Kangaroo 8884752505 (UDI 30884521009664)Recalls & reported problems
Cardinal Health, Inc. · Naso-Jejunal Feeding Tube,7 g Weighted Tip · Class 2
All models of Kangaroo (145)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 30884521009664), received between Oct 14, 2021 and Oct 14, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Twisted/Bent4
- Complete Blockage2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 14, 2021Complete Blockage; Material Twisted/Bent; Complete Blockage; Material Twisted/Bent
THE CUSTOMER REPORTED THAT THE TUBE KINKED WHILE THE USER WAS INSERTING THE TUBE INTO THE PATIENT. NO PATIENT INJURY REPORTED. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER ON (B)(6) 2021 STATED THAT THE NJ FEEDING TUBE WAS INSERTED ON (B)(6) 2021, AND KINKED AT ONE OF THE SIDE HOLES ON (B)(6) 2021, RESULTING IN A BLOCK TUBE REQUIRING A CHANGE OF THE NJ FEEDING TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12636961 ↗MalfunctionOct 14, 2021Material Twisted/Bent; Material Twisted/Bent
THE CUSTOMER REPORTED THAT THE TUBE KINKED WHILE THE USER WAS INSERTING THE TUBE INTO THE PATIENT. NO PATIENT INJURY REPORTED. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER ON 13-OCT-2021 STATED THAT THE NJ FEEDING TUBE WAS INSERTED ON (B)(6) 2021, BUT WAS FOUND TO BE KINKED AT THE LEVEL OF THE SIDE HOLES SEEN ON XRAY ON (B)(6) 2021. THEY CONSULTED WITH THE INTERVENTIONAL RADIOLOGIST TO ADJUST THE POSITION OF THE TUBE, BUT THEY WERE UNABLE TO PASS THE GUIDEWIRE THROUGH THE KINK AREA. A NEW TUBE WAS PLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12636825 ↗MalfunctionOct 14, 2021Material Twisted/Bent; Material Twisted/Bent
THE CUSTOMER REPORTED THAT THE TUBE KINKED WHILE THE USER WAS INSERTING THE TUBE INTO THE PATIENT. NO PATIENT INJURY REPORTED. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER ON 13-OCT-2021 STATED THAT THE PATIENT UNDERWENT A TUBE CHANGE WITH FLUOROSCOPY ON (B)(6) 2021 HOWEVER AT THE WARD ONE HOUR AFTER CHANGING THE TUBE, THE X-RAY SHOWED THAT THE TUBE HAD KINKED SUBSEQUENTLY AT ONE OF THE SIDE HOLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12636808 ↗MalfunctionOct 14, 2021Complete Blockage; Material Twisted/Bent; Complete Blockage; Material Twisted/Bent
THE CUSTOMER REPORTED THAT THE TUBES KINKED WHILE THE USER WAS INSERTING THE TUBE INTO THE PATIENT. NO PATIENT INJURY. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER ON 13-OCT-2021 STATED THAT THE NJ FEEDING TUBE WAS INSERTED ON (B)(6) 2021, AND KINKED AT ONE OF THE SIDE HOLES ON (B)(6) 2021, RESULTING IN A BLOCK TUBE REQUIRING CHANGING OF THE NJ FEEDING TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12636719 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page