Skip to content

WarmTouch 5030810Recalls & reported problems

Covidien LP · Full-Body /Multi Access Blanket · Class 2

All models of WarmTouch (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20884521061795), received between Dec 22, 2020 and Aug 11, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem2

Patient problems

  • Burn(s)1
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 11, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE BLANKET WAS FUNCTIONING CORRECTLY PRIOR TO AND DURING THE PROCEDURE. THE PATIENT WAS LAID ON TOP OF THE BLANKET, INSTEAD OF THE BLANKET BEING DRAPED ON TOP OF THE PATIENT. THEN THE PATIENT URINATED ONTO THE BLANKET, WHICH CAUSED BURNS ON THE PATIENT'S BUTTOCKS. THE BLANKET TEMPERATURE WAS SET TO 42 DEGREES CELSIUS HALFWAY THROUGH TO 38 DEGREES CELSIUS AND WAS WITH THE PATIENT FOR TWO HOURS AND 30 MINUTES. THERE WERE NO SIGNS OF BURNS AFTER THE PROCEDURE AND NOTICED THE FOLLOWING DAY. THE PATIENT WAS REFERRED TO PLASTICS, AND THE WOUND WAS CLEANED.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 22771418 ↗
  • InjuryDec 22, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, THE PATIENT SUFFERED A BURN WHILE USING A BLANKET POST-OPERATIVE. IT WAS CONFIRMED THAT THERE WAS NO ABNORMALITY IN THE TESTING RELATED EQUIPMENT.

    Patient problems: Burn(s)

    MAUDE report 11055020 ↗

Other System, thermal regulating devices

All System, thermal regulating devices →

Sources and updates