Shiley SSVO (UDI 40884522005372)Recalls & reported problems
Covidien LP · Speaking Valve O2 Port · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 40884522005372), received between May 15, 2018 and Nov 6, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Noise, Audible” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Noise, Audible3
- Defective Component2
- Human-Device Interface Problem2
- Material Split, Cut or Torn1
- Mechanical Problem1
- Physical Resistance/Sticking1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- Unintended Extubation2
- Dyspnea1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2024Mechanical Problem; Defective Component; Physical Resistance/Sticking; Mechanical Problem; Defective Component; Physical Resistance/Sticking
ACCORDING TO THE REPORTER, DURING USE, THE SPEAKING VALVE COULD NOT BE DISCONNECTED FROM THE TRACHEOSTOMY, IT PRESENTED A COLLAPSED APPEARANCE AND NO VALVE DIAPHRAGM MOVEMENT DURING INSPIRATION. THE SPEAKING VALVE WAS REMOVED WITH DIFFICULTY USING EXTERNAL PROTECTION AND MANAGED TO BREATHE INDEPENDENTLY, BUT IT BECAME MORE SECRETIVE, WHICH MADE VENTILATION DIFFICULT AND MADE IT AN EMERGENCY TO REMOVE THE PROBE SO THAT HE COULD BE VACUUMED SAFELY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20620706 ↗MalfunctionJun 14, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING USE, THE OXYGEN INLET CAP OF THE SPEAKING VALVE WAS LOOSED AND DETACHED EASILY WHEN THE PATIENT COUGHED AND DURING VALVE DEVELOPMENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19542649 ↗MalfunctionJan 23, 2023Human-Device Interface Problem; Human-Device Interface Problem
ACCORDING THE REPORTER, DEVICE WAS IN USED FOR TWO DAYS, DURING USE, THE FILM MAKES A NOISE EVERY TIME THE PATIENT EXHALES. THE FILM/MEMBRANE MOVED AND TOUCHED THE VALVE STRUCTURE CAUSED THE NOISE. THE NOISE BOTHERED THE PATIENT TO THE POINT OF PREVENTED FROM FALLING ASLEEP/SLEEP. WHEN THE PATIENT IS LYING DOWN, THE NOISE IS GREATER. THERE WAS NO CHANGE IN THE WAY OF USE, NOR IN THE PATIENT'S CONDITIONS, BUT THE DURABILITY WAS EXTREMELY REDUCED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation
MAUDE report 16225733 ↗MalfunctionJan 23, 2023Human-Device Interface Problem; Human-Device Interface Problem
ACCORDING THE REPORTER, DURING USE, THE FILM MAKES A NOISE EVERY TIME THE PATIENT EXHALES. THE FILM/MEMBRANE MOVED AND TOUCHED THE VALVE STRUCTURE CAUSED THE NOISE. THE NOISE BOTHERED THE PATIENT TO THE POINT OF PREVENTED FROM FALLING ASLEEP/SLEEP. WHEN THE PATIENT WAS LAY DOWN, THE NOISE WAS GREATER. THERE WAS NO CHANGE IN THE WAY OF USE, NOR IN THE PATIENT'S CONDITIONS, BUT THE DURABILITY WAS EXTREMELY REDUCED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation
MAUDE report 16225501 ↗MalfunctionFeb 23, 2022Use of Device Problem; Defective Component; Use of Device Problem; Defective Component
ACCORDING TO THE REPORTER, THE PILOT VALVE HAD A DIMENSIONAL ISSUE AS THE SOUND WAS NOT COMING OUT CORRECTLY, THE VOICE WAS LOW. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13585206 ↗MalfunctionDec 21, 2018Noise, Audible; Noise, Audible
ACCORDING TO THE REPORTER, THE DEVICE WAS A NEWLY OPENED ONE AND IT GAVE OUT WEIRD NOISE WHEN ATTACHED TO THE PATIENT'S TRACHEOSTOMY TUBE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8192008 ↗InjuryNov 2, 2018Noise, Audible; Noise, Audible
ACCORDING TO THE REPORTER, THE UNITS WERE GENERATING A 'SSSSST" SOUND WHEN IT'S BEING USED. THE CUSTOMER REPORTED THAT THE PATIENT HAD DIFFICULTY BREATHING WHEN USING THE DEVICE.
Patient problems: Dyspnea
MAUDE report 8032618 ↗MalfunctionMay 15, 2018Noise, Audible; Noise, Audible
ACCORDING TO THE REPORTER, THE DEVICE'S SPEAKING VALVE PRODUCED A SOUND WHEN USED. THE CUSTOMER REPORTED THAT PATIENT HAD NO INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7517072 ↗
Other Tube, tracheostomy (w/wo connector) devices
- Portex 101/860/080CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)188 reports
- Portex 101/860/075CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)115 reports
- Bivona 60SP040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall20 reports
- Bivona 60SP050Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall13 reports
- Bivona 60P040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall12 reports
- Portex 101/860/070CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)60 reports
- Bivona 60SP045Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall8 reports
- Bivona 60SN040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.