Skip to content

IRM® Intermediate Restorative Material 610007Recalls & reported problems

Dentsply International Inc. · IRM® Intermediate Restorative Material, Ivory, Standard Package · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI D0026100071), received between Mar 19, 2025 and Mar 19, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient-Device Incompatibility2

Patient problems

  • Hypersensitivity/Allergic reaction2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    IN THIS EVENT IT IS REPORTED THAT IRM POWDER 40G THAT WAS USED IN TREATMENT, THE PATIENT DEVELOPED AN ALLERGIC REACTION AND HAD TO BE HOSPITALIZED. NO INFORMATION ABOUT SIGNS AND SYMPTOMS HAS BEEN SHARED.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 21641515 ↗
  • InjuryMar 19, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    IN THIS EVENT IT IS REPORTED THAT IRM LIQUID 15ML THAT WAS USED IN TREATMENT, THE PATIENT DEVELOPED AN ALLERGIC REACTION AND HAD TO BE HOSPITALIZED. NO INFORMATION ABOUT SIGNS AND SYMPTOMS HAS BEEN SHARED.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 21641488 ↗

Other CEMENT, DENTAL devices

All CEMENT, DENTAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.