SMARTeZ Pump 481062Recalls & reported problems
Epic International(Thailand) Company Limited · SMARTeZ Pump 250-175-1.5h · Class 2
All models of SMARTeZ Pump (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 28859299101068), received between Nov 15, 2018 and Feb 19, 2019 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Failure to Infuse” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Infuse5
- Leak/Splash4
- Infusion or Flow Problem2
- Failure to Deliver1
- Fluid/Blood Leak1
- Improper Flow or Infusion1
- Insufficient Flow or Under Infusion1
Patient problems
- No Known Impact Or Consequence To Patient12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2019Improper Flow or Infusion
SMARTEZ PUMP (SE0175-250, LOT# S8H02) CONTAINING VANCOMYCIN 1 GRAM IN 0.9% SODIUM CHLORIDE 250 ML IS TAKING 2-3 HOURS TO INFUSE. LINE IS FLUSHING GOOD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350446 ↗MalfunctionFeb 19, 2019Fluid/Blood Leak; Failure to Infuse
SMARTEZ PUMPS (SE0175-250 , LOT# S8H02 X3) CONTAINING VANCOMYCIN 1.5 GRAM IN 0.9% SODIUM CHLORIDE 250 ML; 1 WOULD NOT INFUSE, 1 WAS TAKING OVER 5 HOURS TO INFUSE (BOTH DRIPPED WHEN DISCONNECTED AND UNCLAMPED), AND 1 LEAKED OUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350437 ↗MalfunctionFeb 19, 2019Leak/Splash; Failure to Infuse
SMARTEZ PUMPS (SE0175-250, LOT# S8H02 X3) CONTAINING VANCOMYCIN 1.5 GRAM IN 0.9% SODIUM CHLORIDE 250 ML; 1 WOULD NOT INFUSE, 1 WAS TAKING OVER 5 HOURS TO INFUSE (BOTH DRIPPED WHEN DISCONNECTED AND UNCLAMPED), AND 1 LEAKED OUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350436 ↗InjuryFeb 19, 2019Insufficient Flow or Under Infusion
SLOW FLOW WITH VANCOMCYIN 2 G/NS 250 ML SMARTEZ {SEOL75-250; LOT: S8H02} AND CAFTRIAXONE 2 G/NS 100ML SMARTEZ {SE0200-100; LOT S8D33}. {SIX (6) DOSES OF VANCOMYCIN AND FOUR (4) DOSES OF CEFTRIAXONE HAD TO BE REPLACED); 100 ML, 200 ML /HR AND SMARTEZ 250 ML. 175 ML/HR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350429 ↗MalfunctionFeb 19, 2019Infusion or Flow Problem
SLOW FLOW WITH VANCOMYCIN 2 G/NS 250 ML SMARTEZ {SE0175-250; LOT· S8H02} AND CEFTRIAXONE 2 G/NS 100ML SMARTEZ {SE0200-100; LOT S8D33): 100 ML, 200ML /HR AND SMARTEZ 250 ML, 175 ML/HR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350394 ↗MalfunctionFeb 19, 2019Leak/Splash; Failure to Infuse
SMARTEZ PUMPS (SE0175-250 , LOT# S8H02 X3) CONTAINING VANCOMYCIN 1.5 GRAM IN 0.9% SODIUM CHLORIDE 250 ML; 1 WOULD NOT INFUSE, 1 WAS TAKING OVER 5 HOURS TO INFUSE (BOTH DRIPPED WHEN DISCONNECTED AND UNCLAMPED), AND 1 LEAKED OUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350393 ↗MalfunctionFeb 19, 2019Failure to Infuse
SMARTEZ PUMP ((B)(4), LOT# S8H02) TELAVANCIN 750 MG IN 0.9% SODIUM CHLORIDE 250ML TOOK 4 HOURS TO INFUSE. PATIENT LINE FLUSHING WITHOUT PROBLEMS. ED DRIPS QUICKLY WHEN DISCONNECTED AND UNCLAMPED. RECONNECTED ED TIGHTLY WITHOUT IMPROVEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8350386 ↗MalfunctionJan 23, 2019Leak/Splash
PHARMACY TECHNICIAN ATTEMPTED TOFILL A SMARTEZ PUMP (SE0L75-250, LOT# S8H02) WITH 0.9% SODIUM CHLORIDE WHEN FLUID BEGAN SPRAYING OUT AT THE FILL PORT SEAL AREA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8272152 ↗MalfunctionDec 10, 2018Failure to Infuse
SMARTEZ PUMP (SE0175-250. LOT# S8H02) CONTAINING TELEVANCIN 1300 MG IN 0.9% SODIUM CHLORIDE 250 ML WOULD NOT INFUSE. DOSE REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8147453 ↗MalfunctionDec 10, 2018Infusion or Flow Problem
SMARTEZ PUMP (MODEL SE0175-250, LOT# S8H02) CONTAINING VANCOMYCIN 1500 MG IN 0.9% SODIUM CHLORIDE 250 ML WOULD NOT INFUSE. THE PUMP SEEMS TO WORK FINE WHEN UNCLAMPED AND NOT ATTACHED TO LINE. PRS LINE WORKING FINE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8147449 ↗MalfunctionDec 10, 2018Leak/Splash
TECH USING REPEATER PUMP TO FILL SMARTEZ PUMP ((B)(4), LOT# S8H02) WITH 0.9% SODIUM CHLORIDE AND FLUID BEGAN LEAKING OUT AT BASE OF ELASTOMERIC DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8147444 ↗MalfunctionNov 15, 2018Failure to Deliver
TWO (2) SMARTEZ PUMPS (250 ML VOLUME; 175 MUHR) FILLED WITH VANCOMYCIN 1 G/0.9% SODIUM CHLORIDE WOULD NOT INFUSE. SMARTEZ PUMP LOT: S8H11. THERAPY START DATE: (B)(6) 2018. NO PATIENT INVOLVEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8074954 ↗
Other Pump, Infusion, Elastomeric devices
- EASYPUMP 4540018-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC313 reports
- Halyard 13568Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC86 reports
- ON-Q* Pain Relief System with SELECT-A-FLOW 13472Avanos Medical, Inc. · Pump, infusion, elastomeric73 reports
- Halyard 13472Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC68 reports
- EASYPUMP 4540008-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC54 reports
- DOSI-FUSER L25915-500D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- DOSI-FUSER L25915-600D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- Halyard 13561Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page