Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B458201550000), received between Jan 16, 2026 and Jan 16, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1
  • Solder Joint Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Material Separation; Solder Joint Fracture

    DURING A CLINICAL PROCEDURE, AN INFILTRATION CANNULA WAS REPORTED TO HAVE SEPARATED AT THE SHAFT-TO-BASE INTERFACE DURING USE. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE CANNULA, AND NO PATIENT INJURY OR ADVERSE CLINICAL OUTCOME OCCURRED. THE AFFECTED DEVICE WAS NOT RETURNED TO THE MANUFACTURER; THEREFORE, THE INVESTIGATION WAS LIMITED TO REVIEW OF CUSTOMER PROVIDED PHOTOGRAPHS AND INTERNAL MANUFACTURING RECORDS. REVIEW OF THE PHOTOGRAPHS CONFIRMED SEPARATION OF THE CANNULA SHAFT FROM THE BASE, WITH NO OBVIOUS BENDING, DEFORMATION, OR EXTERNAL DAMAGE VISIBLE; INTERNAL JOINT INTEGRITY COULD NOT BE EVALUATED IN THE ABSENCE OF THE DEVICE. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH APPROVED SPECIFICATIONS, WITH ALL REQUIRED IN-PROCESS AND FINAL INSPECTIONS COMPLETED AND ACCEPTED AND NO DEVIATIONS, NONCONFORMANCE'S, OR REWORK DOCUMENTED. BASED ON THE AVAILABLE INFORMATION, THE EVENT IS CLASSIFIED AS AN UNDETERMINED MECHANICAL FAILURE AT THE SHAFT-TO-BASE INTERFACE. THE REPORTED FAILURE MODE IS CONSISTENT WITH SEPARATION THAT MAY OCCUR UNDER LATERAL STRESS OR EXCESSIVE MECHANICAL LOADING WHICH CAN COMPROMISE STRUCTURAL INTEGRITY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24101610 ↗

Other Cannula, Surgical, General & Plastic Surgery devices

All Cannula, Surgical, General & Plastic Surgery devices →