NA 6007-003-150Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · NAVIGATION PIN 3 MM X 150 MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 37613154107408), received between Jun 25, 2018 and Sep 30, 2022 — classified as 5 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Break2
- Fracture2
- Material Deformation1
Patient problems
- Bone Fracture(s)3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Paralysis1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PER THE CUSTOMER, WHEN USING THE NAVIGATION DEVICE TWO PINS WERE USED TO STAND THE TRACKER ON THE FEMUR, ONE OF WHICH PENETRATED THE CORTICAL BONE, CAUSING A SUPRACONDYLAR FRACTURE. THE CUSTOMER STATES THAT AFTER THE TOTAL KNEE ARTHROPLASTY PROCEDURE WAS COMPLETED A SUPRACONDYLAR FRACTURE OCCURRED AND THE PATIENT UNDERWENT REVISION SURGERY AND PLATE FIXATION AT A LATER DATE. PER FURTHER CLARIFICATION FROM THE SURGEON, "THE PATIENT'S BONE QUALITY IS NOT STRONG," IN THE AREA WHERE THE PINS WERE PLACED. ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE USER FACILITY.
Patient problems: Bone Fracture(s)
MAUDE report 15518905 ↗InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(4) 2021 A TOTAL KNEE PROCEDURE WAS PERFORMED AT THE USER FACILITY, ON (B)(6) 2021 THE PATIENT EXPERIENCED A SUPRA FEMORAL FRACTURE IN THE SAME AREA. THE PATIENT WAS TREATED FOR THE FRACTURE. THERE ARE NO FURTHER KNOWN ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION ASSOCIATED WITH THE EVENT.
Patient problems: Bone Fracture(s)
MAUDE report 11757499 ↗InjuryNov 18, 2020Break; Fracture; Break; Fracture
PER THE SURGEON WHOM PERFORMED THE PROCEDURE AT THE USER FACILITY THE ORTHOLOCK NAVIGATION PIN INSERTION WAS LATERALLY DISPLACED AND BECAME TRANSCORTICAL. SOMETIME AFTER THE PROCEDURE THE PATIENT'S FEMUR BROKE. PER THE SURGEON THERE IS NO ALLEGATION OF A MALFUNCTION BY THE ORTHOLOCK PIN OR ANY MEDICAL DEVICE USED DURING THE PROCEDURE. MEDICAL INTERVENTION WAS NEEDED TO TREAT THE BROKEN FEMUR. THERE ARE NO FURTHER KNOWN ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION ASSOCIATED WITH THE PROCEDURE.
Patient problems: Bone Fracture(s)
MAUDE report 10864133 ↗InjuryOct 1, 2019Break; Fracture; Material Deformation; Break; Fracture; Material Deformation
IT WAS REPORTED THAT DURING A UKA SURGERY, THE DEVICE WAS INSERTED INTO THE FEMUR, BUT IT WAS DIFFICULT TO ENTER. THE IMAGE WAS CHECKED AND THE PIN WAS BENT. WHEN THE HEALTH CARE PROFESSIONAL TRIED TO REMOVE IT, THE PIN BROKE FROM THE THREAD AT THE TIP. THE OPERATION WAS COMPLETED WITH AN ALTERNATIVE PIN. AT THE TIME OF CREATING A PEG HOLE AFTER CUTTING THE FEMUR, THE PEG HOLE WAS FURTHER DEEPENED WITH A HANSON PIN DRILL, AND THE BROKEN PART WAS EXTRACTED WITH FORCEPS. IT WAS REPORTED THAT THERE WERE NO PARTS OF THE PIN LEFT IN THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9139869 ↗InjuryJun 25, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE NAVIGATION PIN WAS STUCK TOO DEEP INTO THE FIBULA. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; OR MEDICAL INTERVENTION. IT WAS FURTHER REPORTED THAT A SYMPTOM LIKE PERONEAL PARALYSIS WAS OBSERVED AFTER THE PROCEDURE. ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE USER FACILITY.
Patient problems: Paralysis
MAUDE report 7635706 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page