ACIST CVi® Angiographic Injection System 014658Recalls & reported problems
Acist Medical Systems, Inc. · CVi Injector p/n 800620-022 Contents: CVi Injector Head (Siemens), Bed Mount, & Power Supply · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 40841716101425), received between Feb 24, 2018 and Jan 21, 2025 — classified as 1 death, 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Insufficient Device Problem Information1
Patient problems
- Air Embolism4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
DeathJan 21, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
DURING A PERCUTANEOUS CORONARY INTERVENTION USING THE ACIST CVI INJECTION SYSTEM, AIR WAS INJECTED INTO A PATIENT. THE AIR INJECTION OCCURRED AFTER INSERTION OF A CORONARY GUIDING CATHETER AND ENGAGEMENT INTO THE LEFT MAIN CORONARY ARTERY AND UPON INJECTION OF CONTRAST MEDIA. THE PATIENT EXPERIENCED CARDIOGENIC CRASH AND SHOCK AND EXPIRED THE SAME DAY.
Patient problems: Air Embolism
MAUDE report 21203140 ↗InjuryDec 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AIR WAS INJECTED INTO A PATIENT DURING AN ANGIOGRAPHY. THE PATIENT WAS REVIVED AND WAS REPORTED TO HAVE RECOVERED AFTER THE AIR INJECTION.
Patient problems: Air Embolism
MAUDE report 20922050 ↗InjuryDec 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A PLANNED ELECTROPHYSIOLOGICAL EXAMINATION WITH ABLATION OF THE PULMONARY VEINS, WHEN USING SALINE TO FLUSH THE ACIST CONSUMABLE KITS ATTACHED TO THE ACIST INJECTION SYSTEM, MODEL CVI, THERE WERE REPEATED MASSIVE AIR BUBBLES IN THE CONSUMABLE TUBING. THE USER RESOLVED THE ERROR BY CHANGING THE DELTAMEDICA SALINE INFUSION CONTAINER CONNECTED TO THE CVI CONSUMABLE KIT WITH ANOTHER MANUFACTURER'S SALINE CONTAINER (BRAUN). THE PROCESS OF CHANGING THE SALINE BAG RESULTED IN AN AIR BEING INJECTED INTO THE PATIENT'S CORONARY ARTERIES. THE PATIENT REQUIRED RESUSCITATION. THE USER NOTICED THAT THE WALLS OF THE DELTAMEDICA GMBH, SALINE CONTAINER ARE HARDER THAN THOSE OF THE BRAUN SALINE CONTAINER. EMPTYING THE BOTTLE COULD RESULT IN THE FORMATION OF NEGATIVE INTERNAL PRESSURE WITH A SUBSEQUENT SUCTION EFFECT. ADDITIONALLY, THE RUBBER PLUG IS HARDER THAN THAT THE BRAUN CONTAINER AND IT REQUIRES SIGNIFICANTLY MORE FORCE TO BE PIERCED WITH THE INFUSION CANNULA. THE HARDNESS OF THE RUBBER CAN LEAD TO INADEQUATE SEALING. THE DELTAMEDICA ISOTONIC SODIUM CHLORIDE SOLUTION, 0.9% PRODUCT IS NOT MANUFACTURED OR DISTRIBUTED BY ACIST. IT IS MANUFACTURED BY DELTAMEDIA GMBH.
Patient problems: Air Embolism
MAUDE report 20848139 ↗InjuryFeb 24, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
USER FACILITY REPORTED: DURING A PROCEDURE USING THE ACIST ANGIOGRAPHIC INJECTION SYSTEM, MODEL CVI, AIR WAS INJECTED INTO A PATIENT'S RIGHT CORONARY ARTERY (RCA). THE AIR BUBBLE WAS VISIBLE IN THE RCA. SUBSEQUENTLY, THE PHYSICIAN FLUSHED WITH SALINE AND THE RCA WAS COMPLETELY FREE OF FLOW. THE PATIENT EXPERIENCED ST-SEGMENT ELEVATION AND HYPOTENSION. THE PHYSICIAN PERFORMED SEVERAL FLUSHES OF SALINE INTO THE RCA, AND FLOW RETURNED WITHIN THE RCA AND THE PATIENT STABILIZED. THERE WAS NO INTERVENTION TO THE PATIENT. THE PATIENT REMAINED STABLE AND WAS MONITORED IN THE USER FACILITY'S INTENSIVE CARE UNIT.
Patient problems: Air Embolism
MAUDE report 7294437 ↗
Other Injector And Syringe, Angiographic devices
- Monarch® 00884450298022Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls37 reports
- Blue Diamond™ 00884450498743Merit Medical Systems, Inc. · Injector and syringe, angiographic2 recalls0 reports
- Monarch® 00884450298053Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- Monarch® 00884450298060Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- NAMIC 80085007Medline Industries, Inc. · Injector and syringe, angiographic1 recall36 reports
- NA 622510B. Braun Medical Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC85 reports
- NAMIC 80087007Medline Industries, Inc. · Injector and syringe, angiographic1 recall20 reports
- Blue Diamond™ 00884450381168Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page