Leica Biosystems 38440501Recalls & reported problems
Leica Biosystems Richmond, Inc. · IP Biopsy Cassette II with lid, Taped, Pink · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20849832022095), received between Nov 6, 2025 and Apr 23, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, THE CUSTOMER CONTACTED LEICA BIOSYSTEMS AND REPORTED "...A CASSETTE LID BROKE AND TISSUE ESCAPED DURING PROCESSING AND WAS LOST." NO ADVERSE IMPACT TO ANY PATIENT HAS BEEN REPORTED TO THE MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24973940 ↗MalfunctionNov 6, 2025Break; Break
ON 13 OCTOBER 2025, LEICA BIOSYSTEMS RECEIVED THE FOLLOWING INFORMATION: "ON-SITE WITH CUSTOMER ON (B)(6) 2025. MANAGER [NAME] INFORMED ME OF ISSUE WITH 38440501: IP BIOPSY CASSETTE II, TAPED, PINK. FLANGE HAS BEEN BREAKING OFF, CASSETTES OPENING DURING PROCESSING, AND TISSUE HAS BEEN LOST. ISSUE HAS OCCURRED ALL WEEK WHERE TISSUE IS BEING LOST. PER [NAME], THIS HAS OCCURRED WITH 3 DIFFERENT PAS EVEN AFTER PARTICULAR CARE IS BEING TAKEN. PAS FEEL THAT CASSETTES SEEM "SOFTER" AND MORE FRAGILE THAN PREVIOUSLY. SWITCHED TO NEW LOT# AND TAKING EXTRA PRECAUTIONS. THIS SITE IS ALSO USING THE SAME CASSETTE IN GREEN AND YELLOW, NO REPORTED ISSUES." NO ADVERSE IMPACT TO A PATIENT HAS BEEN REPORTED TO THE MANUFACTURER.
Patient problems: Insufficient Information
MAUDE report 23491128 ↗MalfunctionNov 6, 2025Break; Break
ON (B)(6) 2025, LEICA BIOSYSTEMS RECEIVED THE FOLLOWING INFORMATION: "ON-SITE WITH CUSTOMER ON (B)(6) 2025. MANAGER [NAME] INFORMED ME OF ISSUE WITH 38440501: IP BIOPSY CASSETTE II, TAPED, PINK. FLANGE HAS BEEN BREAKING OFF, CASSETTES OPENING DURING PROCESSING, AND TISSUE HAS BEEN LOST. ISSUE HAS OCCURRED ALL WEEK WHERE TISSUE IS BEING LOST. PER [NAME], THIS HAS OCCURRED WITH 3 DIFFERENT PAS EVEN AFTER PARTICULAR CARE IS BEING TAKEN. PAS FEEL THAT CASSETTES SEEM "SOFTER" AND MORE FRAGILE THAN PREVIOUSLY. SWITCHED TO NEW LOT # AND TAKING EXTRA PRECAUTIONS. THIS SITE IS ALSO USING THE SAME CASSETTE IN GREEN AND YELLOW, NO REPORTED ISSUES." NO ADVERSE IMPACT TO A PATIENT HAS BEEN REPORTED TO THE MANUFACTURER.
Patient problems: Insufficient Information
MAUDE report 23491120 ↗
Other CASSETTES, TISSUE devices
- Leica Biosystems 39LC-500-9-LLeica Biosystems Richmond, Inc. · CASSETTES, TISSUE2 reports
- Tissue-Tek® Paraform® 7019Sakura Finetek U.S.A., Inc. · CASSETTES, TISSUE1 report
- Tissue-Tek® Paraform® 7020Sakura Finetek U.S.A., Inc. · CASSETTES, TISSUE1 report
- Leica Biosystems 38440502Leica Biosystems Richmond, Inc. · CASSETTES, TISSUE1 report
- Leica Biosystems 39LC-550-2-LLeica Biosystems Richmond, Inc. · CASSETTES, TISSUE1 report
- Leica Biosystems 39LC-500-2-LLeica Biosystems Richmond, Inc. · CASSETTES, TISSUE1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page