Synthes GmbH · 2.0MM RAPID RESORBABLE CORTEX SCREW 4MM-STERILE · Class 2 · Implant

All models of RapidSorb (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes RapidSorb screw is a dissolving screw used to fix bone in place. Most reports describe the screw breaking, often during insertion in surgery. The 157 reports include no deaths and no injuries. No recalls name this device, and reports rose from 54 in the prior 12 months to 91 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
146
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 146 adverse event reports that list this exact device (DI 20887587061815), received between Jan 27, 2022 and Sep 15, 2026 — classified as 146 malfunctions.

The most frequently reported specific device problem is “Break” (143 reports).

3 other recalls cover devices cleared under the same FDA submission (K030069, K062789); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 6 reportsMay 2025: 10 reportsJun 2025: 3 reportsJul 2025: 13 reportsAug 2025: 5 reportsSep 2025: 8 reportsOct 2025: 8 reportsNov 2025: 4 reportsDec 2025: 37 reportsJan 2026: 6 reportsFeb 2026: 9 reportsMar 2026: 3 reportsApr 2026: 4 reportsMay 2026: 1 reportsJun 2026: 7 reportsJul 2026: 0 reportsAug 2026: 8 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break143
  • No Apparent Adverse Event3
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions145
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 146 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775827 ↗
  • MalfunctionAug 31, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50522837 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495440 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495438 ↗
  • MalfunctionAug 28, 2026Break

    It was reported during the surgery, when inserting the screw, the screw was broke. All the pieces were removed from the patient. Changed to another eight to continue the surgery, the same problem happened again. Total involved nine screws. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information could be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495432 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495429 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495428 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495426 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495388 ↗
  • MalfunctionJun 23, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25593945 ↗
  • MalfunctionJun 23, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25593912 ↗
  • MalfunctionJun 16, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25522470 ↗
  • MalfunctionJun 16, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25522467 ↗
  • MalfunctionJun 3, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25380991 ↗
  • MalfunctionJun 3, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERTING THE SCREW, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25379029 ↗
  • MalfunctionJun 3, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25378043 ↗
  • MalfunctionMay 28, 2026Break

    IT WAS REPORTED THAT DURING A NEUROSURGERY SKULL CLOSURE SURGERY, TWO SCREWS WERE BROKEN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25317039 ↗
  • MalfunctionApr 27, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24999489 ↗
  • MalfunctionApr 10, 2026Break

    ADDITIONAL INFORMATION RECEIVED: A. WAS THERE A SURGICAL DELAY IN RELATION TO THE REPORTED EVENT? - NO. B. WAS OTHER MEDICAL INTERVENTION REQUIRED? - NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24844898 ↗
  • MalfunctionApr 3, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24779860 ↗
  • MalfunctionApr 3, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24779842 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW BROKE. ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER SCREW TO CONTINUE THE SURGERY. THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24745066 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW BROKE. ALL THE PIECES WERE REMOVED FROM THE PATIENT. CHANGED ANOTHER SCREW TO CONTINUE THE SURGERY. THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24745064 ↗
  • MalfunctionMar 20, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, NOTED THE SCREW THREADS PEELED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649529 ↗
  • MalfunctionFeb 10, 2026Break

    IT WAS REPORTED THAT DURING THE SURGERY, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN, ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24303054 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K030069, K062789. One submission often covers many models, so check the product and lot details in each record.

  • Z-1920-2016 · May 2, 2016 · Terminated
    1.5 mm Rapid Resorb Orbt1 F1 P1 w/Bending Template / 35 mm - Sterile, Rapid Resorbable Fixation System. 1.5 mm Rapid Resorbable Orbital Floor Plate with Bending Template. Intend
  • Z-1918-2016 · May 2, 2016 · Terminated
    1.5 mm Rapid Resorb Orbt1 F1 P1 w/Bending Template / 24 mm - Sterile, Rapid Resorbable Fixation System. 1.5 mm Rapid Resorbable Orbital Floor Plate with Bending Template. Inte
  • Z-1919-2016 · May 2, 2016 · Terminated
    1.5 mm Rapid Resorb Orbt1 F1 P1 w/Bending Template / 30 mm - Sterile, Rapid Resorbable Fixation System. 1.5 mm Rapid Resorbable Orbital Floor Plate with Bending Template. Intend

Other PLATE, BONE devices

All PLATE, BONE devices →