Skip to content

TAYLOR SPATIAL FRAME 71070286Recalls & reported problems

Smith & Nephew, Inc. · THICK ITALIAN NUT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20885556119376), received between Dec 14, 2023 and Dec 14, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Material Deformation” (1 report).

1 other recall cover devices cleared under the same FDA submission (K093047); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • Perforation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 14, 2023Material Deformation; Material Deformation

    IT WAS REPORTED THAT, DURING AN EXTERNAL FIXATION SURGERY, A DIFFERENCE BETWEEN THE METAL STRENGTH OF A THICK ITALIAN NUT AND THE 10MM WRENCHES WAS NOTICED; OR THAT THE 10MM WRENCH MIGHT BE SLIGHTLY TOO LARGE FOR THE NUT. THIS RESULTED IN DEFORMING THE NUTS UPON TIGHTENING, AND BURRING THE WRENCHES. AS A CONSEQUENCE, A WRENCH 10MM FLEX HEAD WAS STRICKEN WITH A MALLET, WHICH INADVERTENTLY MALIGNED THE FRAME. ALL THE FIXATION (4 HA PINS) HAD TO BE REMOVED AND RESTARTED, ADDING AT LEAST 1 HOUR TO THE PROCEDURE AND ADDITIONAL HOLES AND INCISIONS TO THE BONE HAD TO BE PERFORMED. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Perforation

    MAUDE report 18331872 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093047. One submission often covers many models, so check the product and lot details in each record.

  • Z-0245-2022 · Jun 17, 2021 · Open, Classified
    Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →