BD Vacutainer® SST™ II Advance 367955Recalls & reported problems

Becton, Dickinson U.K. Limited · BD Vacutainer® SST™ II Advance · Class 2

All models of BD Vacutainer SST II Advance (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Vacutainer SST II Advance blood collection tube has 1,459 FDA adverse event reports, with 1,456 malfunctions, 3 injuries and no deaths. The reports mostly describe wrong or unreliable test results, tubes not filled to the right amount, and leaks or splashes. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,478
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,478 adverse event reports that list this exact device (DI 30382903679554), received between Nov 27, 2017 and Sep 30, 2026 — classified as 3 injuries, 1,475 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (318 reports).

3 other recalls cover devices cleared under the same FDA submission (BK050036); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 10 reportsJan 2024: 13 reportsFeb 2024: 8 reportsMar 2024: 9 reportsApr 2024: 16 reportsMay 2024: 8 reportsJun 2024: 15 reportsJul 2024: 7 reportsAug 2024: 2 reportsSep 2024: 38 reportsOct 2024: 14 reportsNov 2024: 13 reportsDec 2024: 7 reportsJan 2025: 29 reportsFeb 2025: 17 reportsMar 2025: 27 reportsApr 2025: 17 reportsMay 2025: 33 reportsJun 2025: 25 reportsJul 2025: 26 reportsAug 2025: 13 reportsSep 2025: 30 reportsOct 2025: 23 reportsNov 2025: 31 reportsDec 2025: 42 reportsJan 2026: 45 reportsFeb 2026: 23 reportsMar 2026: 11 reportsApr 2026: 8 reportsMay 2026: 25 reportsJun 2026: 0 reportsJul 2026: 5 reportsAug 2026: 0 reportsSep 2026: 23 reportsOct 2026: 0 reports20242025202645
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings318
  • Short Fill289
  • Device Ingredient or Reagent Problem287
  • Leak/Splash124
  • Device Contamination with Chemical or Other Material119
  • Incorrect Or Inadequate Test Results83
  • Material Deformation78
  • Volume Accuracy Problem61

Patient problems

  • No Clinical Signs, Symptoms or Conditions983
  • No Known Impact Or Consequence To Patient372
  • No Consequences Or Impact To Patient67
  • No Information36
  • No Patient Involvement17
  • Insufficient Information2
  • Blood Loss1
  • Hemolysis1

Recent adverse event reports

The 25 most recent of 1,478 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954889 ↗
  • MalfunctionSep 29, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935964 ↗
  • MalfunctionSep 29, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935962 ↗
  • MalfunctionSep 24, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890862 ↗
  • MalfunctionSep 24, 2026Device Contamination with Chemical or Other Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889794 ↗
  • MalfunctionSep 24, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889054 ↗
  • MalfunctionSep 23, 2026Short Fill

    It was reported while using "[brand name]" there was poor barrier separation seen in (x) tube(s). No adverse impact was reported regarding the patient or user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870275 ↗
  • MalfunctionSep 23, 2026Component Missing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863111 ↗
  • MalfunctionSep 21, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846040 ↗
  • MalfunctionSep 21, 2026Short Fill

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841669 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820963 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient 7 of 13:It was reported that when using the BD Vacutainer® SST¿ II Advance Plus Blood Collection Tubes, elevated Na and K parameters were observed. Upon repeat of the same sample the results were normal. No adverse impact was reported regarding the patient or user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820962 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820961 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820959 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820958 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820957 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820956 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820955 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient 11 of 13:It was reported that when using the BD Vacutainer® SST¿ II Advance Plus Blood Collection Tubes, elevated Na and K parameters were observed. Upon repeat of the same sample the results were normal. No adverse impact was reported regarding the patient or user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820953 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820951 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient 2 of 13:It was reported that when using the BD Vacutainer® SST¿ II Advance Plus Blood Collection Tubes, elevated Na and K parameters were observed. Upon repeat of the same sample the results were normal. No adverse impact was reported regarding the patient or user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820950 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820949 ↗
  • MalfunctionSep 15, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775084 ↗
  • MalfunctionJul 27, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50021570 ↗
  • MalfunctionJul 27, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50021544 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under BK050036. One submission often covers many models, so check the product and lot details in each record.

  • Z-1618-2022 · May 25, 2022 · Open, Classified
    a. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/SEP GLD 3.5ML PLAS VACUTAINER (100/BX), Model Number: 367983. b. BECTON DICKINSON BD Vacutainer SST TUBE, BLD COL CLOT/
  • Z-0968-2018 · Aug 29, 2017 · Terminated
    BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing an
  • Z-2434-2015 · Jun 23, 2015 · Terminated
    BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transporting, separating, and processing blood in a closed tube.

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page