Skip to content

SuperLine FX4510SWCRecalls & reported problems

Dentium USA · Endosseous dental implant · Class 2 · Implant

Quick summary

This Dentium SuperLine dental implant is a screw placed in the jawbone to anchor a replacement tooth. All 129 reports are injuries, with no deaths; most describe the implant not fusing with the bone, and many describe implants removed later, sometimes with infection or bone loss. No reports came in during the last 12 months, compared with 6 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
129
Deaths reported
0
Injuries reported
129
Recalls naming it
0

Summary

The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI D755FX4510SWC0), received between Mar 1, 2016 and Aug 19, 2025 — classified as 129 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (93 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate93
  • Osseointegration Problem87
  • Loss of Osseointegration34
  • Use of Device Problem9
  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1

Patient problems

  • Inadequate Osseointegration127
  • Failure of Implant52
  • Unspecified Infection26
  • Patient Problem/Medical Problem11
  • Pain5

Recent adverse event reports

The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 19, 2025Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 8 YEARS AND 6 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE IMPLANT WAS REMOVED, AND THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 22845602 ↗
  • InjuryJun 3, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. PAIN WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain

    MAUDE report 22135158 ↗
  • InjuryMay 13, 2025Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INFECTION OF NEIGHERING TOOTH SPREAD TO IMPLANT AROUND 2 YEARS AND 6 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION, PERIODONTAL DISEASE, BONE RESORPTION WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 22015377 ↗
  • InjuryDec 31, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 10 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION AND PERI-IMPLANTITIS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21050435 ↗
  • InjuryNov 6, 2024Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 1 YEAR AND 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. BONE RESORPTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20625148 ↗
  • InjuryNov 6, 2024Loss of Osseointegration; Loss of Osseointegration

    THE FIXTURE WITH ATTACHED CROWN WAS RETURNED. SLA SURFACE TREATMENT WAS CONFIRMED. EVIDENCE OF OSSEOINTEGRATION WAS OBSERVED. THE VISUAL EXAMINATION WAS ACCEPTABLE. THE DOCTOR REPORTED ON FORM F1060, REV 8 THAT THE RETURNED FIXTURE WAS FSX4510C. THE RELATED SALES ORDER # (B)(4) CONFIRMED THAT FX4510SWC LOT K21OB460S WAS ORDERED AND SHIPPED. THIS WAS CONFIRMED WITH CASSIDY RYAN FROM HADDON ORAL'S OFFICE BY DENTIUM STAFF MK. THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 1 YEAR AND 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE I. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20616309 ↗
  • InjuryAug 26, 2024Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 1 YEARS AND 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. PERI-IMPLANTITIS AND BONE RESORPTION WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20071749 ↗
  • InjuryJul 5, 2024Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 9 YEARS AND 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. PERI-IMPLANTITIS WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 19684789 ↗
  • InjuryFeb 12, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 18692026 ↗
  • InjuryJul 28, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 9 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 17420925 ↗
  • InjuryJul 24, 2023Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 7 YEARS AND 2 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 17385627 ↗
  • InjuryJul 10, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 17294532 ↗
  • InjuryJul 10, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 11 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED, BUT WILL ALSO NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 17293877 ↗
  • InjuryApr 21, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 2 MONTHS AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. THE DOCTOR REPORTED NO SIGNIFICANT FINDING DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 16792056 ↗
  • InjuryJan 27, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND THREE MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 16257647 ↗
  • InjuryDec 21, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, LESS THAN THREE MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 16024788 ↗
  • InjuryOct 12, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE AROUND THREE MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 15588174 ↗
  • InjuryAug 20, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INSUFFICIENT PRIMARY STABILITY, LESS THAN ONE MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 15263172 ↗
  • InjuryAug 20, 2022Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION AROUND ONE YEAR AND ONE MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. ABNORMAL SENSATION AROUND THE PLACEMENT WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 15263137 ↗
  • InjuryMay 25, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INSUFFICIENT PRIMARY STABILITY, LESS THAN TWO MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 14500083 ↗
  • InjuryMar 14, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN ONE YEAR, APPROXIMATELY 3.5 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED NO SIGNIFICANT FINDINGS DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 13759844 ↗
  • InjuryMar 14, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN ONE YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED LOCAL INFECTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 13758945 ↗
  • InjuryFeb 23, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN ONE YEAR, APPROXIMATELY 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED LOCAL INFECTION DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 13588824 ↗
  • InjuryFeb 23, 2022Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION AFTER ONE YEAR, APPROXIMATELY 3 YEARS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. THE DOCTOR REPORTED THAT THE IMPLANT FELL OUT, AND NO SURGERY WAS REQUIRED. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 13586577 ↗
  • InjuryFeb 21, 2022Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION WITHIN ONE YEAR, APPROXIMATELY 3 YEARS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. THE DOCTOR REPORTED THAT THE IMPLANT FELL OUT, AND NO SURGERY WAS REQUIRED. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 13573402 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.