Skip to content

Protek 1-519-3230CIRecalls & reported problems

Aspen Surgical Products, Inc. · Cone; Endocavity Probe Cover; 3.5cm x 20cm; Latex; rolled; 4 bands; sterile; 1/pouch; 40/box; (934412) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 60841912100882), received between May 13, 2026 and May 13, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Perforation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Perforation1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2026Material Puncture/Hole; Material Perforation

    COMPLAINANT ALLEGES "THE ITEM # IS 934412 AND OUR CLINICAL STAFF HAS BEEN FINDING HOLES IN THEM WHEN THEY PUT THE PROBE COVER OVER THE PROBE. YOU CAN'T SEE THE HOLES UNTIL THE COVER IS STRETCHED OUT, IF THAT MAKES SENSE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25164579 ↗

Other Drape, Surgical devices

All Drape, Surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page