Skip to content

ELEOS™ 25001090ERecalls & reported problems

Onkos Surgical, Inc. · ELEOS™ MALE-FEMALE MIDSECTION 90MM · Class 2 · Implant

All models of ELEOS (127)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI B27825001090E0), received between Feb 13, 2020 and Mar 19, 2025 — classified as 10 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

11 other recalls cover devices cleared under the same FDA submission (K161520); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem9
  • Appropriate Device Problem Term/Code Not Available1
  • Loose or Intermittent Connection1

Patient problems

  • Unspecified Infection5
  • Medical device site infection4
  • Insufficient Information3
  • Ulcer1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ONKOS SALES REPRESENTATIVE REPORTED THAT A 58-YEAR-OLD FEMALE PATIENT WITH AN ELEOS DISTAL FEMUR REPLACEMENT UNDERWENT REVISION SURGERY ON (B)(6) 2025 DUE TO ALLEGED PERIPROSTHETIC JOINT INFECTION.

    Patient problems: Medical device site infection

    MAUDE report 21642602 ↗
  • InjuryNov 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY (B)(4), AN ONKOS SALES REPRESENTATIVE, THAT A 32-YEAR-OLD MALE PATIENT WITH AN ELEOS DISTAL FEMUR REPLACEMENT UNDERWENT REVISION SURGERY ON (B)(6) 2024 DUE TO AN ALLEGED PERI-OPERATIVE INFECTION.

    Patient problems: Medical device site infection

    MAUDE report 20789585 ↗
  • InjuryAug 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY (B)(4), AN ONKOS SALES REPRESENTATIVE, THAT AN 81-YEAR-OLD MALE PATIENT WITH AN ELEOS DISTAL FEMUR REPLACEMENT UNDERWENT A REVISION SURGERY DUE TO AN ALLEGED INFECTION ON (B)(6) 2024. THE SURGEON, (B)(4) WASHED OUT THE OPERATIVE JOINT AND REPLACED THE ELEOS DEVICES LISTED ABOVE.

    Patient problems: Medical device site infection

    MAUDE report 19911449 ↗
  • InjuryJul 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY (B)(6), AN ONKOS SALES REPRESENTATIVE, THAT AN 80-YEAR-OLD FEMALE PATIENT WITH AN ELEOS DISTAL FEMUR REPLACEMENT UNDERWENT A REVISION SURGERY DUE TO AN ALLEGED INFECTION ON (B)(6) 2024. THE SURGEON, DOCTOR (B)(6), WASHED OUT THE OPERATIVE JOINT AND REPLACED THE ELEOS DEVICES LISTED ABOVE.

    Patient problems: Medical device site infection

    MAUDE report 19740089 ↗
  • InjuryMar 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED ON (B)(6) 2021 DUE TO AN ALLEGED INFECTION. DURING THIS REVISION SURGERY THE FOLLOWING IMPLANTS WERE REVISED: ELEOS TIBIAL HINGE COMPONENT, ELEOS TIBIAL POLY SPACER, ELEOS DISTAL FEMUR AXIAL PIN AND ELEOS DISTAL FEMUR. THE FOLLOWING ELEOS IMPLANTS REMAIN IMPLANTED FROM THE PATIENT'S ORIGINAL SURGERY: CANAL-FILLING STEM EXTENSION, TIBIAL BASEPLATE, MALE-FEMALE MIDSECTIONS, MALE-MALE MIDSECTION, AND PROXIMAL FEMUR. NO ADDITIONAL INFORMATION REGARDING THIS ADVERSE EVENT HAS BEEN REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 16482881 ↗
  • InjuryFeb 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED ON (B)(6) 2023 DUE TO AN ALLEGED INFECTION. DURING THIS REVISION SURGERY THE FOLLOWING IMPLANTS WERE REVISED: ELEOS TIBIAL HINGE COMPONENT, ELEOS TIBIAL POLY SPACER AND ELEOS DISTAL FEMUR AXIAL PIN. NO ADDITIONAL INFORMATION REGARDING THIS ADVERSE EVENT HAS BEEN PROVIDED.

    Patient problems: Insufficient Information; Unspecified Infection

    MAUDE report 16297132 ↗
  • InjuryDec 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT HAD AN ALLEGED INFECTION AND UNDERWENT A REVISION SURGERY ON (B)(6) 2022. THE FOLLOWING IMPLANTS WERE REVISED DURING THE REVISION: ELEOS DISTAL FEMUR AXIAL PIN, ELEOS TIBIAL HINGE W/ ROTATIONAL STOP AND ELEOS TIBIAL POLY SPACER.

    Patient problems: Insufficient Information; Unspecified Infection

    MAUDE report 16028279 ↗
  • InjurySep 8, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    A PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2021 PERFORMED BY DR. (B)(6). THE SURGERY WAS PERFORMED DUE TO ULCERATION AROUND ELEOS TOTAL FEMUR COMPONENT. DURING THE SURGERY, THE SURGEON REMOVED ALL IMPLANTS THAT WERE IMPLANTED AT THE TIME AS IT WAS DECIDED THAT THE PATIENT'S LEG WOULD BE AMPUTATED ABOVE THE KNEE. TO GIVE THE PATIENT'S THIGH STRUCTURE, A MALE-FEMALE MIDSECTION WITH A CUSTOM END CAP IMPLANT WAS PLACED IN THE PATIENT'S UPPER LEG.

    Patient problems: Ulcer

    MAUDE report 12441350 ↗
  • MalfunctionOct 6, 2020Loose or Intermittent Connection; Loose or Intermittent Connection

    THE PATIENT UNDERWENT A REVISION SURGERY (B)(6) 2020 PERFORMED BY (B)(6) DUE TO DISSOCIATION OF THE DISTAL FEMUR COMPONENT, AND A MALE-FEMALE MIDSECTION THAT CONNECTED TO THE SEGMENTAL STEM. THE SURGEON BELIEVED THAT RAISING THE JOINT LINE AND SHORTENING THE LIMB CAUSED INCREASED MECHANICAL FORCES ON THE COMPONENTS, WHICH LED TO THE DISSOCIATION. THE SURGEON LOWERED THE JOINT LINE AND INCREASED THE LIMB LENGTH BACK TO NORMAL. THE PATIENT'S PROXIMAL TIBIA COMPONENT, DISTAL FEMUR COMPONENT, AND TIBIAL HINGE COMPONENT WERE EXPLANTED AND REVISED ON (B)(6) 2020. THE PATIENT'S 10MM TIBIAL POLY SPACER WAS EXPLANTED AND REPLACED WITH AN 8MM TIBIAL POLY SPACER ON (B)(6) 2020. A 140MM MALE-FEMALE MIDSECTION, A 70MM MALE-FEMALE MIDSECTION, AND A 50 MALE-FEMALE MIDSECTION WERE PLACED ON (B)(6) 2020 TO REPLACE A 90MM MALE-FEMALE MIDSECTION, TWO 40MM MALE-FEMALE MIDSECTIONS, AND A 100MM MALE-FEMALE MIDSECTION.

    Patient problems: Insufficient Information

    MAUDE report 10634925 ↗
  • InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT UNDERWENT A REVISION SURGERY DUE TO AN ALLEGED INFECTION AND THE FEMORAL STEM LOOSENING.

    Patient problems: Unspecified Infection

    MAUDE report 9707195 ↗
  • InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT UNDERWENT A REVISION SURGERY DUE TO AN ALLEGED INFECTION AND THE FEMORAL STEM LOOSENING.

    Patient problems: Unspecified Infection

    MAUDE report 9706887 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K161520. One submission often covers many models, so check the product and lot details in each record.

  • Z-2423-2025 · Jul 11, 2025 · Open, Classified
    ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS BOWED CANAL FILLING STEMS. Canal filling stem implants.
  • Z-2424-2025 · Jul 11, 2025 · Open, Classified
    ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS SEGMENTAL STEM CANAL FILLING STRAIGHT SLOTTED SPLINED 2/3 PLASMA. Canal filling stem implants.
  • Z-2425-2025 · Jul 11, 2025 · Open, Classified
    ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS STEM EXTENSION, STRAIGHT, SPLINED, SLOTTED. Canal filling stem implants.
  • Z-1857-2025 · Sep 22, 2024 · Open, Classified
    ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
  • Z-1879-2025 · Sep 1, 2023 · Open, Classified
    a. Segmental Stem, Cemented, Straight, 100MM, 1 EACH; Model Numbers: CS-09100-03M, CS-10100-03M. b. Segmental Stem, Cemented, Straight, 120MM, 1 EACH. Model Numbers: CS-11120-03M,
  • Z-1884-2025 · Sep 1, 2023 · Open, Classified
    Male-Female Midsection with the following sizes and model numbers. a. Male-Female Midsection, 40mm, 1 EACH; Model Number: 25001040E b. Male-Female Midsection, 50mm, 1 EACH; Model
  • Z-1881-2025 · Sep 1, 2023 · Open, Classified
    Segmental Stem, Canal Filling, Straight, 120MM, 1 EACH. Model Numbers: FS-11120-03M, FS-12120-03M, FS-13120-03M, FS-14120-03M, FS-15120-03M, FS-16120-03M, FS-17120-03M, FS-18120-03
  • Z-1882-2025 · Sep 1, 2023 · Open, Classified
    Segmental Stem, Canal Filling, Bowed 152MM, 1 EACH. Model Numbers: FB-11152-03M, FB-12152-03M, FB-13152-03M, FB-14152-03M, FB-15152-03M, FB-16152-03M, FB-17152-03M, FB-18152-03M, F
  • Z-2647-2017 · May 23, 2017 · Terminated
    ELEOS LIMB SALVAGE SYSTEM SEGMENTAL STEM, CANAL FILLING, DIAMETER 20MM, LENGTH 120MM, STYLE STRAIGHT, TYPE CANAL FILLING, COLLAR 25MM; REF 2500SP20E
  • Z-2648-2017 · May 23, 2017 · Terminated
    ELEOS LIMB SALVAGE SYSTEM SEGMENTAL STEM, CANAL FILLING, DIAMETER 20MM, LENGTH 120MM, STYLE STRAIGHT, TYPE CANAL FILLING, COLLAR 25MM; REF 2500SP22E
  • Z-2646-2017 · May 23, 2017 · Terminated
    ELEOS LIMB SALVAGE SYSTEM SEGMENTAL STEM, CANAL FILLING, DIAMETER 18MM, LENGTH 120MM, STYLE STRAIGHT, TYPE CANAL FILLING, COLLAR 25MM, REF 2500SP18E

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page