BIS 186-0106-PHRecalls & reported problems
Covidien LP · Quatro Sensor Kit, (25) 186-0106 · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 90884521657515), received between Jan 9, 2023 and May 21, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K072286, K143506); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem2
- Defective Component1
Patient problems
- Erythema1
- Inflammation1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
ACCORDING TO THE LITERATURE, THE SENSOR CAUSED A REVERSIBLE SKIN LESION IN THE PATIENT UNDERGOING SURGERY. THE ELECTRODES WERE APPLIED, THE PATIENT WAS POSITIONED, AND PRESSURE AREAS WERE CAREFULLY PADDED AND USED FOR MONITORING DURING THIS UNEVENTFUL 2.5-HOUR PROCEDURE. AFTER REMOVING THE ELECTRODES, WELL-DEFINED, ERYTHEMATOUS MARKINGS WERE OBSERVED ON THE SKIN, CORRESPONDING TO THE ELECTRODE PLACEMENT SITES. THESE MARKINGS WERE CLASSIFIED AS NON-BLISTERING, NON-URTICARIAL, AND BLANCHING; THERE WERE NO OTHER SIGNS OF A SYSTEMATIC ALLERGIC REACTION. THE MARKS WERE NOT TYPICAL OF ALLERGIC CONTACT DERMATITIS BUT MORE LIKELY DUE TO AN IRRITANT AND THE PRESSURE EFFECT OF THE MONITOR. A FOLLOW-UP VISIT THREE WEEKS LATER REVEALED THE NEAR DISAPPEARANCE OF THE MARKS.
Patient problems: Inflammation
MAUDE report 19367215 ↗MalfunctionJan 9, 2023Defective Component; Patient Device Interaction Problem; Defective Component; Patient Device Interaction Problem
ACCORDING TO THE REPORTER, POST OPERATIVELY, SENSOR REDNESS MARKS ON PATIENT'S FOREHEAD. THREE DAYS AFTER THE SURGERY ON NOVEMBER 30, THE PATIENT TOOK A SHOWER ON DECEMBER 3, THE NURSE FOUND REDNESS AROUND THE SENSOR MARKS. THERE WAS NO REDNESS BEFORE THE SHOWER OR AFTER THE OPE.
Patient problems: Erythema; Skin Inflammation/ Irritation
MAUDE report 16122869 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072286, K143506. One submission often covers many models, so check the product and lot details in each record.
- Z-2958-2018 · Jul 26, 2018 · Open, ClassifiedLi-ion battery packs used in BIS Vista and BIS View Monitoring Systems. Product Number/CFN: 186-0208 UDI-Device Identifier (GTIN/UPN): 10884521130319
Other Index-generating electroencephalograph software devices
- BIS 186-1046Covidien LP · Index-generating electroencephalograph software102 reports
- BIS 186-0224-AMSCovidien LP · Index-generating electroencephalograph software1 death41 reports
- BIS 186-0195-AMSCovidien LP · Index-generating electroencephalograph software38 reports
- BIS Module IntelliVue BIS ModulePhilips Medizin Systeme Böblingen GmbH · Index-Generating Electroencephalograph Software15 reports
- BIS 186-0195-NKCovidien LP · Index-generating electroencephalograph software11 reports
- BIS 186-1001-RFBCovidien LP · Index-generating electroencephalograph software6 reports
- BIS PMB4000Covidien LP · Index-generating electroencephalograph software5 reports
- BIS 186-0199-RFBCovidien LP · Index-generating electroencephalograph software5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.