BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761Recalls & reported problems
Becton, Dickinson And Company · TUBE CPTCIT GC 16X125 8.0 MLBL BL/BLK · Class 1
All models of BD Vacutainer CPT Cell Preparation Tube with Sodium Citrate (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This BD Vacutainer CPT is a blood collection tube used to separate blood cells for lab tests. Many of its 286 reports describe underfilled tubes, poor separation of the blood layers and inaccurate test results. The reports include no deaths and no injuries. One open recall began in May 2022 and concerns tubes exposed to higher temperatures before delivery, which may have reduced their effectiveness. Reports fell from 27 in the previous 12 months to 7 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 286
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 286 adverse event reports that list this exact device (DI 30382903627616), received between Oct 17, 2017 and Jul 27, 2026 — classified as 286 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (75 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 25, 2022Class IIOpen, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Z-1618-2022 · Root cause: Environmental control · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings75
- Incorrect Or Inadequate Test Results61
- Short Fill43
- Break39
- Device Contamination with Chemical or Other Material22
- Device Ingredient or Reagent Problem21
- Volume Accuracy Problem19
- Leak/Splash14
Patient problems
- No Known Impact Or Consequence To Patient125
- No Clinical Signs, Symptoms or Conditions120
- No Consequences Or Impact To Patient25
- No Information13
- No Patient Involvement3
- Exposure to Body Fluids1
Recent adverse event reports
The 25 most recent of 286 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2026Short Fill; Improper Chemical Reaction
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50020589 ↗MalfunctionApr 22, 2026Short Fill
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, ONE TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954300 ↗MalfunctionFeb 2, 2026Device Ingredient or Reagent Problem
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED POOR BARRIER SEPARATION AFTER PROCESSING. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24239969 ↗MalfunctionJan 30, 2026Short Fill
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, ONE (1) OF TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24228424 ↗MalfunctionDec 17, 2025Short Fill; Short Fill
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF TUBES HAD VACUUM LOSS RESULTING IN UNDERFILLED TUBES. THE TUBES WERE REPLACED. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23835405 ↗MalfunctionDec 3, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED POOR BARRIER SEPARATION (BLOOD IS NOT PASSING THROUGH FICOLL LAYER). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23711545 ↗MalfunctionNov 20, 2025Improper Chemical Reaction; Improper Chemical Reaction
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE AN UNSPECIFIED NUMBER OF TUBES EXHIBITED POOR BARRIER SEPARATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609558 ↗MalfunctionOct 22, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED PRIOR TO USE OF 21 CASES OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, THE OUTER CASE HAD AN INCORRECT EXPIRATION DATE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23357688 ↗MalfunctionOct 16, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
IT WAS REPORTED THAT WHEN USING THE BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED POOR BARRIER SEPARATION. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23311101 ↗MalfunctionJul 9, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
IT WAS REPORTED DURING USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 5 SAMPLES EXHIBITED POOR BARRIER SEPARATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22462505 ↗MalfunctionJun 23, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
IT WAS REPORTED DURING USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 3 SAMPLES EXHIBITED POOR BARRIER SEPARATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22311096 ↗MalfunctionJun 20, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 3 SAMPLES EXHIBITED POOR BARRIER SEPARATION AFTER PROCESSING. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22297370 ↗MalfunctionJun 9, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, EXCESS RED BLOOD CELLS IN GEL WAS OBSERVED IN ONE MATERNAL SAMPLE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22171447 ↗MalfunctionJun 9, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
REPORT 1 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, POOR BARRIER SEPARATION WAS SEEN IN ONE FETAL SAMPLE. THE SAMPLE WAS CENTRIFUGED FURTHER FOR AN ADDITIONAL 10 MINUTES (PER INSTRUCTION FOR USE) WITH THE SAME OUTCOME (POOR BARRIER SEPARATION). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22171429 ↗MalfunctionApr 30, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE AN UNSPECIFIED NUMBER OF SAMPLES EXHIBITED POOR BARRIER SEPARATION AND THE GEL CONSISTENCY WAS NOT ALWAYS FIRM (COULD BE GOOEY OR STRINGY). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21936959 ↗MalfunctionApr 25, 2025Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF SAMPLES EXHIBITED CLUMPING DURING THE CELL WASHING/RESUSPENSION PHASE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21908418 ↗MalfunctionApr 23, 2025Short Fill; Short Fill
IT WAS REPORTED DURING USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 10 TUBES WERE UNDERFILLED. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21893135 ↗MalfunctionApr 3, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
REPORT 1 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE IN A CLINICAL TRIAL, 1 PATIENT SAMPLE EXHIBITED POOR BARRIER SEPARATION AFTER PROCESSING. PATIENT WAS RECOLLECTED AND AGAIN EXHIBITED POOR BARRIER SEPARATION. STUDY PARTICIPANT WAS REMOVED FROM TRIAL. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21766803 ↗MalfunctionApr 3, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE IN A CLINICAL TRIAL, 1 PATIENT SAMPLE EXHIBITED POOR BARRIER SEPARATION AFTER PROCESSING. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21766227 ↗MalfunctionApr 2, 2025Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE,1 SAMPLE EXHIBITED CLUMPING DURING THE CELL WASHING/RESUSPENSION PHASE POST-CENTRIFUGATION. CELLS ARE NOT CLUMPED INITIALLY UPON HARVESTING; CLUMPING APPEARS DURING THE WASH. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21753739 ↗MalfunctionMar 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 1 SAMPLE CONTAINED OIL GEL GLOBULES AFTER CENTRIFUGATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21723970 ↗MalfunctionMar 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 1 SAMPLE CONTAINED OIL GEL GLOBULES AFTER CENTRIFUGATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21708478 ↗MalfunctionMar 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 4 SAMPLES EXHIBITED POOR BARRIER SEPARATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21644423 ↗MalfunctionMar 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
REPORT 2 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE ,1 PATIENT SAMPLE EXHIBITED POOR BARRIER SEPARATION. 2 SAMPLES WERE COLLECTED ON THE PATIENT DURING THE SAME COLLECTION, AND AFTER PROCESSING ONLY ONE SAMPLE EXHIBITED POOR BARRIER SEPARATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21595029 ↗MalfunctionMar 12, 2025Break; Break
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, 1 TUBE BROKE IN THE CENTRIFUGE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21588807 ↗
FDA clearance and classification
The FDA regulates product code JCF (lymphocyte separation medium) under 21 CFR 864.8500 as a Class I device. 21 CFR 864.8500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761
Has the BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1618-2022 (Class II), began on May 25, 2022: Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761?
The FDA's MAUDE database lists 286 adverse event reports for this device, received between Oct 17, 2017 and Jul 27, 2026. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (75), “Incorrect Or Inadequate Test Results” (61) and “Short Fill” (43); the most common patient problem is “Exposure to Body Fluids” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761?
None of the reports are classified as deaths, 0 as injuries and 286 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761 FDA approved?
Devices of this type (product code JCF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Becton, Dickinson And Company.
Who makes the BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362761, and is it still sold?
The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate, model 362761). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other LYMPHOCYTE SEPARATION MEDIUM devices
- BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Heparin 362753Becton, Dickinson And Company · LYMPHOCYTE SEPARATION MEDIUM271 reports
- BD Vacutainer® CPT™ Cell Preparation Tube with Sodium Citrate 362760Becton, Dickinson And Company · LYMPHOCYTE SEPARATION MEDIUM129 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page