE ANAPHYLACTIC SHOCK. THE BWI REPRESENTATIVE REQUESTED INFORMATION ABOUT THE ADHESIVE MATERIAL ON THE PATCHES AND WAS DIRECTED TO MEDICAL AFFAIRS. THE PATCHES WERE DISCARDED AND ARE NOT AVAILABLE FOR RETURN. CONSERVATIVELY THIS EVENT WILL BE AN ADVERSE EVENT FOR CARDIAC ARREST, ANAPHYLACTIC REACTION. IT IS MOST LIKELY THAT THE VANCOMYCIN AND /OR THE ANESTHESIA MEDICATIONS THAT WER

    Patient problems: Anaphylactic Shock; Cardiac Arrest

    MAUDE report 15636615 ↗

FDA clearance and classification

The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Reference Patch CARTO 3 CREFP6

Has the Reference Patch CARTO 3 CREFP6 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20846835000129), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Reference Patch CARTO 3 CREFP6?

The FDA's MAUDE database lists 16 adverse event reports for this device, received between Oct 19, 2022 and Nov 6, 2025. In those reports, the most common specific device problems are “Tear, Rip or Hole in Device Packaging” (7), “Device Markings/Labelling Problem” (4) and “Expiration Date Error” (4); the most common patient problems are “Anaphylactic Shock” (1), “Cardiac Arrest” (1) and “Skin Inflammation/ Irritation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Reference Patch CARTO 3 CREFP6?

None of the reports are classified as deaths, 2 as injuries and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Reference Patch CARTO 3 CREFP6, and is it still sold?

The FDA's device database lists Biosense Webster Inc as the company responsible for it (brand Reference Patch CARTO 3, model CREFP6). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →