Ranger™ 24355 (UDI 50707387781441)Recalls & reported problems

3m Company · 3M™ Ranger™ Fluid Warming Set, High Flow, 10 per case 24355 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This 3M Ranger high flow fluid warming set has 107 reports, with no deaths and no injuries. Most reports describe fluid or blood leaking at connections, such as a connector coming loose or a drip chamber leaking during priming. The reports note that no patient or staff injury was alleged. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
107
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 107 adverse event reports that list this exact device (DI 50707387781441), received between Sep 22, 2017 and Feb 13, 2023 — classified as 107 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (104 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak104
  • Appropriate Device Problem Term/Code Not Available3
  • Connection Problem1
  • Crack1
  • Disconnection1
  • Increase in Pressure1
  • Material Integrity Problem1
  • Unintended Compatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions80
  • Not Applicable21
  • Insufficient Information5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 107 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL REPORTED THAT UPON THE INITIAL INFUSION OF BLOOD VIA A 3M¿ RANGER¿ PRESSURE INFUSER, 14510 WITH THE USE OF 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX9090, BLOOD LEAKED AT THE LEUR CONNECTION. UPON TRANSFERRING THE PATIENT FROM THE OPERATING TABLE TO THE BED, THE COUPLING ALLEGEDLY BECAME LOOSE AND RELEASED QUICKLY RESULTING IN A LARGE OPENING IN THE LUMEN WITH BLOOD LEAKAGE. NO ISSUES OCCURRED WITH THE 3M¿ RANGER¿ PRESSURE INFUSER, AND NO PATIENT OR STAFF INJURY WAS ALLEGED. THE ADMINISTRATION OF FLUIDS WAS RESTARTED WITH A NEW CASSETTE. NO MEDICAL INTERVENTIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16362795 ↗
  • MalfunctionFeb 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL REPORTED WHEN PURGING THE 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX9090 PIPELINE, THE COUPLING BECAME LOOSE, AND BLOOD LEAKED AT THE LEUR CONNECTION. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16362496 ↗
  • MalfunctionFeb 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED AT THE LEUR CONNECTION PROXIMAL TO THE PATIENT END WHILE PRIMING. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16283190 ↗
  • MalfunctionFeb 1, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX9181 LEAKED.. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16276844 ↗
  • MalfunctionJan 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX8986 LEAKED AT THE DRIP CHAMBER. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16196494 ↗
  • MalfunctionJan 17, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX8986 LEAKED AT THE DRIP CHAMBER. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16185545 ↗
  • MalfunctionJan 16, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED AT THE DRIP CHAMBER WHILE PRIMING WITH LACTATED RINGERS. DEVICE WAS NOT CONNECTED TO A PATIENT. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16180005 ↗
  • MalfunctionJan 16, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE INLET LINE CAME OUT OF THE SECOND RINGER'S LACTATE BAG AFTER USE OF 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355). NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16179935 ↗
  • MalfunctionJan 5, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED AT THE HEAT EXCHANGER. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16109275 ↗
  • MalfunctionDec 31, 2022Fluid/Blood Leak; Fluid/Blood Leak

    REPORT RECEIVED VIA MEDWATCH FORM MW (B)(4). A HOSPITAL REPORTED THAT WHEN INFUSING FFP [FRESH FROZEN PLASMA] AND BLOOD WITH USE OF 3M¿ RANGER¿ FLUID WARMING SETS, HIGH FLOW (MODEL 24355) LOT HX9190, THE AIR CHAMBER FROM TWO WARMERS ALLEGEDLY "EXPLODED," REPORTEDLY CAUSING THE FFP AND BLOOD TO SPILL ONTO THE FLOOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16081284 ↗
  • MalfunctionDec 15, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A 3M¿ RANGER¿ HIGH FLOW WARMING SETS (MODEL 24355) LEAKED. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15992451 ↗
  • MalfunctionNov 18, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX9190 LEAKED AT THE BUBBLE TRAP ON MULTIPLE OCCASIONS. IT WAS INDICATED THAT A PRESSURE DEVICE WAS USED. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15828717 ↗
  • MalfunctionOct 28, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE 3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW (MODEL 24370) LOT HM9047 LEAKED. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15691126 ↗
  • MalfunctionOct 25, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED DURING USE. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15669174 ↗
  • MalfunctionOct 25, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE 3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW (MODEL 24370) LOT HM9196 LEAKED AT THE BUBBLE TRAP. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15669134 ↗
  • MalfunctionOct 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED AT THE BUBBLE TRAP WHILE IN USE. IT WAS INDICATED THAT A PRESSURE DEVICE WAS USED. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15661199 ↗
  • MalfunctionOct 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX9181 LEAKED DURING USE. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15659724 ↗
  • MalfunctionOct 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED THE 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LOT HX9090 CASSETTE SLIPPED OUT 2-3CM AND LEAKED BLOOD AT THE FLUID WARMING CASSETTE/HEAT EXCHANGER. NO PATIENT OR STAFF INJURY WAS ALLEGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15659382 ↗
  • MalfunctionOct 21, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED MULTIPLE 3M¿ RANGER¿ HIGH FLOW WARMING SETS (MODEL 24355) LEAKED DURING USE. LOCATION OF LEAK WAS IDENTIFIED TO BE THE BUBBLE TRAP AND AT THE LEUR CONNECTIONS. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15653355 ↗
  • MalfunctionOct 19, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED MULTIPLE 3M¿ RANGER¿ HIGH FLOW WARMING SETS (MODEL 24355) LEAKED AT THE BUBBLE TRAP. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15632159 ↗
  • MalfunctionOct 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED MULTIPLE SETS OF 3M¿ RANGER¿ HIGH FLOW WARMING SETS (MODEL 24355) LEAKED AT THE BUBBLE TRAP WHILE USING. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15577722 ↗
  • MalfunctionOct 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED MULTIPLE SETS OF 3M¿ RANGER¿ HIGH FLOW WARMING SETS (MODEL 24355) LEAKED AT THE BUBBLE TRAP WHILE USING. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15577610 ↗
  • MalfunctionSep 30, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED BLOOD WHILE IN USE. LOCATION OF LEAK WAS NOT IDENTIFIED. NO PATIENT OR STAFF INJURY WAS ALLEGED. INFUSION WAS DISCONTINUED AND TUBING WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15513999 ↗
  • MalfunctionSep 30, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED MULTIPLE SETS OF 3M¿ RANGER¿ HIGH FLOW WARMING SETS (MODEL 24355) LEAKED AT THE BUBBLE TRAP WHILE USING A MANUAL PUMP. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15509062 ↗
  • MalfunctionSep 29, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A HOSPITAL ALLEGED 3M¿ RANGER¿ HIGH FLOW WARMING SET (MODEL 24355) LEAKED AT THE HEAT EXCHANGER WHILE IN USE. NO PATIENT OR STAFF INJURY WAS ALLEGED. NO MEDICAL INTERVENTIONS WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15506729 ↗

Other Warmer, blood, non-electromagnetic radiation devices

All Warmer, blood, non-electromagnetic radiation devices →