SuperLine FX3610SWCRecalls & reported problems
Dentium USA · Endosseous dental implant · Class 2 · Implant
Quick summary
This Dentium SuperLine FX3610SWC is a root-form dental implant placed in the jawbone. Its 93 reports, with 93 injuries and no deaths, mostly describe the implant not bonding to the jawbone, which led to removal and a further surgery for the patient. Some reports also mention infection and bone loss around the implant. No recalls name this device. Reports for the last 12 months (0) are fewer than in the previous 12 months (4).
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 93
- Deaths reported
- 0
- Injuries reported
- 93
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 93 adverse event reports that list this exact device (DI D755FX3610SWC0), received between Feb 28, 2017 and Jun 3, 2025 — classified as 93 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (76 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate76
- Osseointegration Problem63
- Loss of Osseointegration13
- Use of Device Problem12
- Fracture4
Patient problems
- Inadequate Osseointegration89
- Failure of Implant42
- Unspecified Infection24
- Patient Problem/Medical Problem13
- Pain1
Recent adverse event reports
The 25 most recent of 93 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2025Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 5 YEARS AND 9 MONTHS AFTER IMPLANT PLACEMENT. X-RAY WAS PROVIDED. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 22135127 ↗InjuryJun 3, 2025Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 5 YEARS AND 9 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 22135124 ↗InjuryMay 13, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 7 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. PERIODONTAL DISEASE, BONE RESORPTION AND PERI-IMPLANTITIS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 22015378 ↗InjuryNov 6, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE FIXTURE WITH ATTACHED ABUTMENT WAS RETURNED. SLA SURFACE TREATMENT WAS CONFIRMED. EVIDENCE OF OSSEOINTEGRATION WAS OBSERVED. THE VISUAL EXAMINATION WAS ACCEPTABLE. THE DOCTOR REPORTED ON FORM F1060, REV 8 THAT THE RETURNED FIXTURE WAS FXS3612C LOT C15LB739S. THE RELATED SALES ORDER #(B)(4) CONFIRMED THAT FX3610SWC LOT G20NB641S WAS ORDERED AND SHIPPED. THIS WAS CONFIRMED WITH (B)(6), RDA, FROM (B)(6) ORAL'S OFFICE BY DENTIUM STAFF MK. THERE WAS NO MINGLING ISSUE. THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 8 MONTHS AFTER IMPLANT PLACEMENT. THE PATIENT'S X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. BONE RESORPTION WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 20622756 ↗InjuryMar 13, 2024Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION 2 YEARS AND 4 MONTHS AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED BONE RESORPTION, PAIN, AND ABNORMAL SENSATION AROUND IMPLANT PLACEMENT WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain
MAUDE report 18897465 ↗InjuryJan 30, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 18607842 ↗InjuryAug 29, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 6 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 17647278 ↗InjuryAug 29, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 6 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 17647274 ↗InjuryOct 12, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE AROUND TWO MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. ABNORMAL SENSATION AROUND THE IMPLANT PLACEMENT WAS OBSERVED DURING THE REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 15588111 ↗InjuryMay 20, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY ONE MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 14459436 ↗InjuryApr 20, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE AROUND 4 MONTHS AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED PERI-IMPLANTITIS, LOCAL INFECTION AND BONE RESORPTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 14164347 ↗InjuryApr 20, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE AROUND 4 MONTHS AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED PERI-IMPLANTITIS, LOCAL INFECTION AND BONE RESORPTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 14164345 ↗InjuryFeb 23, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN ONE YEAR, APPROXIMATELY 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED BONE RESORPTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 13588794 ↗InjuryFeb 21, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE DOCTOR REPORTED NO SIGNIFICANT FINDINGS DURING THE IMPLANT REMOVAL SURGERY. BONE WAS GRAFTED AT SURGERY LOCATION AND PATIENT WILL RETRY IMPLANT PLACEMENT AT A FUTURE DATE.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 13573293 ↗InjuryFeb 17, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INSUFFICIENT PRIMARY STABILITY, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED LOCAL INFECTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 13552862 ↗InjuryNov 8, 2021Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 2 YEARS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED BONE RESORPTION DURING THE IMPLANT PLACEMENT SURGERY. THE PATIENT WILL REQUIRE ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12776004 ↗InjurySep 15, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 24 DAYS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12477928 ↗InjurySep 15, 2021Fracture; Fracture
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE AT INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant
MAUDE report 12477925 ↗InjurySep 14, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE ONE MONTH FOUR DAYS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE IMPLANT REMOVAL SURGERY, LOCAL INFECTION AROUND IMPLANT PLACEMENT WAS REPORTED. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 12471902 ↗InjuryJul 30, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 2 WEEKS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. DURING THE SURGERY NO SIGNIFICANT FINDINGS WERE REPORTED. IT WAS REPORTED THAT THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12256944 ↗InjuryJul 13, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 2 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY BONE RESORPTION WAS OBSERVED. PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12162962 ↗InjuryJun 18, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 2 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE REPORTED. PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12026296 ↗InjuryJun 18, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE LESS THAN A MONTH AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE OBSERVED. PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12026294 ↗InjuryJun 18, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE LESS THAN A MONTH AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE OBSERVED. PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12026289 ↗InjuryMay 2, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY A MONTH AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 3, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE OBSERVED. PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 11758786 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.