FRAZIER 0033100 (UDI 30653405000123)Recalls & reported problems

Conmed Corporation · FRAZIER Suction Instrument 10 Fr W/Control Vent & Obturator · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 30653405000123), received between Dec 13, 2021 and Sep 12, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product10

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 12, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE 0033100 SURGICAL SUCTION INSTRUMENT, 10FR W/CONTROL VENT & OBTURATOR, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. PERFORMED A FUNCTIONAL INSPECTION, THE DEVICES WERE DYE LEAK TESTED WHICH INDICATED THAT THE PACKAGING HAD AN INSUFFICIENT HEAT SEAL. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23037024 ↗
  • MalfunctionApr 25, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21908618 ↗
  • MalfunctionDec 5, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, SURGICAL SUCTION INSTRUMENT, 10FR W/CONTROL VENT & OBTURATOR, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20852342 ↗
  • MalfunctionSep 6, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20155640 ↗
  • MalfunctionJan 19, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED TWO DEVICES, 0033100, FRAZIER INSTRUMENT 10 FRENCH 50/CASE, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18545095 ↗
  • MalfunctionOct 20, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. PERFORMED A FUNCTIONAL INSPECTION, THE DEVICES WERE DYE LEAK TESTED WHICH INDICATED THAT THE PACKAGING HAD AN INSUFFICIENT HEAT SEAL. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17974506 ↗
  • MalfunctionApr 26, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16818067 ↗
  • MalfunctionJan 26, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16247456 ↗
  • MalfunctionNov 18, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0033100, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15824249 ↗
  • MalfunctionDec 13, 2021Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, CATALOG # 0033100, LOT 202006221, FOR A POSSIBLE BREACH OF STERILITY DUE TO AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH ANY PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION IN JAPAN. THE COMPLETION OF THE DEVICE EVALUATION CONFIRMED AN INSUFFICIENT HEATSEAL; THEREFORE, THIS COMPLAINT MEETS THE CRITERIA FOR A REPORTABLE EVENT. DUE TO THE BREACH IN STERILITY, THIS REPORT IS BEING RAISED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12987744 ↗

Other APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices

All APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices →