CS 3600Recalls & reported problems
Carestream Dental LLC · CS 3600 I/O 3D SCANNER · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 60192156568748), received between Feb 23, 2021 and Apr 12, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Defective Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Defective Component2
- Material Separation2
Patient problems
- Foreign Body In Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2021Material Separation; Defective Component; Material Separation; Defective Component
THIS IS A CORRECTED REPORT. ORIGINAL MDR NUMBER USED CFN FOR CARESTREAM HEALTH. THE ORIGINAL MDR NUMBER WAS: 131703-2021-00001. THE CORRECTED MDR NUMBER USES THE CORRECT FEI FOR CARESTREAM DENTAL LLC: 3002522649-2021-00001. WHEN USING THE CS 3600 ACCESSORY: CS 3600 SINGLE USE NORMAL TIP, THE MIRROR CAN BE DETACHED FROM THE UNIT. WHEN THIS EVENT OCCURRED, THE MIRROR FELL FROM THE UNIT INTO THE PATIENT'S MOUTH. THE PATIENT WAS ABLE TO REMOVE THE MIRROR AND THE PATIENT DID NOT SUSTAIN ANY INJURIES. THE SIZE OF THE MIRROR (22MMX13MMX1.5MM) MAKES IT POSSIBLE FOR THE MIRROR TO BE SWALLOWED OR ASPIRATED. IF THE MIRROR IS SWALLOWED OR ASPIRATED, SURGICAL INTERVENTION WOULD BE REQUIRED TO REMOVE THE MIRROR.
Patient problems: Foreign Body In Patient
MAUDE report 11648699 ↗MalfunctionFeb 23, 2021Material Separation; Defective Component; Material Separation; Defective Component
WHEN USING THE CS 3600 ACCESSORY: CS 3600 SINGLE USE NORMAL TIP, THE MIRROR CAN BE DETACHED FROM THE UNIT. WHEN THIS EVENT OCCURRED, THE MIRROR FELL FROM THE UNIT INTO THE PATIENT'S MOUTH. THE PATIENT WAS ABLE TO REMOVE THE MIRROR AND THE PATIENT DID NOT SUSTAIN ANY INJURIES. THE SIZE OF THE MIRROR (22MMX13MMX1.5MM) MAKES IT POSSIBLE FOR THE MIRROR TO BE SWALLOWED OR ASPIRATED. IF THE MIRROR IS SWALLOWED OR ASPIRATED, SURGICAL INTERVENTION WOULD BE REQUIRED TO REMOVE THE MIRROR.
Patient problems: Foreign Body In Patient
MAUDE report 11366583 ↗
Other System, Optical Impression, Computer Assisted Design And Manufacturing (Cad/Cam) Of Dental Restorations devices
- iTero Element 2 202965Align Technology, Inc. · System, Optical Impression, Computer Assisted Design And Manufacturing (Cad/Cam) Of Dental Restorations2 reports
- iTero Element Flex 202964Align Technology, Inc. · System, Optical Impression, Computer Assisted Design And Manufacturing (Cad/Cam) Of Dental Restorations1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.