Implantium FX3812MLCRecalls & reported problems
Dentium USA · Endosseous dental implant · Class 2 · Implant
All models of Implantium (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI D755FX3812MLC0), received between May 22, 2017 and Jun 3, 2025 — classified as 13 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (9 reports).
1 other recall cover devices cleared under the same FDA submission (K041368); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate9
- Osseointegration Problem7
- Loss of Osseointegration2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- Inadequate Osseointegration12
- Failure of Implant5
- Unspecified Infection3
- Patient Problem/Medical Problem1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 22135157 ↗InjuryNov 17, 2023Fracture; Fracture
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE APPROXIMATELY 9 YEARS AND 2 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THERE WAS NO RELATED TRAUMA HISTORY REPORTED. THE PATIENT HAS SINCE RECOVERED BUT WILL NEED ANOTHER IMPLANT.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 18162561 ↗InjuryJan 4, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 16099544 ↗InjurySep 16, 2022Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 2 YEARS 11 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 15442819 ↗InjuryJul 13, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, FIVE MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 15002720 ↗InjuryMar 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE, APPROXIMATELY 5 YEARS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS NOT REPORTED, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE DOCTOR REPORTED NO SIGNIFICANT FINDINGS DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
MAUDE report 13765787 ↗InjurySep 6, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 6 MONTHS AFTER PLACEMENT SURGERY. BONE QUALITY WAS TYPE II AND ORAL HYGIENE AT THE SITE OF THE IMPLANT WAS GOOD. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE OBSERVED. THE PATIENT HAS RECOVERED.
Patient problems: Inadequate Osseointegration; Patient Problem/Medical Problem
MAUDE report 10498281 ↗InjuryJan 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2019 DUE TO FAILURE TO OSSEOINTEGRATE 7 MONTHS AND 19 DAYS AFTER IMPLANTATION. THE PATIENT HAS NO SIGNIFICANT MEDICAL HISTORY. THE DOCTOR NOTED LOCAL INFECTION DURING EXPLANTATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 9541008 ↗InjuryDec 9, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2019 DUE TO FAILURE TO OSSEOINTEGRATE 3 MONTHS AND 9 DAYS AFTER IMPLANTATION. THE PATIENT HAS HISTORY OF BRUXISM. THE DOCTOR NOTED BONE RESORPTION DURING EXPLANTATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.
Patient problems: Inadequate Osseointegration
MAUDE report 9440874 ↗InjuryAug 7, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2019 DUE TO FAILURE TO OSSEOINTEGRATE 19 DAYS AFTER IMPLANTATION. THE PATIENT HAS HISTORY OF HYPERTENSION. THE DOCTOR NOTED LOCAL INFECTION DURING EXPLANTATION. THE PATIENT REQUIRED BONE AUGMENTATION FOR THE OPERATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 8871834 ↗InjuryOct 24, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2018 DUE TO FAILURE TO OSSEOINTEGRATE 3 MONTHS AND 10 DAYS AFTER IMPLANTATION. THE PATIENT HAS MEDICAL HISTORY OF HYPERTENSION, DIABETES, AND DRUG ABUSE. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.
Patient problems: Inadequate Osseointegration
MAUDE report 8002658 ↗InjuryJul 18, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2017 DUE TO LOSS OF OSSEOINTEGRATION 1 MONTH AFTER IMPLANTATION. THE DOCTOR INDICATED LACK OF PRIMARY STABILITY WAS CAUSE FOR IMPLANT REMOVAL. THE PATIENT HAS NO SIGNIFICANT MEDICAL HISTORY. THE PATIENT HAS BEEN RECOVERED WITHOUT COMPLICATIONS.
Patient problems: Inadequate Osseointegration
MAUDE report 6722024 ↗InjuryMay 22, 2017Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem
THE DENTAL IMPLANT, FX3812MLC, WAS REMOVED ON (B)(6) 2017 DUE TO LOSS OF OSSEOINTEGRATION 2 YEARS AND 5 MONTHS AFTER IMPLANTATION. THE DOCTOR INDICATED THAT NO FINDINGS WERE OBSERVED DURING THE SURGERY. THE PATIENT HAS MEDICAL HISTORY OF BRUXISM. THE PATIENT NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Inadequate Osseointegration
MAUDE report 6583519 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041368. One submission often covers many models, so check the product and lot details in each record.
- Z-1119-2007 · Jun 29, 2007 · TerminatedBiomedicare Implantium Fixtures, Model numbers : FX3408, FX3410, FX3412, FX3414, FX3808, FX3810, FX3812, FX3814, FX4308, FX4310, FX4312, FX4314, FX4808, FX4810, FX4812, FX4814, FX4
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page