Optimotion Blue CR HCCR Total Knee System 1014-B-1300Recalls & reported problems
Optimotion Implants LLC · Tibial Tray Component, Porous Titanium, Size 3 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B6961014B13000), received between Aug 25, 2021 and Aug 25, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size1
- Inadequate or Insufficient Training1
- Misassembled During Installation1
Patient problems
- Bone Fracture(s)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 25, 2021Inadequacy of Device Shape and/or Size; Inadequate or Insufficient Training; Misassembled During Installation; Inadequacy of Device Shape and/or Size; Inadequate or Insufficient Training; Misassembled During Installation
USER REPORTED EVENT OF TIBIAL STEM DISASSOCIATION FROM A TIBIAL BASEPLATE. THE USER IS A HEALTHCARE PROFESSIONAL WHO DETERMINED THAT THE PATIENTS WERE, "OBESE AND SUSTAINED CATASTROPHIC SETTLING/FRACTURE OF THE BONE UNDERNEATH THE BASE PLATE WHICH IS WHAT I ATTRIBUTE TO THE DISLODGEMENT OF THE TIBIAL STEM FROM THE BASEPLATE". THIS INJURY REQUIRED REVISION SURGERY TO RESTORE OSSEOINTEGRATION. USER REPORT SUGGESTS THAT THE "BASE PLATE WAS HUGELY UNDERSIZED AND MALPOSITIONED". THIS EVENT OCCURRED CONCURRENTLY IN ONE ADDITIONAL PATIENT WHERE THE IMPLANTS WERE INSTALLED (29 DAYS APART) BY THE SAME SURGEON (WHO IS NOT THE REPORTING USER). ADDITIONAL COMMUNICATION ON (B)(6) 2021 CONFIRMED THAT, "[INSTALLING SURGEON] CONFIRMS BY RADIOGRAPHIC EXAMINATION THAT HE HAS GROSSLY UNDERSIZED [PATIENT]." IN THE INITIAL COMMUNICATION THE USER INDICATED THEIR, "SUSPICION IS THAT OUR OPERATING TEAM DID NOT IMPACT THE STEM SUFFICIENTLY TO THE TIBIAL TRAY AND WHEN THERE WAS CATASTROPHIC SETTLING FORCES CAUSED THE STEM TO DISENGAGE". ADDITIONAL COMMUNICATION ON (B)(6) 2021 CONFIRMED THAT JUST PRIOR TO THE EVENT, "SURGEONS HAD TRANSITIONED TO ALLOWING THE SURGICAL TECHS TO IMPACT THE STEM. THE TECHS WERE
Patient problems: Bone Fracture(s)
MAUDE report 12366002 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page