SmartMonitor 2PSL w/ Modem 1028970Recalls & reported problems
Circadiance, LLC · SmartMonitor 2PSL w/ Modem · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI B0361028970), received between Mar 6, 2018 and Nov 16, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K061256); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Loose or Intermittent Connection2
- Communication or Transmission Problem1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
A USER FACILITY REPORTED THAT "ALARM CONNECTOR BROKEN INSIDE ALARM PORT" OCCURRED ON A SMART MONITOR 2PSL MONITOR, MODEL 1028970, SERIAL NUMBER (B)(4). THE ALARM PORT REFERRED TO BY THE CUSTOMER IN THIS COMPLAINT IS A FEMALE JACK THAT ALLOWS THE MONITOR TO BE CONNECTED TO A HOSPITAL NURSE CALL SYSTEM. IN THIS CASE, THE ALARM PORT CONNECTIONS HAVE BEEN DETERMINED TO BE PHYSICALLY BROKEN AT THE POINT WHERE THE JACK IS ATTACHED TO THE MAIN CIRCUIT BOARD. AT LEAST ONE OF THE CONNECTIONS IS STILL INTACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8080958 ↗MalfunctionAug 28, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
A USER FACILITY REPORTED THAT "ALARM CONNECTOR BROKEN INSIDE ALARM PORT" OCCURRED ON A SMART MONITOR 2PSL MONITOR, MODEL 1028970, SERIAL NUMBER (B)(4). THE ALARM PORT REFERRED TO BY THE CUSTOMER IN THIS COMPLAINT IS A FEMALE JACK THAT ALLOWS THE MONITOR TO BE CONNECTED TO A HOSPITAL NURSE CALL SYSTEM. IN THIS CASE, THE ALARM PORT CONNECTIONS HAVE BEEN DETERMINED TO BE PHYSICALLY BROKEN AT THE POINT WHERE THE JACK IS ATTACHED TO THE MAIN CIRCUIT BOARD. AT LEAST ONE OF THE CONNECTIONS IS STILL INTACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7824789 ↗MalfunctionMar 6, 2018Communication or Transmission Problem; Communication or Transmission Problem
SAINT FRANCIS HEALTH SYSTEM REPORTED THAT A SMART MONITOR 2PS, MODEL NUMBER 1028970, SERIAL NUMBER (B)(4) HAD AN AUXILLIARY ALARM INTERFACE THAT WAS NOT WORKING. THE BIOMED REPORTED THAT STAFF INFORMED BIOMED THAT THE ALARM WOULD NOT CALL THE DESK THROUGH THE NURSE CALL SYSTEM. THE BIOMED REPORTED THAT SHE VISITED THE LOCATION WHERE THE MONITOR WAS USED AND VERIFIED THAT ALARMS WERE NOT BEING TRANSMITTED TO THROUGH THE NURSE CALL SYSTEM. THE BIOMED ALSO REPORTED THAT THE PRIMARY ALARM IN THE MONITOR WAS WORKING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7318882 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061256. One submission often covers many models, so check the product and lot details in each record.
- Z-2250-2017 · May 1, 2017 · TerminatedSmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.
- Z-1789-2014 · Apr 29, 2014 · TerminatedPhilips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
- Z-1086-2008 · Oct 15, 2007 · TerminatedAPNEA monitor accessories under the following brand names: a) SM 2 PS/PSL, Prof. Operators Manual, NL-NL model number 1025422; b) SM 2/2 PS Europe Patient Cable, Metal model number
Other Monitor, Apnea, Home Use devices
- SmartMonitor 2PS w/ Modem 1014557Circadiance, LLC · Monitor, Apnea, Home Use18 reports
- SmartMonitor 2PS Hospital Alarm 1030271Circadiance, LLC · Monitor, Apnea, Home Use1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page