Shiley SSV (UDI 40884522005365)Recalls & reported problems

Covidien LP · Speaking Valve · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 40884522005365), received between Nov 29, 2018 and Nov 14, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem3
  • Ambient Noise Problem1
  • Material Integrity Problem1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Unintended Extubation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, DURING USE, THE SPEAKING VALVE WAS NOTICED A HOLE, WHEREIN THE PATIENT CAN BREATHE THROUGH. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20683222 ↗
  • MalfunctionOct 26, 2022Connection Problem; Connection Problem

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE PRODUCT WAS INSTALLED WITH INNER CANNULA IN PLACE AND WAS NOT STAYING ON THE TRACHEOSTOMY TUBE. THE SPEAKING VALVE WAS NOT USE AT ALL AS IT DIDN¿T FIT AND IT WAS FALLING OFF AS IT WAS TOO LOOSE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15672886 ↗
  • MalfunctionOct 3, 2022Connection Problem; Connection Problem

    ACCORDING TO THE REPORTER, DURING USE, IT WAS STATED THAT WHEN THE PATIENT COUGH'S THE DEVICE WOULD COME OFF THE TRACH. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 15528512 ↗
  • MalfunctionSep 26, 2022Connection Problem; Connection Problem

    ACCORDING TO THE REPORTER, DURING USE, THE SPEAKING VALVE WAS UNABLE TO SIT ON THE TRACH WITH THE INNER CANNULA IN PLACE. THE DEVICE HAD TO BE CHANGED TO A NEW ONE. MULTIPLE BRAND-NEW SPEAKING VALVES AND INNER CANNULAS WERE ATTEMPTED. THE SPEAKING VALVES WORKED FINE WITH OTHER TUBES. THE PROBLEM SEEMS TO BE AN ISSUE WITH THE TAPERING OF THE 15-MM TRACHEAL CONNECTION, WHICH ALLOWS THE SPEAKING VALVE TO SLIDE TOO FAR AND THE INNER CANNULA TO HIT THE BOTTOM OF THE SPEAKING VALVE, PREVENTING SEATING. THE DEVICE WAS ABLE TO CONNECT TO THE VENTILATOR CIRCUIT, BUT THE SPEAKING VALVE WAS LOOSE TO THE POINT OF FALLING OFF WITH BREATHING. THE PATIENT WAS BREATHING SPONTANEOUSLY. HOWEVER, THE PATIENT WOULD BE MECHANICALLY VENTILATED AT OTHER TIMES DURING THE DAY AS THE PATIENT WAS PARTICIPATING IN TRACH HOOD TRIALS. THE PATIENT WAS LIKELY RECEIVING AEROSOLIZED MEDICATIONS PRIMARILY WHILE ON THE VENTILATOR, AND NOT WITH THE TRACH HOOD OR SPEAKING VALVE IN PLACE.

    Patient problems: Unintended Extubation

    MAUDE report 15486458 ↗
  • MalfunctionJun 11, 2021Ambient Noise Problem; Ambient Noise Problem

    ACCORDING TO THE REPORTER, THE PRODUCT HAD A CLICKING NOISE DURING USE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11987841 ↗
  • MalfunctionNov 29, 2018Material Integrity Problem; Material Integrity Problem

    ACCORDING TO THE REPORTER, THE SPEAKING VALVE SHOOTS ACROSS THE ROOM WHEN PATIENT COUGHS. THE PATIENT WANTS TO KNOW IF THERE IS A WAY TO SECURE THE TRACH ACCESSORY. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8114038 ↗

Other Tube, tracheostomy (w/wo connector) devices

All Tube, tracheostomy (w/wo connector) devices →