BD GasPak™ EZ Anaerobe Container System Sachets with Indicator 260001Recalls & reported problems
Becton, Dickinson And Company · BD GasPak™ EZ Anaerobe Container System Sachets with Indicator · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 30382902600016), received between Jun 18, 2021 and Aug 29, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Contamination of Device Ingredient or Reagent” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Contamination of Device Ingredient or Reagent2
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 29, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD GASPAK¿ EZ ANAEROBE CONTAINER SYSTEM WITH INDICATOR, 20 SACHETS DON'T PRODUCE AN ANEROBIC CONDITION. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE SACHETS DON'T PRODUCE ANAEROBIC CONDITION. THE CUSTOMER PUT ONE SACHET WITH INDICATOR INTO A JAR, AFTER A NIGHT THEY WAS NOTICE THE INDICATOR STAY BLUE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17647305 ↗MalfunctionJun 2, 2022Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent
IT WAS REPORTED THAT BD GASPAK¿ EZ ANAEROBE CONTAINER SYSTEM WITH INDICATOR WERE CONTAMINATED. THERE WAS NO REPORT OF CONFIRMATORY TESTING. NO PATIENT IMPACT WAS REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14580232 ↗MalfunctionJun 18, 2021Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent
IT WAS REPORTED THAT WHILE USING 6 BD GASPAK¿ EZ ANAEROBE CONTAINER SYSTEM WITH INDICATOR MOLD CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IN TERMS OF THE SACHETS THEY APPEAR TO BE CONTAMINATED PRIOR TO BEING USED. THEY LOOK MOULDY TO ME FROM THE IMAGES & THE CUSTOMER SAID THEY APPEARED TO BE CONTAMINATED PRIOR TO USE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12025877 ↗
Other DEVICE, GAS GENERATING devices
- BD GasPak™ EZ CO2 Gas Generating Pouch System 260684Becton, Dickinson And Company · DEVICE, GAS GENERATING1 recall4 reports
- BD GasPak™ EZ Campy Gas Generating Pouch System 260685Becton, Dickinson And Company · DEVICE, GAS GENERATING1 recall0 reports
- BD GasPak™ EZ Anaerobe Gas Generating Pouch System with Indicator 260683Becton, Dickinson And Company · DEVICE, GAS GENERATING4 reports
- BD GasPak™ EZ Anaerobe Container System Sachets 260678Becton, Dickinson And Company · DEVICE, GAS GENERATING2 reports
- BD BBL™ Dry Anaerobic Indicator Strips 271051Becton, Dickinson And Company · DEVICE, GAS GENERATING1 report
- BD BBL™ GasPak™ 100 Anaerobic System 260626Becton, Dickinson And Company · DEVICE, GAS GENERATING1 report
- BD BBL™ CO2 Indicators 271055Becton, Dickinson And Company · DEVICE, GAS GENERATING1 report
- BD GasPak™ EZ Campy Container System Sachets 260680Becton, Dickinson And Company · DEVICE, GAS GENERATING1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page