GUARDIAN 65651-220Recalls & reported problems
Cardinal Health 200, LLC · 2000cc Guardian II · Class 2
All models of GUARDIAN (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 50630140152540), received between Mar 5, 2018 and Jan 5, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Suction Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Suction Problem2
- Burst Container or Vessel1
- Fracture1
- Suction Failure1
Patient problems
- Pain2
- Vomiting2
- Exposure to Body Fluids1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2022Burst Container or Vessel; Suction Failure; Burst Container or Vessel; Suction Failure
CUSTOMER REPORTED THE PROCEDURE WAS A PLANNED AIRWAY SUCTIONING. THE CANNISTER WITH SMALL AMOUNTS OF ORAL SECRETIONS SHATTERED IN MULTIPLE LOCATIONS AFTER COMPLETING ROUTINE SUCTIONING OF THE PATIENT. THERE WAS A LOUD BANG HEARD WHEN THE CANISTER BROKE. THE SUCTION CANISTER WAS CONTAMINATED WITH ORAL SECRETIONS WHEN THE PRODUCT LOST ITS SUCTION CAPABILITY. THERE WAS NO INJURY OR ADVERSE EFFECT TO THE PATIENT BUT THERE WAS BODILY FLUID EXPOSURE AS A PORTION OF THE CONTAINER DID CONTACT THE INTACT SKIN OF THE PRIMARY RN. THERE WERE NO ADVERSE EFFECTS TO RN. AT PRESENT RN IS DOING FINE.
Patient problems: Exposure to Body Fluids
MAUDE report 13172737 ↗InjuryMar 5, 2018Suction Problem
BASED ON INFORMATION RECEIVED FROM THE CUSTOMER THE CANISTERS PLASTIC WAS FRACTURED AND DID NOT PROVIDE ADEQUATE SUCTION FOR THE PATIENT. PATIENT HAD A NG TUBE TO LOW INTERMITTENT WALL SUCTION. REPORTEDLY DUE TO LACK OF POOR SUCTION THE PATIENT EXPERIENCED VOMITING, RETCHING AND PAIN. PATIENT WAS DISCHARGED ON (B)(6) 2018.
Patient problems: Pain; Vomiting
MAUDE report 7314780 ↗InjuryMar 5, 2018Fracture; Suction Problem
BASED ON INFORMATION RECEIVED FROM THE CUSTOMER THE CANISTERS PLASTIC WAS FRACTURED AND DID NOT PROVIDE ADEQUATE SUCTION FOR THE PATIENT. PATIENT HAD A NG TUBE SET TO LOW INTERMITTENT WALL SUCTION. REPORTEDLY DUE TO LACK OF POOR SUCTION THE PATIENT EXPERIENCED VOMITING, RETCHING AND PAIN. PATIENT WAS DISCHARGED ON (B)(6) 2018.
Patient problems: Pain; Vomiting
MAUDE report 7314771 ↗
Other APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices
- StrykeFlow 0250070520Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED2 recalls250 reports
- StrykeFlow 0250070505Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall3 reports
- Sentinel Seal 8888571562Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Aqua-Seal 8888571299Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- spineX BONE DUST COLLECTOR S-4500Oto-Med, Inc. · Apparatus, Suction, Operating-Room, Wall Vacuum Powered1 recall0 reports
- Aqua-Seal 8888571406Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Altitude 8888571370Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- CRD 65651-517Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page