Implantium FX3810MLCRecalls & reported problems

Dentium USA · Endosseous dental implant · Class 2 · Implant

All models of Implantium (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Implantium FX3810MLC endosseous dental implant has 40 adverse event reports, all describing injuries with no deaths and no recalls. Most describe the implant failing to bond with bone and being removed, sometimes with infection, pain, or bone loss, and a few report fractures. Reports span 2017 to 2026.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
40
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI D755FX3810MLC0), received between Oct 10, 2017 and Jun 11, 2026 — classified as 40 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (26 reports).

1 other recall cover devices cleared under the same FDA submission (K041368); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate26
  • Osseointegration Problem14
  • Loss of Osseointegration9
  • Fracture3
  • Adverse Event Without Identified Device or Use Problem1
  • Use of Device Problem1

Patient problems

  • Inadequate Osseointegration35
  • Failure of Implant26
  • Pain11
  • Unspecified Infection11
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2026Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 3 YEARS AND 2 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS NOT PROVIDED. THE ORAL HYGIENE AROUND IMPLANT SITE WAS POOR. LOCAL INFECTION, BONE RESORPTION, AND PERI-IMPLANTITIS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT NEEDS ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 25480282 ↗
  • InjuryDec 2, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INSUFFICIENT PRIMARY STABILITY, 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. NO SIGNIFICANT FINDINGS WERE FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 23700640 ↗
  • InjuryMay 9, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21996521 ↗
  • InjuryMar 17, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 5 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21627484 ↗
  • InjuryFeb 18, 2025Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 1 YEAR AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21406470 ↗
  • InjuryFeb 18, 2025Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 1 YEAR AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION AND PERI-IMPLANTITIS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21404802 ↗
  • InjuryJan 8, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS NOT REPORTED, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. LOCAL INFECTION AND BONE RESORPTION WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED AND WILL BE GOING WITH A BRIDGE OPTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 21105989 ↗
  • InjuryOct 14, 2024Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 2 YEARS AND 7 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS NOT REPORTED. LOCAL INFECTION, BONE RESORPTION, PAIN WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 20442349 ↗
  • InjurySep 20, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20268187 ↗
  • InjuryOct 25, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 18006394 ↗
  • InjuryJun 9, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 17101584 ↗
  • InjuryJun 6, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 17073363 ↗
  • InjuryApr 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY ONE MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE DOCTOR REPORTED LOCAL INFECTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 16666146 ↗
  • InjuryJan 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 16099543 ↗
  • InjuryJan 3, 2023Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION 4 YEARS AND 9 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION AND PERI-IMPLANTITIS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 16092184 ↗
  • InjuryMar 4, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO FAILURE OF OSSEOINTEGRATION APPROXIMATELY 3 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE DOCTOR REPORTED THE PATIENT HAD LOCAL INFECTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 13676261 ↗
  • InjuryJun 18, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 10 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 3, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, BONE RESORPTION WAS OBSERVED. PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 12026284 ↗
  • InjuryJun 17, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 10 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 3, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, BONE RESORPTION WAS OBSERVED. PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 12022022 ↗
  • InjuryApr 16, 2021Fracture; Fracture

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE ALMOST TWO YEARS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 1, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. DURING THE SURGERY, LOCAL INFECTION, BONE RESORPTION, PERI-IMPLANTITIS, PAIN, AND FIXTURE NECK FRACTURE WERE OBSERVED. PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Pain; Unspecified Infection

    MAUDE report 11680634 ↗
  • InjuryApr 16, 2021Fracture; Fracture

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE ALMOST TWO YEARS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 1, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. DURING THE SURGERY, LOCAL INFECTION, BONE RESORPTION, PERI-IMPLANTITIS, PAIN, AND FIXTURE NECK FRACTURE WERE OBSERVED. PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Pain; Unspecified Infection

    MAUDE report 11680625 ↗
  • InjuryApr 7, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE LESS THAN A MONTH AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. DURING THE SURGERY, LOCAL INFECTION, PERI-IMPLANTITIS, AND PAIN WERE OBSERVED. PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 11632149 ↗
  • InjuryApr 7, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 3 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. DURING THE SURGERY, LOCAL INFECTION, PERI-IMPLANTITIS, AND PAIN WERE OBSERVED. PATIENT WILL LATER REQUIRE ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 11632145 ↗
  • InjuryMar 20, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THAT THE IMPLANT WAS REMOVED DUE OSSEOINTEGRATION FAILURE APPROXIMATELY 4 MONTHS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE 2, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE OBSERVED. PATIENT HAS RECOVERED SINCE.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 11536696 ↗
  • InjuryOct 26, 2020Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    THE DOCTOR REPORTED THAT IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 1 YEAR AND 6 MONTHS SINCE THE DATE OF SURGERY. BONE TYPE OBSERVED IN THE ARE WAS TYPE 1, AND ORAL HYGIENE AROUND THE SITE OF SURGERY WAS MODERATE. DURING THE SURGERY, PERI-IMPLANTITIS WAS OBSERVED. PATIENT HAS RECOVERED SINCE.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Patient Problem/Medical Problem

    MAUDE report 10738266 ↗
  • InjuryJul 31, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DENTAL IMPLANT FX3810MLC WAS REMOVED ON (B)(6) 2020 DUE TO OSSEOINTEGRATION FAILURE AROUND 3 MONTHS AND 6 DAYS AFTER IMPLANTATION. THE PATIENT HAS HISTORY OF CANCER AND HYPERTENSION. THE DOCTOR NOTED LOCAL INFECTION AND PAIN DURING EXPLANTATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.

    Patient problems: Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 10353221 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041368. One submission often covers many models, so check the product and lot details in each record.

  • Z-1119-2007 · Jun 29, 2007 · Terminated
    Biomedicare Implantium Fixtures, Model numbers : FX3408, FX3410, FX3412, FX3414, FX3808, FX3810, FX3812, FX3814, FX4308, FX4310, FX4312, FX4314, FX4808, FX4810, FX4812, FX4814, FX4

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →