Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 40801902005128), received between Aug 31, 2020 and Nov 11, 2025 — classified as 4 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Material Separation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 5 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation6
  • Material Twisted/Bent5
  • Complete Blockage2
  • Failure to Advance2
  • Fluid/Blood Leak1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Foreign Body In Patient4
  • Pain1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT: "EPIDURAL SPINAL CATHETER WHICH IS BROKEN 3 CM BELOW THE SKIN SURFACE. THE BROKEN CATHETER TIP IS AT THE INTERSPINOUS SPACE OF L3-4. THE PATIENT UNDERWENT A RIGHT LAMINOTOMY AND REMOVAL OF RETAINED EPIDURAL CATHETER." ASSOCIATED COMPLAINTS: 9680794-2025-00939.

    Patient problems: Foreign Body In Patient

    MAUDE report 23520195 ↗
  • InjuryNov 7, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT "DURING THE EPIDURAL CATHETER REMOVAL SEGMENT OF THE CATHETER WAS RETAINED. CT SCAN - THE IMAGES SHOW BROKEN EPIDURAL CATHETER IN THE RIGHT-SIDE SPINAL CANAL OUTSIDE OF THE THECAL SAC AT L3-4 EXTENDING TO THE SUBCUTANEOUS FAT 5 CM FROM THE SKIN SURFACE. THE PATIENT HAS A SCHEDULED PRE-SURGERY APPOINTMENT WITH THE NEUROSURGEON THIS MONTH."

    Patient problems: Foreign Body In Patient

    MAUDE report 23494671 ↗
  • InjuryAug 13, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT: "NURSE REMOVED EPIDURAL CATHETER ON (B)(6), UNAWARE THAT PART OF THE CATHETER WAS STILL IN THE PATIENT. FAST FORWARD A COUPLE WEEKS LATER, WITH A COUPLE INSTANCES OF THE PATIENT CALLING IN FOR BACK PAIN AND SHE DID HAVE SOME PARESTHESIAS IN HER LOWER EXTREMITIES. THE PATIENT NEEDED A STAT CT FOR IMAGING, TO FIND OUT EXACTLY WHAT PART WAS LEFT INSIDE, AND THEN NEEDED EMERGENCY LAMINECTOMY, FOREIGN BODY REMOVAL. FOR THE PROCEDURE, SHE NEEDED GENERAL ANESTHESIA, INTUBATION, MEDICATIONS FOR THE PROCEDURE ITSELF, ADDITIONAL PAIN MEDICATION AND THEN NEUROSURGICAL OBSERVATION DURING HER ADMISSION. PATIENT HAS RECOVERED AND IS AT HOME, AFTER DELIVERY AND HER EMERGENCY SURGERY. SHE'S BEING FOLLOWED UP BY OBSTETRICS AND NEUROSURGERY."

    Patient problems: Foreign Body In Patient; Pain

    MAUDE report 22793745 ↗
  • MalfunctionFeb 19, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT: "CATHETER LEAKED WHILE ADMINISTERING DRUGS FOR ADDUCTOR CANAL BLOCK. THERE WAS NO REPORTED PATIENT HARM OR CON SEQUENCE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21409436 ↗
  • MalfunctionDec 20, 2024Material Separation; Failure to Advance; Material Separation; Failure to Advance

    IT WAS REPORTED THAT: "ANESTHESIOLOGIST WAS NOT ABLE TO PASS THE CATHETER THROUGH THE EPIDURAL NEEDLE. SHE REMOVED THE CATHETER AND NOTICED THE WIRE REINFORCED SMALL SPRING INSIDE WAS BROKEN INTO PIECES INSIDE THE CATHETER. NO PATIENT INJURY OCCURRED. CATHETER WAS REMOVED AND A COMPETITIVE CATHETER WAS SUCCESSFULLY PLACED." THERE WAS NO PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 9680794-2024-01257, 9680794-2024-01258 AND 9680794-2024-01259.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20986326 ↗
  • MalfunctionDec 20, 2024Material Separation; Failure to Advance; Material Separation; Failure to Advance

    IT WAS REPORTED THAT: "ANESTHESIOLOGIST WAS NOT ABLE TO PASS THE CATHETER THROUGH THE EPIDURAL NEEDLE. SHE REMOVED THE CATHETER AND NOTICED THE WIRE REINFORCED SMALL SPRING INSIDE WAS BROKEN INTO PIECES INSIDE THE CATHETER. NO PATIENT INJURY OCCURRED. CATHETER WAS REMOVED AND A COMPETITIVE CATHETER WAS SUCCESSFULLY PLACED." THERE WAS NO PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 9680794-2024-01257, 9680794-2024-01258, 9680794-2024-01259.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20986148 ↗
  • MalfunctionJun 27, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE INTERNAL WIRE IN THE EPIDURAL CATHETER IS KINKING / BINDING UP WHEN TRYING TO PUSH THROUGH THE NEEDLE CAUSING THEM TO USE MULTIPLE CATHETERS". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19624007 ↗
  • MalfunctionJun 27, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE INTERNAL WIRE IN THE EPIDURAL CATHETER IS KINKING / BINDING UP WHEN TRYING TO PUSH THROUGH THE NEEDLE CAUSING THEM TO USE MULTIPLE CATHETERS". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19623962 ↗
  • MalfunctionJun 27, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE INTERNAL WIRE IN THE EPIDURAL CATHETER IS KINKING / BINDING UP WHEN TRYING TO PUSH THROUGH THE NEEDLE CAUSING THEM TO USE MULTIPLE CATHETERS". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19623939 ↗
  • MalfunctionJun 27, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE INTERNAL WIRE IN THE EPIDURAL CATHETER IS KINKING / BINDING UP WHEN TRYING TO PUSH THROUGH THE NEEDLE CAUSING THEM TO USE MULTIPLE CATHETERS". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19623867 ↗
  • MalfunctionJun 27, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE INTERNAL WIRE IN THE EPIDURAL CATHETER IS KINKING / BINDING UP WHEN TRYING TO PUSH THROUGH THE NEEDLE CAUSING THEM TO USE MULTIPLE CATHETERS". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19623865 ↗
  • MalfunctionMar 21, 2022Complete Blockage; Complete Blockage

    RAN INTO THIS ISSUE TODAY ON L & D. EPIDURAL CATHETER WAS SEALED AT THE PROXIMAL END. ADDITIONAL INFORMATION: THE CATHETER HAD ALREADY BEEN INSERTED INTO THE PATIENT. BECAUSE THE DEFECT WAS THE END DISTAL TO THE PATIENT IT WAS NOT DISCOVERED UNTIL THE TUOCHY NEEDLE WAS REMOVED AND SOMETHING CAUGHT AS THE CATHETER PASSED THROUGH THE LUMEN. I ENDED UP TAKING STERILE SCISSORS AND CUTTING OFF THE SEALED PORTION WHICH ALLOWED THE CATHETER TO FUNCTION AS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13839474 ↗
  • MalfunctionDec 13, 2021Complete Blockage; Complete Blockage

    I HAD AN EPIDURAL KIT WITH AN EPIDURAL CATHETER WITH THE TIP STILL SEALED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12983548 ↗
  • InjuryAug 31, 2020Material Separation; Physical Resistance/Sticking

    PER ANESTHESIOLOGIST, ATTEMPT AT L4/5, OS AT MULTIPLE ANGLES BEFORE LOR (LOSS OF RESISTANCE), UNABLE TO THREAD CATHETER SMOOTHLY. CONSIDERED THIN EPIDURAL SPACE AT THAT LEVEL NOT SUITABLE FOR EPIDURAL CATHETER. ATTEMPT AT L3/4, OS AT MULTIPLE ANGLES BEFORE LOR, AGAIN UNABLE TO THREAD CATHETER SMOOTHLY. ATTEMPT AT L5/S1, LOR AT 7 CM, EPIDURAL CATHETER ADVANCED SMOOTHLY INTO EPIDURAL SPACE. RESISTANCE WHEN RETRIEVING NEEDLE, STOPPED PULLING IMMEDIATELY ON NEEDLE, AND DECIDED TO PULL NEEDLE AND EPIDURAL CATHETER TOGETHER. WHEN NEEDLE REMOVED, SAW EXPOSED INNER WIRE FROM CATHETER. PULLED ALL OF WIRE OUT. COMPARED PULLED CATHETER TO NEW CATHETER AND NOTED APPROXIMATELY 5 CM OF CATHETER PRESUMED TO REMAIN IN EPIDURAL SPACE. PATIENT DENIED PARAESTHESIAS OR NEUROLOGICAL SYMPTOMS DURING ALL THE ATTEMPTS AND AFTER REMOVAL OF CATHETER. ANOTHER CATHETER WAS PLACED WITHOUT COMPLICATION. PATIENT UNDERWENT CESAREAN SECTION WITHOUT COMPLICATION. NEUROSURGEON CONSULTED. CT SCAN ORDER OF THE LUMBAR SPINE TO DETERMINE THE POSITION OF THE RETAINED CATHETER FRAGMENT. CT SCAN CONFIRMED PRESENCE OF RETAINED CATHETER AT THE L4-5 LEVEL, APPEARED TO BE EPIDURAL, LEFT SIDE EXTENDING INTO THE PROXIMAL PORTION

    Patient problems: Foreign Body In Patient

    MAUDE report 10471035 ↗

Other CATHETER, CONDUCTION, ANESTHETIC devices

All CATHETER, CONDUCTION, ANESTHETIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.