Implantium FX3812MLCRecalls & reported problems

Dentium USA · Endosseous dental implant · Class 2 · Implant

All models of Implantium (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI D755FX3812MLC0), received between May 22, 2017 and Jun 3, 2025 — classified as 13 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (9 reports).

1 other recall cover devices cleared under the same FDA submission (K041368); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate9
  • Osseointegration Problem7
  • Loss of Osseointegration2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1

Patient problems

  • Inadequate Osseointegration12
  • Failure of Implant5
  • Unspecified Infection3
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 22135157 ↗
  • InjuryNov 17, 2023Fracture; Fracture

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE APPROXIMATELY 9 YEARS AND 2 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THERE WAS NO RELATED TRAUMA HISTORY REPORTED. THE PATIENT HAS SINCE RECOVERED BUT WILL NEED ANOTHER IMPLANT.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 18162561 ↗
  • InjuryJan 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 16099544 ↗
  • InjurySep 16, 2022Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 2 YEARS 11 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 15442819 ↗
  • InjuryJul 13, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, FIVE MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 15002720 ↗
  • InjuryMar 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO IMPLANT FRACTURE, APPROXIMATELY 5 YEARS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS NOT REPORTED, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE DOCTOR REPORTED NO SIGNIFICANT FINDINGS DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    MAUDE report 13765787 ↗
  • InjurySep 6, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 6 MONTHS AFTER PLACEMENT SURGERY. BONE QUALITY WAS TYPE II AND ORAL HYGIENE AT THE SITE OF THE IMPLANT WAS GOOD. DURING THE SURGERY, NO SIGNIFICANT FINDINGS WERE OBSERVED. THE PATIENT HAS RECOVERED.

    Patient problems: Inadequate Osseointegration; Patient Problem/Medical Problem

    MAUDE report 10498281 ↗
  • InjuryJan 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2019 DUE TO FAILURE TO OSSEOINTEGRATE 7 MONTHS AND 19 DAYS AFTER IMPLANTATION. THE PATIENT HAS NO SIGNIFICANT MEDICAL HISTORY. THE DOCTOR NOTED LOCAL INFECTION DURING EXPLANTATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 9541008 ↗
  • InjuryDec 9, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2019 DUE TO FAILURE TO OSSEOINTEGRATE 3 MONTHS AND 9 DAYS AFTER IMPLANTATION. THE PATIENT HAS HISTORY OF BRUXISM. THE DOCTOR NOTED BONE RESORPTION DURING EXPLANTATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 9440874 ↗
  • InjuryAug 7, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2019 DUE TO FAILURE TO OSSEOINTEGRATE 19 DAYS AFTER IMPLANTATION. THE PATIENT HAS HISTORY OF HYPERTENSION. THE DOCTOR NOTED LOCAL INFECTION DURING EXPLANTATION. THE PATIENT REQUIRED BONE AUGMENTATION FOR THE OPERATION. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 8871834 ↗
  • InjuryOct 24, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2018 DUE TO FAILURE TO OSSEOINTEGRATE 3 MONTHS AND 10 DAYS AFTER IMPLANTATION. THE PATIENT HAS MEDICAL HISTORY OF HYPERTENSION, DIABETES, AND DRUG ABUSE. THE PATIENT HAS RECOVERED WITHOUT COMPLICATIONS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 8002658 ↗
  • InjuryJul 18, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    THE DENTAL IMPLANT FX3812MLC WAS REMOVED ON (B)(6) 2017 DUE TO LOSS OF OSSEOINTEGRATION 1 MONTH AFTER IMPLANTATION. THE DOCTOR INDICATED LACK OF PRIMARY STABILITY WAS CAUSE FOR IMPLANT REMOVAL. THE PATIENT HAS NO SIGNIFICANT MEDICAL HISTORY. THE PATIENT HAS BEEN RECOVERED WITHOUT COMPLICATIONS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 6722024 ↗
  • InjuryMay 22, 2017Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    THE DENTAL IMPLANT, FX3812MLC, WAS REMOVED ON (B)(6) 2017 DUE TO LOSS OF OSSEOINTEGRATION 2 YEARS AND 5 MONTHS AFTER IMPLANTATION. THE DOCTOR INDICATED THAT NO FINDINGS WERE OBSERVED DURING THE SURGERY. THE PATIENT HAS MEDICAL HISTORY OF BRUXISM. THE PATIENT NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 6583519 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041368. One submission often covers many models, so check the product and lot details in each record.

  • Z-1119-2007 · Jun 29, 2007 · Terminated
    Biomedicare Implantium Fixtures, Model numbers : FX3408, FX3410, FX3412, FX3414, FX3808, FX3810, FX3812, FX3814, FX4308, FX4310, FX4312, FX4314, FX4808, FX4810, FX4812, FX4814, FX4

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices