Skip to content

DUALXS T/PLIF 8 DEG x 7-9MM 12941001003Recalls & reported problems

Amplify Surgical, Inc. · Lordosis & Heights: 8° x 7-9mm Collapsed Width (mm): 10 Expanded Width (mm): 20 Collapsed Height (mm): 7 Expanded Height (mm): 9 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B707129410010030), received between Jan 29, 2026 and Jan 29, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Migration1

Patient problems

  • Spinal deformity1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2026Loosening of Implant Not Related to Bone-Ingrowth; Migration

    DR. DID NOT USE EXPANDABLE TRIAL. DR. IMPLANTED A DUALX SLIM INTO PATIENT. DURING EXPANSION SURGICAL SITE GAVE AUDIBLE POP SOUND DURING EXPANSION (CANNOT CONFIRM IMPLANT OR ANATOMY). UNDER LATERAL XRAY IMPLANT SHOWS EXPANDED AND IN CORRECT POSITION BUT POSTERIOR PORTION APPEARS TO NOT BE CLEAR OF POSTERIOR FORAMEN. LOCK SCREW WOULD NOT THREAD INTO NOSE. INSTRUMENTS WERE DISCONNECTED FROM IMPLANT IN AN ATTEMPT TO GET LOCK SCREW TO THREAD INTO NOSE. LOCK SCREW WOULD NOT THREAD INTO NOSE. LOCK SCREW WAS REMOVED. IMPLANT WAS MALLETED FORWARD TO CLEAR POSTERIOR NEURAL ELEMENTS. LATERAL XRAY SHOWS EXPANDED IMPLANT BUT ANTERIOR OF THE VERTEBRAL BODY. AP XRAY SHOWS POSTERIOR PORTION OF IMPLANT ONLY PARTIALLY EXPANDED (IMPLANT APPEARS TO BE INTACT AND COMPONENTS NOT BROKEN OR DISLOCATED). ATTEMPTS TO RE-ATTACH INSTRUMENTS TO IMPLANT FAILED. IMPLANT LEFT IN PATIENT PARTIALLY LATERALLY EXPANDED AND PARTIALLY VERTICALLY EXPANDED WITHOUT LOCK SCREW.

    Patient problems: Spinal deformity

    MAUDE report 24218202 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page