Cavitron® 81302Recalls & reported problems

Dentsply Professional · Cavitron® SELECT™ SPS™ Ultrasonic Scaler. Also contains: Steri-Mate® Handpiece with Swivel Cable 120/11V · Class 2

Quick summary

This Dentsply Cavitron Select SPS ultrasonic scaler is a dental cleaning device with a handpiece that sprays water. Most of the 67 reports describe the water flow stopping or dropping during use while the handpiece gets hot; a couple mention burns. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
67
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI D003813021), received between Jan 24, 2024 and May 1, 2026 — classified as 67 malfunctions.

The most frequently reported specific device problem is “No Flow” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 5 reportsJul 2025: 4 reportsAug 2025: 7 reportsSep 2025: 7 reportsOct 2025: 3 reportsNov 2025: 4 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 2 reportsMar 2026: 5 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Flow36
  • Restricted Flow rate29
  • Overheating of Device17

Patient problems

  • No Clinical Signs, Symptoms or Conditions65
  • Burn(s)2

Recent adverse event reports

The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056934 ↗
  • MalfunctionApr 16, 2026Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24903798 ↗
  • MalfunctionApr 8, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816445 ↗
  • MalfunctionMar 30, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24733677 ↗
  • MalfunctionMar 24, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP. NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24674102 ↗
  • MalfunctionMar 19, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24640538 ↗
  • MalfunctionMar 19, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE CAVITRON IS GETTING HOT. NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24640537 ↗
  • MalfunctionMar 13, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS GETTING HOT; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24594699 ↗
  • MalfunctionFeb 24, 2026Overheating of Device; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428791 ↗
  • MalfunctionFeb 11, 2026No Flow

    IN THIS EVENT IT IS REPORTED THAT A SELECT SPS SWVL/RESRV, 230VUK/I IS ALLEGED TO HAVE NO WATER COMING THROUGH. NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24316874 ↗
  • MalfunctionJan 22, 2026No Flow

    IN THIS EVENT IT IS REPORTED THAT A SELECT SPS SWVL/RESRV,230VUK/I ALLEGEDLY HAD NO WATER FLOW. NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144659 ↗
  • MalfunctionJan 13, 2026Restricted Flow rate

    WHILE USING A CAVITRON G124 GENERATOR THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24056743 ↗
  • MalfunctionDec 28, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP,NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23903724 ↗
  • MalfunctionDec 3, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124 GENERATOR (DNA), THEY ALLEGE THAT THE INSERTS HEAT UP. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23705391 ↗
  • MalfunctionNov 13, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE GETS HOT; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23543196 ↗
  • MalfunctionNov 10, 2025Restricted Flow rate; No Flow; Restricted Flow rate; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512397 ↗
  • MalfunctionNov 5, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THE INSERT HEATED UP WHILE IN USE. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476237 ↗
  • MalfunctionNov 5, 2025Restricted Flow rate; No Flow; Restricted Flow rate; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476231 ↗
  • MalfunctionOct 16, 2025No Flow; No Flow

    WHILE USING A CAVITORN SELELCT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310318 ↗
  • MalfunctionOct 16, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310315 ↗
  • MalfunctionOct 16, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THE INSERT AND THE HANDPIECE ARE HEATING UP. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310314 ↗
  • MalfunctionSep 15, 2025No Flow; No Flow

    IN THIS EVENT IT IS REPORTED THAT A SELECT SPS SWVL/RESRV, 230VUK/I IS ALLEGED TO HAVE OVERHEATED AND NO WATER FLOW. NO INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23050399 ↗
  • MalfunctionSep 10, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THE INSERT OVER HEATED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017456 ↗
  • MalfunctionSep 10, 2025No Flow; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017455 ↗
  • MalfunctionSep 8, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22994270 ↗

Other Scaler, ultrasonic devices

All Scaler, ultrasonic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.