Skip to content

SURFLO SR-OX1451CARecalls & reported problems

Terumo (Philippines) Corporation · SURFLO IV Catheter · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 34806017510146), received between Nov 9, 2024 and Nov 9, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE 14-GAUGE SURFLO CATHETER WAS PLACED BY A CRNA PRIOR TO CARDIAC SURGERY. ON POST-OPERATIVE DAY 5, WHILE ATTEMPTING TO REMOVE THE CATHETER, THE PCU RN NOTICED THAT THE HUB HAD DETACHED, LEAVING THE CATHETER RETAINED IN THE PATIENT'S VEIN. CONSEQUENTLY, THE PATIENT REQUIRED A MINOR SURGICAL PROCEDURE BY THE ORTHO - HAND SURGICAL SERVICE TO REMOVE THE RETAINED CATHETER ON (B)(6) 2024. THE SURGEON HANDED THE RETAINED CATHETER TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 16 OCT 2024: THERE WAS NO DOCUMENTED DIFFICULTY INTRODUCING THE CATHETER WHEN IT WAS PLACED ON (B)(6) 2024, NOR WAS THERE ANY DOCUMENTATION OF FLUID LEAKAGE DURING THE FIVE DAYS IT WAS IN PLACE. THE DRUG BEING DELIVERED THROUGH THE CATHETER IS UNKNOWN. THE CATHETER WAS SECURED WITH 3M TEGADERM IV 1633 DURING THIS TIME. THE METHOD USED TO REMOVE THE SECURE DRESSING IS BY HAND. IT APPEARS THAT THE WHITE CATHETER PORTION SIMPLY DISCONNECTED FROM THE HUB. THE CATHETER DID NOT TRAVEL FAR WITHIN THE ANATOMY AND REMAINED IN THE RADIAL ASPECT OF THE DISTAL RIGHT FOREARM. EXTENT OF THE MINOR SURGICAL PROCEDURE: THE PROCEDURE REPORT STATES THAT A PALPABLE THIN OBJECT APPROXIMATELY 5 CM

    Patient problems: Foreign Body In Patient

    MAUDE report 20646846 ↗
  • InjuryNov 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE 14-GAUGE SURFLO CATHETER WAS PLACED BY A CRNA PRIOR TO CARDIAC SURGERY. ON POST-OPERATIVE DAY 5, WHILE ATTEMPTING TO REMOVE THE CATHETER, THE PCU RN NOTICED THAT THE HUB HAD DETACHED, LEAVING THE CATHETER RETAINED IN THE PATIENT'S VEIN. CONSEQUENTLY, THE PATIENT REQUIRED A MINOR SURGICAL PROCEDURE BY THE ORTHO - HAND SURGICAL SERVICE TO REMOVE THE RETAINED CATHETER ON (B)(6) 2024. THE SURGEON HANDED THE RETAINED CATHETER TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 16 OCT 2024: THERE WAS NO DOCUMENTED DIFFICULTY INTRODUCING THE CATHETER WHEN IT WAS PLACED ON (B)(6) 2024, NOR WAS THERE ANY DOCUMENTATION OF FLUID LEAKAGE DURING THE FIVE DAYS IT WAS IN PLACE. THE DRUG BEING DELIVERED THROUGH THE CATHETER IS UNKNOWN. THE CATHETER WAS SECURED WITH 3M TEGADERM IV 1633 DURING THIS TIME. THE METHOD USED TO REMOVE THE SECURE DRESSING IS BY HAND. IT APPEARS THAT THE WHITE CATHETER PORTION SIMPLY DISCONNECTED FROM THE HUB. THE CATHETER DID NOT TRAVEL FAR WITHIN THE ANATOMY AND REMAINED IN THE RADIAL ASPECT OF THE DISTAL RIGHT FOREARM. EXTENT OF THE MINOR SURGICAL PROCEDURE: THE PROCEDURE REPORT STATES THAT A PALPABLE THIN OBJECT APPROXIMATELY 5 CM

    Patient problems: Foreign Body In Patient

    MAUDE report 20646786 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →