Solana SARS-CoV-2 Assay CoronavirusRecalls & reported problems

Diagnostic Hybrids, Inc. · Solana SARS-CoV-2 Assay · Class N

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Diagnostic Hybrids Solana SARS-CoV-2 test is a reagent used to detect the coronavirus. Its 328 reports describe false positive results, false negative results, and erratic readings. The reports include no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
328
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 328 adverse event reports that list this exact device (DI 30014613339656), received between Apr 20, 2021 and Aug 15, 2022 — classified as 328 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (244 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Apparent Adverse Event322
  • False Positive Result244
  • False Negative Result43
  • Imprecise or Erratic Results32
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions325
  • Insufficient Information7

Recent adverse event reports

The 25 most recent of 328 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 15, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event

    REPORTED FALSE NEGATIVE SARS RESULT. THE CUSTOMER COMMUNICATED THE RESULT WAS POSITIVE BY ANOTHER MOLECULAR METHOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15227085 ↗
  • MalfunctionAug 15, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event

    REPORTED FALSE NEGATIVE SARS RESULT. THE CUSTOMER COMMUNICATED THE RESULT WAS POSITIVE BY ANOTHER MOLECULAR METHOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15227016 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090069 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090067 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090066 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090065 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090064 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090063 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090062 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090061 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090060 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090058 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090057 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090028 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090026 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090024 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090022 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090021 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090020 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090010 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090009 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090007 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090006 ↗
  • MalfunctionJul 22, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT. CUSTOMER COMMUNICATED THE ISSUE IS OCCURRING OVER SEVERAL INSTRUMENTS AND WITH NEGATIVE CONTROLS (EACH PATIENT EVENT WILL BE FILED SEPARATELY). SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN AND NO CONFIRMATION TESTING WAS PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15090001 ↗
  • MalfunctionJun 30, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event

    REPORTED FALSE POSITIVE SARS RESULT FOR 1 PATIENT. IT IS UNKNOWN IF THE PATIENT WAS SYMPTOMATIC. THE CUSTOMER COMMUNICATED THE PATIENT LATER TESTED NEGATIVE WITH MOLECULAR (PCR TESTING) METHOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14876136 ↗

Other Reagents, 2019-novel coronavirus nucleic acid devices

All Reagents, 2019-novel coronavirus nucleic acid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page