Level 1 EQ-5000Recalls & reported problems
Icu Medical, Inc. · Class 2
Quick summary
This ICU Medical Level 1 EQ-5000 is a warming system used to keep patients at a controlled temperature. Most of the 46 reports describe temperature control problems — the unit failing to heat, running too hot, shutting down unexpectedly, or sounding alarms — and a few mention power cutting out mid-infusion, a broken connector, or electric leakage. Five injuries were reported, including second-degree burns, redness, and blistering, and no deaths. There are no recalls. Reports fell sharply, from 21 in the earlier 12 months to 1 in the most recent.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 46
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 50695085850005), received between May 13, 2021 and Nov 13, 2025 — classified as 5 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Temperature Problem” (12 reports).
1 other recall cover devices cleared under the same FDA submission (K011907); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Temperature Problem12
- Insufficient Heating11
- Device Alarm System6
- Excessive Heating4
- Unexpected Shutdown3
- Break2
- Device Displays Incorrect Message2
- Erratic or Intermittent Display2
Patient problems
- No Clinical Signs, Symptoms or Conditions41
- Insufficient Information3
- Partial thickness (Second Degree) Burn2
- Blister1
- Burn(s)1
- Erythema1
- Skin Burning Sensation1
Recent adverse event reports
The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2025Loss of Power; Loss of Power
IT WAS REPORTED THAT THE POWER SUPPLY CUT OFF MIDWAY AND FAILED TO TURN ON AFTERWARD. THIS OCCURRED DURING INFUSION AT THE FACILITY. THERE WAS A PATIENT INVOLVEMENT, BUT NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23546854 ↗MalfunctionAug 14, 2025Material Fragmentation; Power Problem; Material Fragmentation; Power Problem
IT WAS REPORTED THAT THE POWER SUPPLY WAS UNSTABLE, INTERMITTENTLY TURNING ON AND OFF DURING OPERATION. THIS OCCURRED DURING PRIMING AT THE FACILITY. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22811134 ↗MalfunctionJul 22, 2025Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THE CONNECTOR IS BROKEN AND SOMETIMES, AIR DOES NOT BLOW OUT. THE EVENT OCCURRED BEFORE PATIENT USE AT THE FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569536 ↗MalfunctionJul 14, 2025Temperature Problem; Excessive Heating; Temperature Problem; Excessive Heating
IT WAS REPORTED THAT THE DEVICE EXHIBITED A HIGH TEMPERATURE. THERE WAS NO PATIENT INVOLVEMENT, NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22496138 ↗MalfunctionApr 29, 2025Temperature Problem; Temperature Problem
IT WAS REPORTED THAT A LOW TEMPERATURE ALARM OCCURRED. THIS OCCURRED DURING INFUSION AT THE FACILITY. THERE WAS A PATIENT INVOLVEMENT, BUT NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926355 ↗MalfunctionApr 7, 2025Insufficient Heating; Temperature Problem; Insufficient Heating; Temperature Problem
IT WAS REPORTED THAT THE TEMPERATURE DID NOT RISE. EVENT OCCURRED BEFORE PATIENT USE/DURING TESTING AT THE FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARMED REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21788563 ↗MalfunctionMar 31, 2025Leak/Splash; Electrical Power Problem; Leak/Splash; Electrical Power Problem
IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRIC LEAKAGE. THIS OCCURRED DURING INFUSION AT THE FACILITY. THERE WAS A PATIENT INVOLVEMENT, BUT NO HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21734293 ↗MalfunctionMar 26, 2025Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE SENSOR HAS A FALSE CONTACT, SO THE GENERATOR DOES NOT HEAT UP THE AIR. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21698480 ↗MalfunctionMar 18, 2025Electrical /Electronic Property Problem; Device Emits Odor; Electrical /Electronic Property Problem; Device Emits Odor
IT WAS REPORTED THAT THE POWER DOES NOT TURN ON. THERE'S A STRANGE SMELL. EVENT OCCURRED BEFORE PATIENT USE/DURING PRIMING AT THE FACILITY. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARMED REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21631527 ↗MalfunctionFeb 19, 2025Erratic or Intermittent Display; Ambient Temperature Problem; Erratic or Intermittent Display; Ambient Temperature Problem
IT WAS REPORTED THAT EVEN WHEN THE POWER WAS TURNED ON, THE AIR WAS NOT BLOWING, AND VARIOUS NUMBERS WERE SEEN ON THE TEMPERATURE DISPLAY. THERE WERE NO PATIENT HARM AND/OR ADVERSE EVENTS REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21417090 ↗InjuryFeb 13, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT PATIENT COMPLAINED EXPERIENCING OVERHEATING AFTER BEING COVERED WITH A FULL BODY BLANKET. AFTER THE HEATER WAS STOPPED, THE HOSE APPEARED TO BE MUCH HOTTER THAN USUAL, EVEN THOUGH THE DIAL SHOWED 25° AND THE SETPOINT WAS 44°. THE PATIENT PRESENTED REDNESS ON BOTH LEGS AND COMPLAINED OF PAIN NEAR HER RIGHT ANKLE. AFTER THE INCIDENT, THE PATIENT WAS REFERRED TO A DERMATOLOGIST AND A PLASTIC SURGERY SPECIALIST FOR WITH FURTHER TREATMENT AND BANDAGING.
Patient problems: Burn(s); Skin Burning Sensation
MAUDE report 21379763 ↗MalfunctionFeb 12, 2025Insufficient Heating; Insufficient Heating
IT WAS REPORTED THAT THE DEVICE TEMPERATURE WAS NOT RISING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21368871 ↗MalfunctionFeb 11, 2025Temperature Problem; Temperature Problem
IT WAS REPORTED THAT THE TEMPERATURE WAS NOT RISING. THERE WERE PATIENT HARM OR ADVERSE EVENT REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21357359 ↗MalfunctionFeb 11, 2025Temperature Problem; Temperature Problem
IT WAS STATED IN THE REPORT THAT THE TEMPERATURE WAS UNSTABLE. EVENT OCCURRED DURING PATIENT USE AT THE FACILITY. THERE WERE NO PATIENT HARM OR ADVERSE EVENTS REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21357279 ↗MalfunctionJan 24, 2025Circuit Failure; Overheating of Device; Unexpected Shutdown; Excessive Heating; Circuit Failure; Overheating of Device; Unexpected Shutdown; Excessive Heating
IT WAS REPORTED THAT THE WARMER WAS STOPPING AND THAT THE OVER-TEMPERATURE AND THE MAINTENANCE INDICATORS WERE RAPIDLY BLINKING. THE DEVICE WAS BEING USED ON A PATIENT WHILE THIS HAPPENED. NO PATIENT HARM OR MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21229499 ↗MalfunctionJan 15, 2025Break; Insufficient Heating; No Device Output; Break; Insufficient Heating; No Device Output
IT WAS REPORTED THAT "THE TEMPERATURE WON'T RISE, THE EQUIPMENT WON'T WORK, AND THE HOSE IS BROKEN". THERE WAS NO PATIENT INVOLVEMENT AND REPORTER CONFIRMED THAT THERE WERE NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21162256 ↗MalfunctionJan 14, 2025Insufficient Heating; Key or Button Unresponsive/not Working; Insufficient Heating; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE PRODUCT DID NOT PRODUCE WARM AIR. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21152674 ↗MalfunctionJan 10, 2025Temperature Problem; Temperature Problem
IT WAS REPORTED THAT THE TEMPERATURE DOESN'T RISE. EVENT OCCURRED BEFORE PATIENT USE, DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARMED REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21132300 ↗MalfunctionDec 31, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED THAT THE INDICATOR WAS FLASHING. THE FAULT OCCURRED WHILE IN USE WITH THE PATIENT. THERE WAS KNOWN PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21047665 ↗MalfunctionDec 26, 2024Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT THE DEVICE STOPPED WORKING WHILE IN PATIENT USE. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21021049 ↗MalfunctionNov 11, 2024Insufficient Heating; Insufficient Heating
IT WAS REPORTED THAT THE TEMPERATURE DID NOT REACH THE SET VALUE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20652331 ↗MalfunctionNov 5, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE TESTING THE TEMPERATURE OF THE WARMER, INCORRECT READINGS WERE RECEIVED. IN ONE INSTANCE, WHEN SETTING THE TEMPERATURE TO 36 DEGREES CELSIUS, THE DEVICE SHOWED A RECORDING OF 12 DEGREES CELSIUS. THERE WAS UNKNOWN PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20609855 ↗MalfunctionOct 28, 2024Device Alarm System; Erratic or Intermittent Display; Device Alarm System; Erratic or Intermittent Display
IT WAS REPORTED THAT THE TEMPERATURE SHOWS ABNORMAL DISPLAY AND AN ALARM SOUNDS. EVENT OCCURRED WHILE PATIENT USE, DURING PATIENT ADMINISTRATION. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20546735 ↗MalfunctionOct 21, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED THAT THERE WAS A POOR CONNECTION ALARM THAT RANG. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARMED REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20495568 ↗MalfunctionOct 21, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED THAT THE DISCONNECT INDICATOR ALARM SOUNDS. THERE WAS KNOWN PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20494599 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K011907. One submission often covers many models, so check the product and lot details in each record.
- Z-0497-05 · Dec 15, 2004 · TerminatedEquator Forced Air Convective Warming System, Reference Number EQ 5000
Other System, thermal regulating devices
- Arctic Sun® 5000 50000000Medivance, Inc. · System, thermal regulating1 recall3 deaths6,903 reports
- Arctic Sun® 5000 50000000EMedivance, Inc. · System, thermal regulating1 recall2 deaths4,413 reports
- Arctic Sun Stat 60000000Medivance, Inc. · System, thermal regulating1 recall1,613 reports
- Arctic Sun® 5000 50000000LMedivance, Inc. · System, thermal regulating1 recall388 reports
- Arctic Sun® Temperature Management System, ArcticGel™ Pads 31802Medivance, Inc. · System, thermal regulating1 recall385 reports
- Altrix 8001Stryker Corporation · System, thermal regulating1 recall160 reports
- Arctic Sun Stat 60000000LMedivance, Inc. · System, thermal regulating1 recall104 reports
- Hillrom powered by Augustine Surgical HD 2083516Augustine Temperature Management LLC · System, Thermal Regulating2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.