Skip to content

BD BBL™ Fluid Thioglycollate Medium 221196Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Fluid Thioglycollate Medium, 8 mL, Carton of 100 size K tubes · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 30382902211960), received between Jul 15, 2022 and Jan 2, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 24, 2025Class IIOpen, Classified

    transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.

    Z-1828-2025 · Root cause: Storage · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Contamination /Decontamination Problem1
  • Contamination of Device Ingredient or Reagent1
  • Particulates1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHEN USING THE BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM HAD NOTICED A GROWTH IN THE TUBE AFTER INOCULATION. THERE WAS NO REPORT OF IMPACT TO PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18428520 ↗
  • MalfunctionMar 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT WHILE USING THE BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM THAT THERE WAS CONTAMINATION GROWTH ON THE PLATE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS THAT THEY HAD GROWTH IN A NEGATIVE CONTROL IN 221196.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16575530 ↗
  • MalfunctionMar 8, 2023Particulates; Contamination /Decontamination Problem

    IT WAS REPORTED THAT WHILE USING THE BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS: AS PART OF OUR ACCEPTANCE PROCEDURES WE INCUBATE SEVERAL TUBES IMMEDIATELY ON RECEPTION AS A NEGATIVE CONTROL BEFORE GROWTH PROMOTION IS COMPLETED. ONE OF THE FTM TUBES THAT WE INCUBATED CAME BACK POSITIVE DESPITE BEING A NEGATIVE CONTROL AND WE FOUND PARTICULATES FLOATING IN THE MEDIA OF 2 OTHER TUBES. LOT 2312524.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16504519 ↗
  • MalfunctionJul 15, 2022Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    IT WAS REPORTED WHILE USING THE BD BBL¿ FLUID THIOGLYCOLLATE MEDIUM THAT THERE WAS CONTAMINATED TUBED MEDIA. THE CUSTOMER REPORTED THAT THIS OCCURRED 400 TIMES. SUBJECT LINE SHOULD READ: MEDIA 221196 CONTAMINATED TUBED MEDIA CUSTOMER REPORTS CONTAMINATED TUBED MEDIA- LOT #1349733.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15029229 ↗

FDA clearance and classification

The FDA regulates product code JSG (culture media, non-selective and non-differential) under 21 CFR 866.2300 as a Class I device. 21 CFR 866.2300 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the BD BBL™ Fluid Thioglycollate Medium 221196

Has the BD BBL™ Fluid Thioglycollate Medium 221196 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1828-2025 (Class II), began on Apr 24, 2025: transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the BD BBL™ Fluid Thioglycollate Medium 221196?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 15, 2022 and Jan 2, 2024. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (2), “Contamination /Decontamination Problem” (1) and “Contamination of Device Ingredient or Reagent” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD BBL™ Fluid Thioglycollate Medium 221196?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BD BBL™ Fluid Thioglycollate Medium 221196 FDA approved?

Devices of this type (product code JSG) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Becton, Dickinson And Company.

Who makes the BD BBL™ Fluid Thioglycollate Medium 221196, and is it still sold?

The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD BBL™ Fluid Thioglycollate Medium, model 221196). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL devices

All CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page