BD MiniDraw™ H&H Capillary Blood Collection Tube 366603Recalls & reported problems
Becton, Dickinson And Company · BD MiniDraw™ H&H Capillary Blood Collection Tube · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This BD MiniDraw capillary blood collection tube is a small tube used to collect blood samples with an anticoagulant. Its 141 reports mostly describe samples that clotted or showed platelet clumping. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 141
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 141 adverse event reports that list this exact device (DI 50382903666039), received between Jun 8, 2024 and Jul 27, 2026 — classified as 141 malfunctions.
The most frequently reported specific device problem is “Coagulation in Device or Device Ingredient” (91 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Coagulation in Device or Device Ingredient91
- Clumping in Device or Device Ingredient54
- Improper Chemical Reaction30
- Incorrect, Inadequate or Imprecise Result or Readings1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions141
Recent adverse event reports
The 25 most recent of 141 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2026Improper Chemical Reaction
Patient 28 of 30:It was reported while using BD MiniDraw¿ H&H Capillary Blood Collection Tube, 1 sample exhibited platelet clumping. There was no health impact or consequences reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50045810 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032715 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032711 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032181 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032177 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, CLOTTING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032174 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032172 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032169 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032164 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032155 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032151 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032146 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032112 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032106 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032099 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032060 ↗MalfunctionApr 29, 2026Improper Chemical Reaction
IT WAS REPORTED WHILE USING BD MINIDRAW H&H CAPILLARY BLOOD COLLECTION TUBE, PLATELET CLUMPING WAS SEEN WITH ONE PATIENT SAMPLE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25031978 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient
PATIENT 2 OF 2: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 SAMPLE WAS CLOTTED. THE SAMPLE WAS RECOLLECTED IF THE PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254252 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 9 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254251 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 1 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254250 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 8 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254249 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 7 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254248 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 2 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254247 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 4 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254238 ↗MalfunctionFeb 3, 2026Coagulation in Device or Device Ingredient; Improper Chemical Reaction
PATIENT 5 OF 9: IT WAS REPORTED WHILE USING BD MINIDRAW¿ H&H CAPILLARY BLOOD COLLECTION TUBE, 1 PATIENT SAMPLE HAD PLATELET AGGREGATES. A RECOLLECT WAS PERFORMED IF PATIENT AGREED. THERE WAS NO OTHER HEALTH IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254237 ↗
FDA clearance and classification
Cleared through 510(k) K230493 on Nov 27, 2023 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “BD MiniDraw Capillary Blood Collection System with BD MiniDraw H&H Capillary Blood Collection Tube”. The FDA review took 277 days. FDA record ↗
The FDA regulates product code GIM (tubes, vacuum sample, with anticoagulant) under 21 CFR 862.1675 as a Class II device. 21 CFR 862.1675 ↗
Common questions about the BD MiniDraw™ H&H Capillary Blood Collection Tube 366603
Has the BD MiniDraw™ H&H Capillary Blood Collection Tube 366603 been recalled?
No FDA recall record in our data names this exact device identifier (DI 50382903666039), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the BD MiniDraw™ H&H Capillary Blood Collection Tube 366603?
The FDA's MAUDE database lists 141 adverse event reports for this device, received between Jun 8, 2024 and Jul 27, 2026. In those reports, the most common specific device problems are “Coagulation in Device or Device Ingredient” (91), “Clumping in Device or Device Ingredient” (54) and “Improper Chemical Reaction” (30). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD MiniDraw™ H&H Capillary Blood Collection Tube 366603?
None of the reports are classified as deaths, 0 as injuries and 141 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD MiniDraw™ H&H Capillary Blood Collection Tube 366603 FDA approved?
It was cleared through 510(k) K230493 on Nov 27, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BD MiniDraw™ H&H Capillary Blood Collection Tube 366603, and is it still sold?
The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD MiniDraw™ H&H Capillary Blood Collection Tube, model 366603). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Tubes, vacuum sample, with anticoagulant devices
- Vacuette 454322Greiner Bio-One GmbH · Tubes, vacuum sample, with anticoagulant3 recalls0 reports
- BD Vacutainer® K2 EDTA (K2E) Plus Blood Collection Tubes 366643Becton, Dickinson U.K. Limited · Tubes, vacuum sample, with anticoagulant124 reports
- Vacuette 454332Greiner Bio-One GmbH · Tubes, vacuum sample, with anticoagulant1 recall0 reports
- Cyto-Chex BCT 218980Streck, Inc. · Tubes, Vacuum Sample, With Anticoagulant1 recall0 reports
- Vacuette 454334Greiner Bio-One GmbH · Tubes, vacuum sample, with anticoagulant1 recall0 reports
- Cyto-Chex BCT 213559Streck, Inc. · Tubes, Vacuum Sample, With Anticoagulant1 recall0 reports
- BD MiniDraw™ Finger Sleeve, Extra-Large 366616Becton, Dickinson And Company · Tubes, vacuum sample, with anticoagulant25 reports
- BD MiniDraw™ Finger Sleeve, Large 366615Becton, Dickinson And Company · Tubes, vacuum sample, with anticoagulant21 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.