Shiley 18875 (UDI 40884521032539)Recalls & reported problems

Covidien LP · TaperGuard Evac Oral Tracheal Tube,Murphy Eye · Class 2

All models of Shiley (387)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien TaperGuard Evac breathing tube is placed through the mouth into the windpipe to help a patient breathe. Of its 571 reports, 407 describe malfunctions, most often air leaks, and 160 describe injuries. Patients mention the tube coming out unintentionally in 274 reports and low oxygen in 30. Its reports record 4 deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
568
Deaths reported
4
Injuries reported
158
Recalls naming it
0

Summary

The FDA's MAUDE database contains 568 adverse event reports that list this exact device (DI 40884521032539), received between Oct 27, 2017 and Sep 21, 2026 — classified as 4 deaths, 158 injuries, 406 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (269 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 7 reportsJan 2024: 6 reportsFeb 2024: 6 reportsMar 2024: 3 reportsApr 2024: 9 reportsMay 2024: 6 reportsJun 2024: 14 reportsJul 2024: 9 reportsAug 2024: 5 reportsSep 2024: 12 reportsOct 2024: 3 reportsNov 2024: 9 reportsDec 2024: 2 reportsJan 2025: 18 reportsFeb 2025: 3 reportsMar 2025: 16 reportsApr 2025: 9 reportsMay 2025: 22 reportsJun 2025: 17 reportsJul 2025: 16 reportsAug 2025: 5 reportsSep 2025: 4 reportsOct 2025: 4 reportsNov 2025: 8 reportsDec 2025: 15 reportsJan 2026: 20 reportsFeb 2026: 1 reportsMar 2026: 5 reportsApr 2026: 7 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 8 reportsAug 2026: 5 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak269
  • Use of Device Problem174
  • Defective Component38
  • Material Split, Cut or Torn38
  • Disconnection29
  • Inflation Problem11
  • Appropriate Device Problem Term/Code Not Available8
  • Connection Problem8

Patient problems

  • Unintended Extubation272
  • No Clinical Signs, Symptoms or Conditions169
  • Extubate61
  • Insufficient Information38
  • Hypoventilation34
  • Low Oxygen Saturation29
  • No Consequences Or Impact To Patient12
  • No Information11

Recent adverse event reports

The 25 most recent of 568 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Gas/Air Leak

    Patient problems: Unintended Extubation

    MAUDE report 50848669 ↗
  • MalfunctionSep 16, 2026Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794103 ↗
  • MalfunctionAug 27, 2026Material Deformation

    It was reported that during use, an endotracheal intubation was performed as ordered by the physician. After the circuit was secured, it was found that the endotracheal tube could not remain upright and tended to fold outward. Manual intervention was provided to keep the circuit in an upright position. The tube was replaced.

    Patient problems: Unintended Extubation

    MAUDE report 50487942 ↗
  • InjuryAug 20, 2026Use of Device Problem

    It was reported that the patient¿s dyspnea gradually worsened, but remained conscious with low skin temperature in the extremities and scattered mottling over the body. Therefore, endotracheal intubation and ventilator-assisted ventilation were provided. Before intubation, a routine cuff seal test was performed, which showed cuff leakage. After the endotracheal tube was replaced, the intubation was completed successfully. The event resulted in a delay to treatment and hypoxia because the cuff leak was not detected in time.

    Patient problems: Hypoxia

    MAUDE report 50359821 ↗
  • InjuryAug 19, 2026Gas/Air Leak

    It was reported that during pre-usage inspection, when the cuff seal of the endotracheal tube was then checked, and after inflation, the pressure dropped rapidly and the cuff rapidly deflated, indicating an air leak in the cuff. The medical staff immediately replaced it with a brand-new endotracheal tube of the same specification, and the second intubation was completed successfully. After connection to the ventilator, the patient¿s breathing gradually stabilized, and no obvious physical injury was observed at this stage.

    Patient problems: Hypoventilation; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50330910 ↗
  • InjuryAug 13, 2026Gas/Air Leak

    It was reported that the ventilator alarmed, the tidal volume had decreased significantly, and the tidal volume fluctuated widely. Gas was observed being exhaled from the patient¿s mouth, inflating the cuff was ineffective, and the cuff pressure was found to be significantly low. External factors such as circuit disconnection, water accumulation, and air trapping were ruled out, and cuff leakage was considered. The ventilator oxygen concentration was adjusted to 100%, secretions and saliva were fully suctioned, the original endotracheal tube was removed and replaced, and the procedure went smoothly. After reconnection to the ventilator, the alarm disappeared, the original ventilator oxygen concentration was restored, and after 5 minutes of observation, oxygen saturation was 100%.

    Patient problems: Hypoventilation; Unintended Extubation

    MAUDE report 50244300 ↗
  • InjuryAug 8, 2026Gas/Air Leak

    It was reported that after the patient developed weak breathing due to intracerebral hemorrhage, endotracheal intubation was ordered by the physician and the intubation was successful. After intubation, the patient was connected to a ventilator for assisted breathing, but the ventilator issued a continuous low-pressure alarm. The alarm persisted even after repeated adjustments, the cuff pressure of the tube was checked and found to be slightly low, then the cuff was inflated. After inflation, the ventilator alarm disappeared then about 2 minutes later, the ventilator again issued a low-pressure alarm, which happened repeatedly. The tube was replaced, and when the removed tube was inspected, it inflated smoothly and the cuff was full. After water was injected into the cuff, continuous water jets were observed from the cuff wall, confirming cuff leakage. The event delayed the patient¿s treatment but no apparent harm.

    Patient problems: Unintended Extubation

    MAUDE report 50184595 ↗
  • MalfunctionJul 30, 2026Gas/Air Leak

    Patient problems: Unintended Extubation

    MAUDE report 50082428 ↗
  • InjuryJul 29, 2026Use of Device Problem

    It was reported that before intubation, the physician tested the cuff pressure and found that the cuff could not be inflated. Inspection showed that the cuff had ruptured, and the endotracheal tube was immediately replaced. This prolonged the patient¿s resuscitation time.

    Patient problems: Insufficient Information

    MAUDE report 50068101 ↗
  • InjuryJul 16, 2026Material Deformation

    IT WAS REPORTED THAT AFTER THE PATIENT WAS TRANSFERRED FROM THE OPERATING ROOM, THE PATIENT HAD LOW PERIPHERAL OXYGEN SATURATION AND RHONCHI WERE HEARD IN THE THROAT. BLOOD GAS ANALYSIS SHOWED A PARTIAL PRESSURE OF OXYGEN OF 41 MMHG. TO IMPROVE THE PATIENT¿S VENTILATION AND OXYGENATION, ENDOTRACHEAL INTUBATION WAS PERFORMED. WHEN THE NURSE PERFORMED SUCTION CARE, THERE WAS OBVIOUS RESISTANCE AFTER THE SUCTION CATHETER ENTERED THE ENDOTRACHEAL TUBE. THE PERIPHERAL OXYGEN SATURATION DECREASED TO 88%, AND THE VENTILATOR TIDAL VOLUME FLUCTUATED WITHIN THE RANGE OF 200¿320 ML. TO ASSESS THE CONDITION OF THE ENDOTRACHEAL TUBE AND LUNGS AND TO ASSIST WITH SPUTUM CLEARANCE, BRONCHOSCOPY WAS THEN PERFORMED. UPON INSERTION OF THE BRONCHOSCOPE THROUGH THE ENDOTRACHEAL TUBE, NARROWING OF THE LUMEN WAS OBSERVED AT THE 20 CM MARKING OF THE TUBE. THE ORIGINAL ENDOTRACHEAL TUBE WAS IMMEDIATELY REMOVED, AND RE-INTUBATION WAS PERFORMED WITH ANOTHER ENDOTRACHEAL TUBE.

    Patient problems: Hypoventilation; Low Oxygen Saturation; Unintended Extubation

    MAUDE report 25839546 ↗
  • MalfunctionJul 16, 2026Gas/Air Leak

    IT WAS REPORTED THAT WHEN THE PATIENT¿S ENDOTRACHEAL TUBE WAS DUE FOR REPLACEMENT. A PRE-INTUBATION TEST SHOWED NO CUFF LEAKAGE. HOWEVER, AFTER INTUBATION, OBVIOUS AIR LEAKAGE WAS NOTED. AFTER CONNECTION TO THE VENTILATOR, THE LEAKAGE VOLUME WAS FOUND TO BE AS HIGH AS 50%, AND THE TUBE WAS IMMEDIATELY REPLACED. THE REMOVED TUBE WAS TESTED IN WATER, WHICH SHOWED NO LEAKAGE FROM THE CUFF ITSELF; THE LEAKAGE SITE WAS FOUND TO BE AT THE CUFF CONNECTION.

    Patient problems: Unintended Extubation

    MAUDE report 25839338 ↗
  • MalfunctionJul 15, 2026Gas/Air Leak

    IT WAS REPORTED THAT AFTER THE STAFF IN THE CARDIAC SURGERY ICU COMPLETED ENDOTRACHEAL INTUBATION FOR THE PATIENT, A LARGE AMOUNT OF GURGLING AIR SOUND WAS HEARD FROM THE PATIENT¿S MOUTH. AFTER THE ENDOTRACHEAL TUBE WAS REMOVED, CUFF LEAKAGE WAS FOUND, AND RE-INTUBATION WAS PERFORMED.

    Patient problems: Unintended Extubation

    MAUDE report 25830295 ↗
  • MalfunctionJul 13, 2026Gas/Air Leak

    IT WAS REPORTED THAT AFTER THE PHYSICIAN SUCCESSFULLY PERFORMED ENDOTRACHEAL INTUBATION, LEAKAGE WAS DETECTED IN THE CUFF DURING INF LATION. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 25804458 ↗
  • MalfunctionJul 9, 2026

    PER REPORTING STAFF, AFTER INTUBATION, IT WAS NOTED THAT THE AIR CUFF WAS NOT HOLDING AIR. PATIENT REQUIRED EMERGENT REINTUBATION. TUBE CHANGED OUT USING GLIDESCOPE AND BOUGIE. AT NO TIME WAS THE PATIENT DESATURATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25763174 ↗
  • MalfunctionJul 7, 2026Use of Device Problem

    IT WAS REPORTED THAT WHILE THE PATIENT WAS ON MECHANICAL VENTILATION VIA THE ENDOTRACHEAL TUBE, THE VENTILATOR ALARMED INDICATING A LEAK IN THE BREATHING CIRCUIT. AFTER CHECKING, THE BREATHING CIRCUIT WAS FOUND TO BE INTACT, AND IT WAS DISCOVERED THAT THE CUFF OF THE ENDOTRACHEAL TUBE HAD RUPTURED. THE ENDOTRACHEAL TUBE WAS REPLACED, AND VENTILATION RETURNED TO NORMAL.

    Patient problems: Unintended Extubation

    MAUDE report 25732601 ↗
  • MalfunctionJun 16, 2026Decoupling

    IT WAS REPORTED THAT DURING USE, THE ORANGE CONNECTOR AT THE END OF THE SUBGLOTTIC SUCTION OF THE ENDOTRACHEAL TUBE WAS FOUND TO BE DETACHED. THIS CAUSED AIR LEAKAGE AND INSUFFICIENT SUCTION PRESSURE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 25527034 ↗
  • MalfunctionMay 28, 2026Gas/Air Leak

    IT WAS REPORTED THAT AT THE TIME OF ENDOTRACHEAL TUBE INSERTION, THE CUFF WAS CHECKED AND NO LEAKAGE WAS FOUND, AND THE CUFF PRESSURE WAS WITHIN THE NORMAL RANGE. CUFF LEAKAGE OCCURRED AFTER APPROXIMATELY 30 HOURS OF USE. THE ENDOTRACHEAL TUBE WAS REPLACED, THE PATIENT¿S VITAL SIGNS REMAINED STABLE.

    Patient problems: Unintended Extubation

    MAUDE report 25325399 ↗
  • MalfunctionMay 14, 2026Use of Device Problem

    IT WAS REPORTED THAT IN PREPARATION FOR INTUBATION, THE CUFF OF THE ENDOTRACHEAL TUBE INFLATED TO ENSURE IT WOULD HOLD AIR. CUFF IMMEDIATELY DEFLATE AND WOULD NOT HOLD AIR. ANOTHER ENDOTRACHEAL TUBE WAS CHECKED OUT AND USED FOR THE INTUBATION. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25172362 ↗
  • InjuryApr 28, 2026Gas/Air Leak

    IT WAS REPORTED THAT AT 19:21 ON (B)(6) 2026, THE PATIENT WAS ADMITTED TO THE DEPARTMENT DUE TO ACUTE HEART FAILURE. AT 19:35, THE PATIENT DEVELOPED SIGNIFICANT DYSPNEA. AS PRESCRIBED, PROPOFOL WAS ADMINISTERED INTRAVENOUSLY FOR SEDATION, AND ORAL ENDOTRACHEAL INTUBATION WAS PERFORMED WITH INSERTION DEPTH OF 24 CM. THE PATIENT WAS THEN CONNECTED TO A VENTILATOR FOR ASSISTED VENTILATION. FROM 23:00 ON (B)(6) TO 08:30 ON (B)(6), THE BEDSIDE NURSE CHECKED THE CUFF PRESSURE EVERY TWO HOURS AND FOUND IT TO BE DEFLATED EACH TIME. THE VENTILATOR FREQUENTLY ALARMED, INDICATING INSUFFICIENT VENTILATION AND LOW TIDAL VOLUME FOR A TOTAL OF 8H . DURING THE NIGHT, MULTIPLE ATTEMPTS TO INFLATE THE CUFF USING A CUFF PRESSURE MONITOR FAILED TO DETECT CUFF PRESSURE OR LEAKAGE, AND CUFF DAMAGE WAS SUSPECTED. AT 08:30, THE ANESTHESIOLOGY TEAM ASSISTED AT THE BEDSIDE TO REPLACE THE TUBE. AFTER REPLACEMENT, THE PATIENT¿S OXYGEN SATURATION RETURNED TO NORMAL AND VENTILATOR PARAMETERS STABILIZED. INSPECTION REVEALED THAT THE CUFF OF THE REMOVED TUBE WAS LEAKING AND DAMAGED.

    Patient problems: Hypoventilation; Low Oxygen Saturation; Unintended Extubation

    MAUDE report 25023402 ↗
  • MalfunctionApr 27, 2026Use of Device Problem

    IT WAS REPORTED THAT THE ENDOTRACHEAL TUBE'S CUFF DID NOT INFLATE WHEN AIR WAS INTRODUCED THROUGH THE PILOT BALLOON. THE CUFF WAS DESCRIBED AS FEELING RIGID DURING THE ATTEMPTED INFLATION TEST. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25009610 ↗
  • MalfunctionApr 20, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE CUFF OF THE ENDOTRACHEAL TUBE HAD AN AIR LEAK DURING USE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 24927997 ↗
  • MalfunctionApr 16, 2026Gas/Air Leak

    IT WAS REPORTED THAT DURING USE, IT WAS FOUND THAT THE ENDOTRACHEAL TUBE'S CUFF PRESSURE WAS INSUFFICIENT. AFTER INFLATION USING A M ANOMETER, THE CUFF PRESSURE REMAINED INADEQUATE. UPON INSPECTION, A SMALL AMOUNT OF MOISTURE WAS OBSERVED INSIDE THE CUFF, AND CUFF AIR LEAKAGE WAS IDENTIFIED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24895025 ↗
  • MalfunctionApr 8, 2026Gas/Air Leak

    IT WAS REPORTED THAT AFTER INTUBATION OF THE ENDOTRACHEAL TUBE, THE CUFF LEAKED AIR. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 24815524 ↗
  • MalfunctionApr 7, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE CUFF OF THE ENDOTRACHEAL TUBE HAD AN AIR LEAK DURING USE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 24811817 ↗
  • InjuryApr 1, 2026Gas/Air Leak

    IT WAS REPORTED THAT DURING INTUBATION IT WAS FOUND THAT THE CUFF OF THE ENDOTRACHEAL TUBE WAS LEAKING AIR , WITH LOW VENTILATOR TIDAL VOLUME AND AUDIBLE AIRWAY LEAK. INTUBATION WAS IMMEDIATELY DISCONTINUED AND A NEW ENDOTRACHEAL TUBE WAS USED INSTEAD.

    Patient problems: Hypoventilation; Unintended Extubation

    MAUDE report 24761526 ↗

FDA clearance and classification

Cleared through Special 510(k) K090352 on Apr 3, 2009 (decision: Substantially Equivalent). Submitted by Covidien as “Taperguard Endotracheal Tubes, Taperguard Evac Endotracheal Tubes”. The FDA review took 51 days. FDA record ↗

The FDA regulates product code BTR (tube, tracheal (w/wo connector)) under 21 CFR 868.5730 as a Class II device. 21 CFR 868.5730 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Shiley 18875

Has the Shiley 18875 been recalled?

No FDA recall record in our data names this exact device identifier (DI 40884521032539), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Shiley 18875?

The FDA's MAUDE database lists 568 adverse event reports for this device, received between Oct 27, 2017 and Sep 21, 2026. In those reports, the most common specific device problems are “Gas/Air Leak” (269), “Use of Device Problem” (174) and “Defective Component” (38); the most common patient problems are “Unintended Extubation” (272), “Extubate” (61) and “Hypoventilation” (34). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Shiley 18875?

4 reports are classified as deaths, 158 as injuries and 406 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Shiley 18875 FDA approved?

It was cleared through Special 510(k) K090352 on Apr 3, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Shiley 18875 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Shiley 18875, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand Shiley, model 18875). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Tube, tracheal (w/wo connector) devices