3M™ 9690 (UDI 30707387789290)Recalls & reported problems
3m Company · 3M™ Single-use Specialty Clipper Blade Assembly, 9690, 20 Each/Case · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 12, 2024Open, Classified
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41C (105.8F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60C (140F) and the Model 9660 blade maximum temperature may reach 77C (170.6F)."
Z-1170-2024 · Root cause: Under Investigation by firm · FDA record ↗
Other RAZOR, SURGICAL devices
- 3M™ 96603m Company · RAZOR, SURGICAL1 recall0 reports
- N/A 4406Carefusion 2200, Inc · RAZOR, SURGICAL17 reports
- MEDLINE CLIP5515Medline Industries, Inc. · RAZOR, SURGICAL6 reports
- CARDINAL HEALTH CAH4414Cardinal Health 200, LLC · RAZOR, SURGICAL5 reports
- 3M™ 96903m Company · RAZOR, SURGICAL2 reports
- Medline DYND70880Medline Industries, Inc. · RAZOR, SURGICAL1 report
- SensiClip 4403ACarefusion 2200, Inc · RAZOR, SURGICAL1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.