Skip to content

BD BBL™ Gram Decolorizer 212527Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Gram Decolorizer, 250 mL · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30382902125274), received between Apr 13, 2022 and Nov 15, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Ingredient or Reagent Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Ingredient or Reagent Problem1
  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED PRIOR TO USING THE BD BBL¿ GRAM DECOLORIZER THE USER NOTED A BOTTLE DID NOT HAVE A LABEL AND COULD NOT BE USED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20695402 ↗
  • MalfunctionApr 13, 2022Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED WHILE USING BD BBL¿ GRAM DECOLORIZER THE SLIDES WERE OVERSTRIPPED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTED USING GRAM STAIN KIT FOR CONTAMINATION CHECKS OF INCOMING MICRO SAMPLES AS PART OF OUR STANDARD PROCESS. DURING THESE CHECKS I HAVE RUN INTO A CONSISTENT ISSUE WITH THE GRAM DECOLORIZER PROVIDED WITH YOUR KIT. MY SLIDES APPEAR TO ALWAYS BE OVER STRIPPED OF CRYSTAL VIOLET. MY PROCEDURE IS AS FOLLOWS FLOOD WITH CRYSTAL VIOLET, ALLOW TO STAND 1 MINUTE, RINSE THOROUGHLY WITH DI WATER, FLOOD WITH GRAM IODINE, ALLOW TO STAND 1 MINUTE, RINSE THOROUGHLY WITH DI WATER, APPLY 1-2 DROPS OF GRAM DECOLORIZER, IMMEDIATELY RINSE THOROUGHLY WITH DI WATER, FLOOD SLIDE WITH SAFRANIN, ALLOW TO STAND 30 SECONDS, RINSE THOROUGHLY WITH DI WATER. DRY IN BOOK OF BIBULOUS PAPER BEFORE MICROSCOPICALLY OBSERVING RESULTS. I WAS WONDERING IF THERE WERE ANY SUGGESTIONS TO PREVENT THE OVER STRIPPING OF THE CRYSTAL VIOLET WITH THE DECOLORIZER? THIS HAS BEEN A CONSISTENT PROBLEM EVEN WITH CONTROL MICROBES SUCH AS B. AGRI AND B. SUBTILIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14108002 ↗

Other Stains, microbiologic devices

All Stains, microbiologic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page