Skip to content

ACIST CVi® Angiographic Injection System 014657Recalls & reported problems

Acist Medical Systems, Inc. · CVi Injector p/n 800620-021 Contents: CVi Injector Head, Bed Mount, & Power Supply · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 40841716101418), received between Feb 9, 2018 and Jul 24, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Gas/Air Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Gas/Air Leak1
  • Insufficient Device Problem Information1

Patient problems

  • Vascular Dissection3
  • Air Embolism2
  • Aortic Dissection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A LEFT HEART CATHETERIZATION FOR CARDIOMYOPATHY AND ACUTE HEART FAILURE, AIR WAS INJECTED INTO A PATIENT. FOLLOWING RADIAL SHEATH PREP FOR SPASM PROPHYLAXIS, A 5 FRENCH JL3.5 CATHETER WAS TAKEN OVER A J-WIRE AND USED TO ENGAGE THE LEFT MAIN OSTIUM FOR LEFT CORONARY ANGIOGRAPHY. THE CATHETER WAS FLUSHED TWICE PER STANDARD PROCESS USING 10CC SYRINGES; THERE WAS NO AIR ALERT MESSAGE DISPLAYED BY THE CVI INJECTION SYSTEM. A SUBSEQUENT TEST INJECTION WAS COMPLETED TO CONFIRM ENGAGEMENT OF THE LEFT MAIN. THERE WAS SIGNIFICANT AIR EMBOLISM INTO THE CIRCUMFLEX AND THE LEFT ANTERIOR DESCENDING ARTERY (LAD), ALONG THE MID CIRCUMFLEX AND MID LAD WITH NO FLOW IN THE DISTAL VESSELS SEEN. AN INITIAL QUICK ATTEMPT TO ASPIRATE THROUGH THE JL 3.5 DIAGNOSTIC CATHETER WAS UNSUCCESSFUL. THE PATIENT QUICKLY DETERIORATED, BECAME BRADYCARDIC, AND WAS GIVEN 1 MG OF ATROPINE. THE PATIENT BECAME UNRESPONSIVE AND AN EMERGENT AIRWAY WAS CALLED; HE WAS INTUBATED. THE PATIENT SUBSEQUENTLY WENT INTO PULSELESS ELECTRICAL ACTIVITY (PEA) ARREST AND CARDIOPULMONARY RESUSCITATION (CPR) WAS INITIATED. AFTER ONE ROUND OF CPR AND 1 MG OF EPINEPHRINE (EPI), THERE WAS RETURN OF SPONTANEOUS CIRCULATION (ROSC). THE U

    Patient problems: Air Embolism

    MAUDE report 22588983 ↗
  • InjuryJan 29, 2020Gas/Air Leak; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER INJECTING INTO THE LEFT VENTRICLE DURING A LEFT VENTRICLE CATHETERIZATION PROCEDURE, THE PHYSICIAN RINSED THE HIGH PRESSURE TUBING WITH SALINE TO PERFORM PRESSURE MEASUREMENTS IN THE LEFT VENTRICLE AND NOTED BUBBLES IN THE HIGH PRESSURE TUBING. THE PATIENT EXPERIENCED A TRANSIENT ISCHEMIC ATTACK (TIA) DURING THE PROCEDURE. A COMPUTED TOMOGRAPHY (CT) SCAN WAS PERFORMED AND THE RESULTS CONFIRMED NO AIR WAS PRESENT IN THE PATIENT'S BRAIN. THE USER FACILITY NOTES THEY DID NOT CONFIRM AIR WAS INJECTED INTO THE PATIENT DURING THE CATHETERIZATION PROCEDURE.

    Patient problems: Air Embolism

    MAUDE report 9645407 ↗
  • InjuryMay 23, 2019Adverse Event Without Identified Device or Use Problem

    DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO THE RIGHT CORONARY ARTERY (RCA), THE PATIENT'S RCA AND AORTA WERE DISSECTED. THE PATIENT WAS TAKEN TO EMERGENCY SURGERY AND A SAPHENOUS VEIN GRAFT TO THE RCA WAS COMPLETED AND THE AORTA WAS REPAIRED.

    Patient problems: Aortic Dissection; Vascular Dissection

    MAUDE report 8634899 ↗
  • InjuryFeb 9, 2018Adverse Event Without Identified Device or Use Problem

    USER FACILITY REPORTED: A MALE PATIENT, AGE (B)(6) YEARS, UNDERGOING A CORONARY ANGIO FOR A NON-ST-ELEVATION MYOCARDIAL INFARCTION (NSTEMI), HAD A CORONARY DISSECTION DURING A CORONARY ANGIO. THERE WERE NO CLINICAL SYMPTOMS. THE EVENT RESULTED IN PROLONGATION OF HOSPITALIZATION. THE PATIENT RECOVERED. RECOVERY DATE WAS NOT PROVIDED.

    Patient problems: Vascular Dissection

    MAUDE report 7261507 ↗
  • InjuryFeb 9, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem

    USER FACILITY REPORTED: A FEMALE PATIENT, AGE (B)(6), UNDERGOING A CORONARY ANGIO FOR A NON-ST-ELEVATION MYOCARDIAL INFARCTION (NSTEMI), HAD A CORONARY DISSECTION FOLLOWING CORONARY ANGIO. DISSECTION OCCURRED AFTER THE SECOND INJECTION INTO THE RIGHT CORONARY ARTERY (RCA). PATIENT EXPERIENCED CHEST PAIN OF LESS THAN ONE HOUR IN DURATION, ST-SEGMENT ELEVATION, AND EVIDENCE OF MYOCARDIAL INFARCTION. THE PATIENT RECOVERED. RECOVERY DATE WAS NOT PROVIDED.

    Patient problems: Vascular Dissection

    MAUDE report 7261503 ↗

Other Injector And Syringe, Angiographic devices

All Injector And Syringe, Angiographic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.