Cavitron® 81302Recalls & reported problems

Dentsply Professional · Cavitron® SELECT™ SPS™ Ultrasonic Scaler. Also contains: Steri-Mate® Handpiece with Swivel Cable 120/11V · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Dentsply Cavitron Select SPS ultrasonic scaler is a dental cleaning device with a handpiece that sprays water. Most of the 67 reports describe the water flow stopping or dropping during use while the handpiece gets hot; a couple mention burns. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
67
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI D003813021), received between Jan 24, 2024 and May 1, 2026 — classified as 67 malfunctions.

The most frequently reported specific device problem is “No Flow” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 5 reportsJul 2025: 4 reportsAug 2025: 7 reportsSep 2025: 7 reportsOct 2025: 3 reportsNov 2025: 4 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 2 reportsMar 2026: 5 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Flow36
  • Restricted Flow rate29
  • Overheating of Device17

Patient problems

  • No Clinical Signs, Symptoms or Conditions65
  • Burn(s)2

Recent adverse event reports

The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056934 ↗
  • MalfunctionApr 16, 2026Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24903798 ↗
  • MalfunctionApr 8, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816445 ↗
  • MalfunctionMar 30, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24733677 ↗
  • MalfunctionMar 24, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP. NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24674102 ↗
  • MalfunctionMar 19, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24640538 ↗
  • MalfunctionMar 19, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE CAVITRON IS GETTING HOT. NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24640537 ↗
  • MalfunctionMar 13, 2026No Flow

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS GETTING HOT; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24594699 ↗
  • MalfunctionFeb 24, 2026Overheating of Device; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428791 ↗
  • MalfunctionFeb 11, 2026No Flow

    IN THIS EVENT IT IS REPORTED THAT A SELECT SPS SWVL/RESRV, 230VUK/I IS ALLEGED TO HAVE NO WATER COMING THROUGH. NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24316874 ↗
  • MalfunctionJan 22, 2026No Flow

    IN THIS EVENT IT IS REPORTED THAT A SELECT SPS SWVL/RESRV,230VUK/I ALLEGEDLY HAD NO WATER FLOW. NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144659 ↗
  • MalfunctionJan 13, 2026Restricted Flow rate

    WHILE USING A CAVITRON G124 GENERATOR THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24056743 ↗
  • MalfunctionDec 28, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP,NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23903724 ↗
  • MalfunctionDec 3, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124 GENERATOR (DNA), THEY ALLEGE THAT THE INSERTS HEAT UP. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23705391 ↗
  • MalfunctionNov 13, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE GETS HOT; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23543196 ↗
  • MalfunctionNov 10, 2025Restricted Flow rate; No Flow; Restricted Flow rate; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512397 ↗
  • MalfunctionNov 5, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THE INSERT HEATED UP WHILE IN USE. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476237 ↗
  • MalfunctionNov 5, 2025Restricted Flow rate; No Flow; Restricted Flow rate; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476231 ↗
  • MalfunctionOct 16, 2025No Flow; No Flow

    WHILE USING A CAVITORN SELELCT SPS G124 THEY ALLEGE THAT THEY HAVE NO WATER FLOW AND THE HANDPIECE IS HEATING UP; NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310318 ↗
  • MalfunctionOct 16, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124 THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310315 ↗
  • MalfunctionOct 16, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THE INSERT AND THE HANDPIECE ARE HEATING UP. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310314 ↗
  • MalfunctionSep 15, 2025No Flow; No Flow

    IN THIS EVENT IT IS REPORTED THAT A SELECT SPS SWVL/RESRV, 230VUK/I IS ALLEGED TO HAVE OVERHEATED AND NO WATER FLOW. NO INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23050399 ↗
  • MalfunctionSep 10, 2025Overheating of Device; Overheating of Device

    WHILE THE CUSTOMER WAS USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THE INSERT OVER HEATED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017456 ↗
  • MalfunctionSep 10, 2025No Flow; No Flow

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE NO WATER AND THE HANDPIECE IS GETTING HOT, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017455 ↗
  • MalfunctionSep 8, 2025Restricted Flow rate; Restricted Flow rate

    WHILE USING A CAVITRON SELECT SPS G124, THEY ALLEGE THAT THEY HAVE LOW WATER FLOW AND THE HANDPIECE IS HEATING UP, NO INJURY RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22994270 ↗

FDA clearance and classification

The FDA regulates product code ELC (scaler, ultrasonic) under 21 CFR 872.4850 as a Class II device. 21 CFR 872.4850 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Cavitron® 81302

Has the Cavitron® 81302 been recalled?

No FDA recall record in our data names this exact device identifier (DI D003813021), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Cavitron® 81302?

The FDA's MAUDE database lists 67 adverse event reports for this device, received between Jan 24, 2024 and May 1, 2026. In those reports, the most common specific device problems are “No Flow” (36), “Restricted Flow rate” (29) and “Overheating of Device” (17); the most common patient problem is “Burn(s)” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Cavitron® 81302?

None of the reports are classified as deaths, 0 as injuries and 67 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Cavitron® 81302, and is it still sold?

The FDA's device database lists Dentsply Professional as the company responsible for it (brand Cavitron®, model 81302). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Scaler, ultrasonic devices

All Scaler, ultrasonic devices →