ELEOS™ KSP17100ERecalls & reported problems
Onkos Surgical, Inc. · ELEOS™ STEM EXTENSION18X100MM STRAIGHT, SPLINED, SLOTTED 18X100MM · Class 2 · Implant
All models of ELEOS (127)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI B278KSP17100E0), received between Jan 2, 2018 and May 11, 2023 — classified as 5 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
11 other recalls cover devices cleared under the same FDA submission (K161520); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Device Dislodged or Dislocated2
- Device Slipped2
- Insufficient Device Problem Information1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Loss of Osseointegration1
- Mechanical Problem1
Patient problems
- Insufficient Information2
- No Information2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Failure of Implant1
- Joint Dislocation1
- Joint Disorder1
- No Code Available1
- Scar Tissue1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 11, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2023 DUE TO AN ALLEGED LOOSENING OF THE PATIENT'S 18X100MM CANAL-FILLING STEM EXTENSION. DURING THE REVISION SURGERY, THE PATIENT'S RESURFACING FEMUR, RESURFACING FEMUR AXIAL PIN, AND 18X100MM CANAL-FILLING STEM EXTENSION WERE REVISED TO A DISTAL FEMUR, DISTAL FEMUR AXIAL PIN, AND 13X120MM CEMENTED SEGMENTAL STEM. THE FOLLOWING IMPLANTS WERE ALSO REVISED: TIBIAL HINGE COMPONENT AND TIBIAL POLY SPACER. THERE WERE NO ALLEGED MALFUNCTIONS WITH THE RESURFACING FEMUR, RESURFACING FEMUR AXIAL PIN, TIBIAL HINGE COMPONENT, AND TIBIAL POLY SPACER. NO ADDITIONAL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED.
Patient problems: Insufficient Information
MAUDE report 16915338 ↗InjuryMay 21, 2021Mechanical Problem; Insufficient Device Problem Information; Mechanical Problem; Insufficient Device Problem Information
THE PATIENT INVOLVED IN THIS COMPLAINT UNDERWENT A REVISION SURGERY ON (B)(6) 2021 PERFORMED BY DR. (B)(6). IT WAS REPORTED BY THE SALES REPRESENTATIVE THAT THE PATIENT PULLED OUT THEIR SURGICAL DRAIN WHEN THE DRAIN WAS SUPPOSED TO REMAIN IN THE PATIENT ACCORDING TO POST-OPERATIVE INSTRUCTIONS. THE DRAIN WAS PLACED ON (B)(6) 2021 DURING A SURGERY PERFORMED BY DR. (B)(6) TO PLACE ELEOS COMPONENTS FOR AN UNKNOWN REASON. DURING THE REVISION SURGERY, THE PATIENT HAD ELEOS COMPONENTS EXPLANTED AND REPLACED. THERE WAS NO REPORTED FAILURE OF THE ELEOS COMPONENTS. IT IS UNKNOWN AT THIS TIME WHY THE COMPONENTS WERE REPLACED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Insufficient Information; Joint Dislocation
MAUDE report 11866634 ↗InjuryMar 19, 2018Detachment of Device or Device Component; Device Slipped; Detachment of Device or Device Component; Device Dislodged or Dislocated
A REVISION SURGERY WAS PERFORMED TO DRAIN AND IRRIGATE THE KNEE DUE TO AN ALLEGED INFECTION. DURING THE REVISION SURGERY, IT WAS FOUND THAT THE RESURFACING FEMUR HAD LOOSENED AND DISSOCIATED FROM THE STEM EXTENSION.
Patient problems: Unspecified Infection
MAUDE report 7350367 ↗InjuryFeb 16, 2018Device Slipped; Detachment of Device or Device Component; Detachment of Device or Device Component
PATIENT WITH ELEOS REVISION HINGE, HEARD A SNAPPING/POPPING NOISE COME FROM THEIR KNEE WHILE DOING SQUATS AT PHYSICAL THERAPY. A REVISION SURGERY WAS PERFORMED DUE TO ARTHROFIBROSIS, FAILED TOTAL KNEE AND MORBID OBESITY, TO REMOVE EXCESS SCAR TISSUE AND PERFORM A POLY SWAP. UPON INSPECTING THE JOINT, THE SURGEON NOTICED A FREE PIECE OF CEMENT AND THAT THE CEMENTED RESURFACING FEMUR HAD LOOSENED AND DISSOCIATED FROM THE STEM EXTENSION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Joint Disorder; No Information; Scar Tissue
MAUDE report 7278343 ↗InjuryJan 2, 2018Loss of Osseointegration; Detachment of Device or Device Component; Device Dislodged or Dislocated; Loss of Osseointegration; Detachment of Device or Device Component
PATIENT WITH ELEOS REVISION HINGE, HEARD A SNAPPING/POPPING NOISE COME FROM THEIR KNEE WHILE DOING SQUATS AT PHYSICAL THERAPY. A REVISION SURGERY WAS PERFORMED TO REMOVE EXCESS SCAR TISSUE AND EXCHANGE THE POLY COMPONENTS FOR A THICKER SPACER. DURING SURGERY THE SURGEON NOTICED THAT THE CANAL FILLING STEM HAD DISSOCIATED FROM THE INTERMEDULLARY CANAL. THE CANAL FILLING STEM WAS REMOVED AND A CEMENTED STEM WAS IMPLANTED.
Patient problems: No Code Available; No Information
MAUDE report 7156567 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K161520. One submission often covers many models, so check the product and lot details in each record.
- Z-2425-2025 · Jul 11, 2025 · Open, ClassifiedELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS STEM EXTENSION, STRAIGHT, SPLINED, SLOTTED. Canal filling stem implants.
- Z-2423-2025 · Jul 11, 2025 · Open, ClassifiedELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS BOWED CANAL FILLING STEMS. Canal filling stem implants.
- Z-2424-2025 · Jul 11, 2025 · Open, ClassifiedELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS SEGMENTAL STEM CANAL FILLING STRAIGHT SLOTTED SPLINED 2/3 PLASMA. Canal filling stem implants.
- Z-1857-2025 · Sep 22, 2024 · Open, ClassifiedELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
- Z-1882-2025 · Sep 1, 2023 · Open, ClassifiedSegmental Stem, Canal Filling, Bowed 152MM, 1 EACH. Model Numbers: FB-11152-03M, FB-12152-03M, FB-13152-03M, FB-14152-03M, FB-15152-03M, FB-16152-03M, FB-17152-03M, FB-18152-03M, F
- Z-1879-2025 · Sep 1, 2023 · Open, Classifieda. Segmental Stem, Cemented, Straight, 100MM, 1 EACH; Model Numbers: CS-09100-03M, CS-10100-03M. b. Segmental Stem, Cemented, Straight, 120MM, 1 EACH. Model Numbers: CS-11120-03M,
- Z-1881-2025 · Sep 1, 2023 · Open, ClassifiedSegmental Stem, Canal Filling, Straight, 120MM, 1 EACH. Model Numbers: FS-11120-03M, FS-12120-03M, FS-13120-03M, FS-14120-03M, FS-15120-03M, FS-16120-03M, FS-17120-03M, FS-18120-03
- Z-1884-2025 · Sep 1, 2023 · Open, ClassifiedMale-Female Midsection with the following sizes and model numbers. a. Male-Female Midsection, 40mm, 1 EACH; Model Number: 25001040E b. Male-Female Midsection, 50mm, 1 EACH; Model
- Z-2648-2017 · May 23, 2017 · TerminatedELEOS LIMB SALVAGE SYSTEM SEGMENTAL STEM, CANAL FILLING, DIAMETER 20MM, LENGTH 120MM, STYLE STRAIGHT, TYPE CANAL FILLING, COLLAR 25MM; REF 2500SP22E
- Z-2647-2017 · May 23, 2017 · TerminatedELEOS LIMB SALVAGE SYSTEM SEGMENTAL STEM, CANAL FILLING, DIAMETER 20MM, LENGTH 120MM, STYLE STRAIGHT, TYPE CANAL FILLING, COLLAR 25MM; REF 2500SP20E
- Z-2646-2017 · May 23, 2017 · TerminatedELEOS LIMB SALVAGE SYSTEM SEGMENTAL STEM, CANAL FILLING, DIAMETER 18MM, LENGTH 120MM, STYLE STRAIGHT, TYPE CANAL FILLING, COLLAR 25MM, REF 2500SP18E
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page