My3D® Personalized Pelvic Reconstruction CS-65055-04MRecalls & reported problems
Onkos Surgical, Inc. · CANCELLOUS BONE SCREW, 6.5MM DIA x 55MM LENGTH · Class 2
All models of My3D Personalized Pelvic Reconstruction (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B278CS6505504M0), received between Dec 11, 2024 and Dec 11, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K212815); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 11, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A CUSTOM ONKOS INTERCALLARY TIBIA CONSTRUCT EXPERIENCED SUBSIDENCE AT THE PROXIMAL PORTION. THE PATIENT'S INDEX PROCEDURE TOOK PLACE ON (B)(6) 2024. THE PATIENT WAS REVISED ON (B)(6) 2024 TO AUGMENT THE PROXIMAL PORTION WITH BONE CEMENT AND REPLACE THE FOUR PROXIMAL SCREWS. THIS EVENT WILL BE REPORTABLE AS A SERIOUS INJURY, DUE TO THE REVISION PROCEDURE. THIS REPORT CAPTURES THE MY3D PERSONALIZED PELVIC RECONSTRUCTION CANCELLOUS BONE SCREW, SECOND OF THE FOUR PROXIMAL SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20903368 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K212815. One submission often covers many models, so check the product and lot details in each record.
- Z-0091-2027 · Jul 24, 2026 · Open, ClassifiedOnkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M
- Z-0090-2027 · Jul 24, 2026 · Open, ClassifiedOnkos Surgical Inc. My3D Personalized Pelvic Implant, Porous, Left REF PV-ACREC-04M
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.