SuperLine II FXS4008CRecalls & reported problems
Dentium USA · Endosseous dental implant · Class 2 · Implant
All models of SuperLine II (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Dentium SuperLine II dental implant is a root-form implant placed in the jaw to support teeth. The 25 reports mostly describe implants failing to bond with the surrounding bone, leading to removal and repeat surgery, sometimes with infection or bone loss. There are no deaths and 25 injuries mentioned, and no recalls name this
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 25
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI D755FXS4008C0), received between Sep 15, 2021 and Jun 11, 2026 — classified as 25 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate21
- Loss of Osseointegration4
Patient problems
- Inadequate Osseointegration25
- Failure of Implant24
- Unspecified Infection8
- Pain2
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
InjuryJun 11, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 3 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. BONE RESORPTION AND PERI-IMPLANTITIS WERE OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 25480281 ↗InjuryMar 25, 2026Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION AFTER 1 YEAR, APPROXIMATELY 4 YEARS AND 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. LOCAL INFECTION AND BONE RESORPTION WERE OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT NEEDS ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 24683560 ↗InjuryMar 24, 2026Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION AFTER 1 YEAR, APPROXIMATELY 4 YEARS AND 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT NEEDS ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 24678288 ↗InjuryFeb 13, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. THE DOCTOR REPORTED THAT THE IMPLANT FELL OUT BY ITSELF DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 24339749 ↗InjurySep 2, 2025Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 1 YEAR AND 2 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 22948020 ↗InjuryJul 28, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, AROUND 14 DAYS AFTER IMPLANT PLACEMENT. THE PATIENT'S X-RAY WAS PROVIDED. THE BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS NOT REPORTED. LOCAL INFECTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 22638631 ↗InjuryJul 15, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 9 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS NOT REPORTED. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 22515591 ↗InjuryJul 15, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 9 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 22515590 ↗InjuryMar 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 2 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. BONE RESORPTION WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 21627475 ↗InjuryFeb 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. NO SIGNIFICANT FINDING WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 21404967 ↗InjuryDec 12, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 5 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 20915342 ↗InjuryNov 26, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 5 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. LOCAL INFECTION, BONE RESORPTION AND PERI-IMPLANTITIS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 20788949 ↗InjurySep 20, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS NOT REPORTED, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. PERI-IMPLANTITIS WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 20268145 ↗InjurySep 20, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 5 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS NOT REPORTED. PERI-IMPLANTITIS WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 20268134 ↗InjuryApr 26, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THERE WERE NO FINDINGS REPORTED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 19192601 ↗InjuryJan 9, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. LOCAL INFECTION, FAILURE OF INTEGRATION AND MOBILITY WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION IN 6 MONTHS.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 18480613 ↗InjuryJan 8, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 2 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. PERI IMPLANTITIS WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 18472160 ↗InjuryNov 17, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. PAIN AND ABNORMAL SENSATION AROUND IMPLANT PLACEMENT WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain
MAUDE report 18162585 ↗InjuryNov 1, 2023Loss of Osseointegration; Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 1 YEAR AND 7 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS POOR. PERIODONTAL DISEASE WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 18049020 ↗InjuryOct 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 18006523 ↗InjuryJul 10, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 17293881 ↗InjuryJun 30, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INFECTION LESS THAN 3 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. OTHER THAN THE INFECTION CAUSE, NO FINDING WAS REPORTED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 17243584 ↗InjuryJun 9, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 2 MONTHS AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED BONE RESORPTION AND ABONORMAL SENSATION AROUND THE IMPLANT PLACEMENT WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 17101581 ↗InjurySep 16, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, ONE MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. PERI-IMPLANTITIS AND PAIN WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain
MAUDE report 15442817 ↗InjurySep 15, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE APPROXIMATELY 24 DAYS AFTER INITIAL IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 12477931 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page