NA 201.928Recalls & reported problems
Synthes GmbH · 2.0MM IMF SCREW SELF-DRILLING 8MM · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 20887587028634), received between Feb 28, 2019 and Feb 5, 2026 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Break” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break12
- Entrapment of Device2
- Device-Device Incompatibility1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient4
- Foreign Body In Patient2
- No Code Available1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Break; Entrapment of Device
IT WAS REPORTED DURING INSERTION, ONE IMF SCREW BROKE. THE HEAD OF THE SCREW BROKE OFF AND WAS REMOVED. THE SHAFT WAS EMBEDDED IN THE PATIENTS BONE AND COULD NOT BE RETRIEVED. NO SURGICAL DELAY. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: Foreign Body In Patient
MAUDE report 24268350 ↗MalfunctionOct 30, 2025Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERTING THE SCREW, THE TIP OF THE SCREW BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23427296 ↗MalfunctionOct 16, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT DURING A MAXILLOFACIAL SURGERY, A IMF SCREW SNAPPED AND FRACTURED IN HALF DURING INSERTION. THE TIP OF THE IMF SCREW HAS BEEN LEFT IN THE PATIENT AS IT WOULD CAUSE TOO MUCH DAMAGE TO REMOVE. THE BROKEN OFF HEAD WAS REMOVED FROM THE PATIENT AND IS BEING SENT FOR FURTHER TESTING.
Patient problems: Foreign Body In Patient
MAUDE report 23317326 ↗MalfunctionAug 5, 2025Break; Break
IT WAS REPORTED THAT THE HARDWARE FRACTURE OCCURRED IN DR. (B)(6) RECONSTRUCTION CASE AT HOSPITAL. UPON REMOVING AN IMF SCREW (PRODUCT CODE: 201.928) TO REMOVE THE PATIENT'S MAXILLOMANDIBULAR FIXATION AT THE END OF THE CASE, THE SCREW HEAD BROKE OFF, LEAVING THE BODY OF THE SCREW INSIDE OF THE PATIENT'S MANDIBLE. NO HARM WAS CAUSED TO THE PATIENT, AS WE WERE ABLE TO BURR OUT THE REMAINDER OF THE SCREW WITH A CARBIDE SIDE CUTTING BURR TO REMOVE ALL REMNANTS OF THE SCREW. THERE WAS NO SURGICAL DELAYED AND THE PROCEDURE SUCCESSFULLY COMPLETED. AND THERE WAS PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22720275 ↗MalfunctionMay 22, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THAT DURING THE SURGERY, WHEN INSERTING THE SCREW, THE SCREW BROKE, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THIS REPORT INVOLVES ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19374889 ↗MalfunctionMay 22, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THAT DURING THE SURGERY, WHEN INSERTING THE SCREW, THE SCREW BROKE, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THIS REPORT INVOLVES ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19374785 ↗MalfunctionApr 1, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWEDEN AS FOLLOWS: IT WAS REPORTED THAT IN (B)(6) 2024, DURING A HIP SURGERY, THE BOX WITH THE CORAIL STEM WITH COLLAR, DID NOT CONTAIN A STEM WITH COLLAR. THE WRONG IMPLANT WAS IN THE BOX. THIS REPORT INVOLVES ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19013178 ↗MalfunctionJan 23, 2023Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THE ANCHOR BROKE. IT WAS REPORTED THAT DURING THE SURGERY ,THE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER ONE WAS USED TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS COMPLAINT INVOLVES TWO (2) DEVICES. 2.0MM IMF SCREW SELF-DRILLING 8MM-5PK. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16226140 ↗MalfunctionJan 23, 2023Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THE ANCHOR BROKE. IT WAS REPORTED THAT DURING THE SURGERY ,THE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER ONE WAS USED TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS COMPLAINT INVOLVES TWO (2) DEVICES. 2.0MM IMF SCREW SELF-DRILLING 8MM-5PK. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16226055 ↗InjuryJun 13, 2019Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, IN THE PAST 4 (FOUR) WEEKS, THE SURGEON HAD (TEN) 10 DIFFERENT ITEMS SCREWS INCLUDING THE SCREW SELF-DRILLING 8MM BROKE. IT IS UNKNOWN HOW MANY SCREWS WERE BROKEN. FRAGMENTS WERE GENERATED FROM BROKEN DEVICES HAD TO DRILL OUT. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. PROCEDURE WAS SUCCESSFULLY COMPLETED JUST HAD TO PUT IN ANOTHER SCREW. PATIENT OUTCOME HAD TO LEAVE SOME OF THE BROKEN SCREW INSIDE THE TISSUE. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS 1 OF 2 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 8697988 ↗MalfunctionJun 7, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING DENTAL SURGERY AND INTRAMAXILLARY FIXATION, THREE (3) INTRAMAXILLARY FIXATION SCREWS BROKE. THE PATIENT HAD A MANDIBLE FRACTURE. THE SCREWS WERE REMOVED. THERE WAS A SURGICAL DELAY OF TWENTY (20) MINUTES. PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: UNKNOWN PRECUT CERCLAGE WIRES (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY: UNKNOWN). THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8679687 ↗MalfunctionJun 7, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING DENTAL SURGERY AND INTRAMAXILLARY FIXATION, THREE (3) INTRAMAXILLARY FIXATION SCREWS BROKE. THE PATIENT HAD A MANDIBLE FRACTURE. THE SCREWS WERE REMOVED. THERE WAS A SURGICAL DELAY OF TWENTY (20) MINUTES. PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: UNKNOWN PRECUT CERCLAGE WIRES (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY: UNKNOWN). THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8679647 ↗MalfunctionJun 7, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING DENTAL SURGERY AND INTRAMAXILLARY FIXATION, THREE (3) INTRAMAXILLARY FIXATION SCREWS BROKE. THE PATIENT HAD A MANDIBLE FRACTURE. THE SCREWS WERE REMOVED. THERE WAS A SURGICAL DELAY OF TWENTY (20) MINUTES. PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: UNKNOWN PRECUT CERCLAGE WIRES (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY: UNKNOWN). THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8679615 ↗MalfunctionFeb 28, 2019Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE. DURING THE PROCEDURE, THE HOLDING SLEEVE OF THE INTERMAXILLARY FIXATION (IMF) SCREW DID NOT HOLD THE HEAD OF THE SCREWS WELL; WHEN THE SCREWS WERE PLACED THEY MOVE A LOT, MAKING PLACEMENT DIFFICULT. IT WAS NOTED A LOT OF FORCE WAS PLACED ON TWO (2) SELF-DRILLING INTERMAXILLARY FIXATION (IMF) SCREWS TO PLACE THEM. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME AND PATIENT STATUS WERE UNKNOWN. THIS REPORT IS FOR AN INTERMAXILLARY FIXATION (IMF) SCREW. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8377909 ↗
Other SCREW, FIXATION, INTRAOSSEOUS devices
- MatrixMIDFACE 04.503.226.01Synthes GmbH · SCREW, FIXATION, INTRAOSSEOUS21 reports
- NA 400.736ESynthes GmbH · SCREW, FIXATION, INTRAOSSEOUS21 reports
- OsteoMed 204-1206Osteomed LLC · SCREW, FIXATION, INTRAOSSEOUS14 reports
- LEVEL ONE CMF 25-092-12-91KLS-Martin L.P. · SCREW, FIXATION, INTRAOSSEOUS11 reports
- OsteoMed 202-2016Osteomed LLC · SCREW, FIXATION, INTRAOSSEOUS9 reports
- LEVEL ONE CMF 25-092-08-91KLS-Martin L.P. · SCREW, FIXATION, INTRAOSSEOUS9 reports
- MatrixMIDFACE 04.503.204.01Synthes GmbH · SCREW, FIXATION, INTRAOSSEOUS8 reports
- PlusDrive 401.062ESynthes GmbH · SCREW, FIXATION, INTRAOSSEOUS8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page