Skip to content

QC Sets and Panels: Helix Elite™ 8234Recalls & reported problems

Microbiologics Inc. · QC Sets and Panels: Helix Elite™ Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet) · Class 1

All models of QC Sets and Panels: Helix Elite (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 20, 2025Class IIOpen, Classified

    Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

    Z-1163-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jan 20, 2025Class IIOpen, Classified

    Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

    Z-1164-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Multi-analyte controls unassayed devices

All Multi-analyte controls unassayed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page